Marcin Cierniak1, Dariusz Timler1, Andrzej Wieczorek2, Przemyslaw Sekalski3, Natalia Borkowska2, Tomasz Gaszynski4.
Abstract
Currently, there are plenty of videolaryngoscopes that appear on the market. They have different specifications. Some of these features favor the fact that they are more suited for educational purposes of future operators and others can be characterized with an excellent clinical use. In this study we compared four types of videolaryngoscopes. The aim of the study was to compare the technical specifications of the above-mentioned devices for usefulness in clinical practice and correlate these parameters with the subjective evaluation of these videolaryngoscopes usage performed in practice by an experienced medical staff. All devices considered in this study participated in another multicenter clinical study on the basis of which we completed the subjective evaluation of the operators. In order to examine the technical parameters of the equipment we established the cooperation with the Department of Microelectronics at Technical University of Lodz. Mechanical and optical parameters and the endoscopic tube current were taken into consideration. The C-MAC has a camera with the widest viewing angle (the OX axis-63.1, the axis OY-47.8), which in combination with the largest diagonal size of the display enables the operator to see the details relevant to clinical practice. It has also the strongest lamp intensity of the devices mentioned in this comparison (7800 Lx). In comparison of the clinical use in almost all compared parameters the Cmac D-blade is a winner, although for clinical education purpose we consider the Vividtrac a better device.Entities:
Keywords: Cmac D-blade; Kingvision; McGrath Mac; Vividtrac
Mesh:
Year: 2015 PMID: 26120020 PMCID: PMC4960280 DOI: 10.1007/s10877-015-9727-2
Source DB: PubMed Journal: J Clin Monit Comput ISSN: 1387-1307 Impact factor: 2.502
Fig. 1McGrath Mac videolaryngoscope. Source http://www.covidien.com/rms/products/video-laryngoscopy/mcgrath-video-laryngoscope
Fig. 2Kingvision videolaryngoscope. Source http://www.owntheairway.com/king-video-laryngoscopes/
Fig. 3Vividtrac. Source http://mercurymed.com/catalog2/index.php?type=87
Fig. 4Cmac D-blade. Source http://www.medgadget.com/2011/01/dblade_from_karl_storz_improves_view_in_difficult_airway.html
Fig. 12The comparison of the images in the same patient. Left image captured using the VividTrack, right one through the C-MAC
Fig. 5The field of view of each device. VividTrack VT-A100 (red), C-MAC 8401 ZX (black), KingVision (blue), McGRATH MAC (green). The dots represent corners of a square with a side measuring 1 cm. The imaging plane at the distal blade tip
The distance between the camera and the distal blade tip
| McGrath MAC | KingVision | VividTrack VT-A100 | C-MAC 8401 ZX | |
|---|---|---|---|---|
| The distance between the camera and the distal blade tip (mm) | 60 | 34 | 60 | 35 |
Individual values of the intensity of the endoscopic light
| McGrath MAC | KingVision | VividTrack VT-A100 | C-MAC 8401 ZX | |
|---|---|---|---|---|
| C mercury (Lx) | 540 | 6800 | 1900 | 8000 |
| L incandescent (Lx) | 620 | 6600 | 1880 | 6600 |
| F fluorescent (Lx) | 540 | 5500 | 1830 | 7200 |
| S daylight (Lx) | 570 | 6000 | 1860 | 7800 |
| Lamp type | White led | LED |
Optical parameters
| Screen type | McGrath MAC | KingVision | VividTrack VT-A100 | C-MAC 8401 ZX |
|---|---|---|---|---|
| Vertical | Horizontal | Horizontal | Horizontal | |
| Area observed in the OX axis (mm) | 30 | 34 | 58 | 43 |
| Area observed in the OY axis (mm) | 41 | 24 | 28 | 23 |
| The virtual image of the OY axis (mm) | 44 | 26 | 45 | 31 |
| The viewing angle of the OX axis (°) | 28.1 | 51.8 | 51.6 | 63.1 |
| The viewing angle of the OY axis (°) | 40.3 | 40.8 | 41.1 | 47.8 |
| Reference to the best result for the OX axis (%) | 44 | 82 | 82 | 100 |
| Reference to the best result for the OY axis (%) | 84 | 85 | 86 | 100 |
Screen parametres
| McGrath MAC | KingVision | VividTrack VT-A100 | C-MAC 8401 ZX | |
|---|---|---|---|---|
| Screen resolution (px) | b.d. | 320 × 240 QVGA | n.d. | 800 × 480 |
| Screen technology | LED | OLED | n.d. | LCD |
| Size of the screen (inch) | 2.5 | 2.4 | n.d. | 7.0 |
| Recording speed (fps) | b.d. | 30 | 30 | 12 |
| Camera resolution (px) | b.d. | 640 × 480 VGA | 640 × 480 VGA | 320 × 240 QVGA |
| Sensor technology | CMOS | CMOS | CMOS | CMOS |
| Pixel density [resolution (px/cm PPCM)] | b.d. | 187 | 109 | 75 |
Fig. 6The comparison of test images recorded with the C-MAC (left) and the VividTrack (right) at a magnification of ×16
Fig. 7Test pattern
Fig. 8Test image of the C-MAC
Fig. 9Test image of the C-MAC
Fig. 10Test image of the VividTrack
Fig. 11Test image of the VividTrack
Fare of each device (in PLN)
| McGrath MAC | KingVision | VividTrack VT-A100 | C-MAC 8401 ZX | |
|---|---|---|---|---|
| Fare | 26,000 | 8000 | 500 | 29,000 |