Yang Liu1, Dong Bian1, Yuanhao Wu2, Nan Li1, Kejin Qiu3, Yufeng Zheng4, Yong Han5.
Abstract
Mg-1Ca samples were implanted with biocompatible alloy ions Ag, Fe and Y respectively with a dose of 2×10(17)ionscm(-2) by metal vapor vacuum arc technique (MEVVA). The surface morphologies and surface chemistry were investigated by SEM, AES and XPS. Surface changes were observed after all three kinds of elemental ion implantation. The results revealed that the modified layer was composed of two sublayers, including an outer oxidized layer with mixture of oxides and an inner implanted layer, after Ag and Fe ion implantation. Y ion implantation induced an Mg/Ca-deficient outer oxidized layer and the distribution of Y along with depth was more homogeneous. Both electrochemical test and immersion test revealed accelerated corrosion rate of Ag-implanted Mg-1Ca and Fe-implanted Mg-1Ca, whereas Y ion implantation showed a short period of protection since enhanced corrosion resistance was obtained by electrochemical test, but accelerated corrosion rate was found by long period immersion test. Indirect cytotoxicity assay indicated good cytocompatibility of Y-implanted Mg-1Ca. Moreover, the corresponding corrosion mechanisms involving implanting ions into magnesium alloys were proposed, which might provide guidance for further application of plasma ion implantation to biodegradable Mg alloys.Entities:
Keywords: Biocompatibility; Corrosion property; Mechanism; Mg–Ca alloy; Plasma ion implantation
Mesh:
Substances:
Year: 2015 PMID: 26094143 DOI: 10.1016/j.colsurfb.2015.05.050
Source DB: PubMed Journal: Colloids Surf B Biointerfaces ISSN: 0927-7765 Impact factor: 5.268