Optimal force magnitude loaded to orthodontic microimplants: A finite element analysis.

Raed H Alrbata1, Moath Q Momani2, Ahmad M Al-Tarawneh3, Ayman Ihyasat3.   

Abstract

OBJECTIVE: To find an optimal force that can be loaded onto an orthodontic microimplant to fulfill the biomechanical demands of orthodontic treatment without diminishing the stability of the microimplant.
MATERIALS AND METHODS: Using the finite element analysis method, 3-D computer-aided design models of a microimplant and four cylindrical bone pieces (incorporating cortical bone thicknesses of 0.5, 1.2, 2.0, and 3.0 mm) into which the microimplant was inserted were used. Various force magnitudes of 0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, and 4.0 N were then horizontally and separately applied to the microimplant head as inserted into the different bone assemblies. For each bone/force assembly tested, peak stresses developed at areas of intimate contact with the microimplant along the force direction were then calculated using regression analysis and compared with a threshold value at which pathologic bone resorption might develop.
RESULTS: The resulting peak stresses showed that bone pieces with thicker cortical bone tolerated higher force magnitudes better than did thinner ones. For cortical bone thicknesses of 0.5, 1.2, 2.0, and 3.0 mm, the maximum force magnitudes that could be applied safely were 3.75, 4.1, 4.3, and 4.45 N, respectively.
CONCLUSIONS: For the purpose of diminishing orthodontic microimplant failure, an optimal force that can be safely loaded onto a microimplant should not exceed a value of around 3.75-4.5 N.

Keywords:  Optimal force; Orthodontic microimplant

Mesh:

Substances:

Year:  2015        PMID: 26098865     DOI: 10.2319/031115-153.1

Source DB:  PubMed          Journal:  Angle Orthod        ISSN: 0003-3219            Impact factor:   2.079


  2 in total

1.  Evaluation of factors influencing the success rate of orthodontic microimplants using panoramic radiographs.

Authors:  Jae Hyun Park; Jong-Moon Chae; R Curtis Bay; Mi-Jung Kim; Keun-Young Lee; Na-Young Chang
Journal:  Korean J Orthod       Date:  2017-11-19       Impact factor: 1.372

2.  The correlation among gripping volume, insertion torque, and pullout strength of micro-implant.

Authors:  Chun-Chan Ting; Kun-Jung Hsu; Szu-Yu Hsiao; Chun-Ming Chen
Journal:  J Dent Sci       Date:  2020-07-16       Impact factor: 2.080

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.