| Literature DB >> 26111971 |
Bin Yu1, Xiaoxue Hu2, Tianxiao Zou1, Miao He1, Guangyu Cai1.
Abstract
BACKGROUND: The aim of this study was to compare the novel needle-over-cannula vs. the Braun continuous peripheral nerve block catheter set (Contiplex A) in postoperative analgesics after total knee arthroplasty (TKA).Entities:
Mesh:
Substances:
Year: 2015 PMID: 26111971 PMCID: PMC4494567 DOI: 10.12659/MSM.893617
Source DB: PubMed Journal: Med Sci Monit ISSN: 1234-1010
Figure 1(A) A schematic of the novel needle-over-cannula; (B) The novel needle-over-cannula (22-gauge needle, 90-mm-length external indwelling cannula; Tuoren Group, Xinxiang, China) invented by Dr. Bin Yu, who is also the first author of the article.
Figure 2The novel needle-over-cannula was inserted out-plane at an angle of about 30° to skin, vertical to arcus cruralis, parallel to femoral nerve.
Figure 3Ultrasound of the perineural fluid space with injection of 20 ml 0.5% ropivacaine after the needle breaking through fascia iliaca and fascia lata.
Figure 4After the needle was pulled out and the cannula was fixed to the skin with no-pressure medical tape.
Figure 5Ultrasound image of the cannula after procedures.
Situations and complications of the 2 groups.
| Braun group | Cannula group | P-value | |
|---|---|---|---|
| No. | 30 | 30 | n.s. |
| Placement failure | 4/30 | 0/30 | n.s. |
| Single attempt success | 17/26 | 29/30 | <0.05 |
| Time of placement (min) | 13.13±3.29 | 5.84±0.77 | <0.05 |
| Indwelling depth (cm) | 9.73±1.49 | 7.24±1.15 | n.s. |
| Catheter or cannula occlusion | 0 | 0 | n.s. |
| Catheter or cannula slipping | 0 | 0 | n.s. |
| Infection at puncture site | 0 | 0 | n.s. |
| Nerve injury related complication | 0 | 0 | n.s. |
| Severe intoxication | 0 | 0 | n.s. |
| The maximal degree of knee flexion | 87.69°±12.43° | 88°±12.97° | n.s. |
| Patient satisfaction | 8.65±0.85 | 8.63±1.03 | n.s. |
n.s. – not significant.
P<0.05 considered statistically significant.
Postoperative follow-up of the 2 groups.
| VAS (no activities/activities) | Muscle strength of four unit head muscle | Modified Bromage score | Local anesthetic leakage at puncture site | Sense nerve block LFCN/FN/ON | Nausea/ vomiting | Parecoxib | ||
|---|---|---|---|---|---|---|---|---|
| 6 h | Braun group | 0.54±1.07/2.35±1.98 | 3.81±0.69 | 0.65±0.63 | 7/26 | 1/7/0 | 0/26 | 1/26 |
| Cannula group | 0.40±0.81/2.07±1.44 | 4±0.53 | 0.67±0.61 | 0/30 | 7/15/1 | 0/26 | 0/30 | |
| n.s. | n.s. | n.s. | <0.05 | n.s | n.s | n.s | ||
| 12 h | Braun group | 1.35±1.41/3.19±2.79 | 4.23±0.76 | 0.46±0.58 | 12/26 | 0/2/1 | 2/26 | 2/26 |
| Cannula group | 1.1±1.40/2.27±2.15 | 4.43±0.68 | 0.50±0.68 | 0/30 | 0/7/0 | 1/30 | 1/30 | |
| n.s | n.s | n.s | <0.05 | n.s | n.s | n.s | ||
| 24 h | Braun group | 1.39±1.36/4.12±1.88 | 4.54±0.58 | 0.38±0.57 | 12/26 | 1/3/1 | 0/26 | 0/26 |
| Cannula group | 0.9±1.34/2.87±2.01 | 4.73±0.52 | 0.20±0.48 | 0/30 | 2/3/0 | 0/30 | 2/30 | |
| <0.05 | n.s | n.s | <0.05 | n.s | n.s | n.s | ||
| 48 h | Braun group | 1.31±1.26/3.85±1.60 | 4.77±0.51 | 0.15±0.37 | 12/26 | 0/0/0 | 0/26 | 0/26 |
| Cannula group | 1.27±1.23/3.33±1.40 | 4.90±0.31 | 0.13±0.35 | 0/30 | 0/0/0 | 0/30 | 1/30 | |
| n.s | n.s | n.s | <0.05 | n.s | n.s | n.s |
n.s. – not significant.
P<0.05considered statistically significant.