Ying Liu1, Feng Chen1, Xiaodong Zhu2, Lisong Shen3, Sean X Zhang4.
Abstract
We evaluated a novel plasma (1,3)-β-d-glucan (BDG) detection assay for the diagnosis of candidemia in children. The median BDG levels were 73.4 pg/ml in patients with candidemia and <10 pg/ml in patients without candidemia (P < 0.001). Receiver operating characteristic analysis revealed a cutoff point of 14 pg/ml and an area under the curve of 0.802. At these values, the assay demonstrated 68% sensitivity, 91% specificity, 66% positive predictive value, and 91% negative predictive value. Plasma BDG levels were undetectable in 18 candidemia cases.Entities:
Mesh:
Substances:
Year: 2015 PMID: 26109441 PMCID: PMC4540920 DOI: 10.1128/JCM.00673-15
Source DB: PubMed Journal: J Clin Microbiol ISSN: 0095-1137 Impact factor: 5.948