Michael P Müller1, Cynthia Poenicke2, Maxi Kurth3, Torsten Richter4, Thea Koch5, Carolin Eisold6, Adrian Pfältzer7, Axel R Heller8.
Abstract
BACKGROUND: Basic life support (BLS) guidelines focus on chest compressions with a minimal no-flow fraction (NFF), early defibrillation, and a short perishock pause. By using an automated external defibrillator (AED) lay persons are guided through the process of attaching electrodes and initiating defibrillation. It is unclear, however, to what extent the voice instructions given by the AED might influence the quality of initial resuscitation.Entities:
Mesh:
Year: 2015 PMID: 26094032 PMCID: PMC4475613 DOI: 10.1186/s13049-015-0123-1
Source DB: PubMed Journal: Scand J Trauma Resusc Emerg Med ISSN: 1757-7241 Impact factor: 2.953
BLS with the use of an AED in scenarios of persistent ventricular fibrillationThe duration of the scenario was 300 seconds
| Time until 1st chest compression [seconds] | Time until completion of 1st rhythm analysis [seconds] | Duration of rhythm analysis [seconds] | Time until 1st shock [seconds] | Interval between 1st shock and 2nd rhythm analysis [seconds] | Perishock pause [seconds] | No-flow fraction | |
|---|---|---|---|---|---|---|---|
| (A) Heartsave PAD | 50 ± 3 I | 47 ± 3D,I | 13 ± 1B,C,D,F,G | 63 ± 4D | 138 ± 0C | 12 ± 0+ | 0.51 ± 0.01C,E,G,+ |
| (B) Philips HS1§ | 125 ± 2 | 77 ± 3C,G,H | 13 ± 2A,C,D,E,F,G,H | 88 ± 2C | 194 ± 3 | 44 ± 4G | 0.72 ± 0.01F |
| (C) Philips HS1* | 103 ± 1G,+ | 77 ± 3B,G,H,+ | 10 ± 1A,B,D,E,F,H,I | 90 ± 2B,+ | 140 ± 1A | 29 ± 1E,H,+ | 0.52 ± 0.01A,G,+ |
| (D) Lifeline VIEW | 68 ± 3+ | 48 ± 2A,I,+ | 11 ± 1A,B,C,E,F,H,I | 56 ± 3A,I,+ | 153 ± 3 | 34 ± 2E,F,+ | 0.61 ± 0.03+ |
| (E) Lifepak CR Plus | 81 ± 1H,+ | 63 ± 1 H,+ | 10 ± 1B,C,D,F,H,I | 75 ± 1G,H,+ | 125 ± 1G,H,I | 31 ± 1C,D,+ | 0.48 ± 0A,G,H,+ |
| (F) PowerHeart G5 | 147 ± 1+ | 93 ± 1+ | 11 ± 1A,B,C,D,E,H | 101 ± 1+ | 183 ± 1 | 38 ± 1D,+ | 0.71 ± 0B,+ |
| (G) Fred Easy Life | 105 ± 1C,+ | 74 ± 1B,C,H,+ | 15 ± 0A,B | 80 ± 1E,H,+ | 124 ± 1E,H,I | 46 ± 0B,+ | 0.50 ± 0A,C,E,+ |
| (H) AED Plus | 81 ± 3E,+ | 71 ± 2B,C,E,G,+ | 9 ± 1B,C,D,E,F,I | 77 ± 2E,G,+ | 121 ± 1E,G,I | 25 ± 1C,I,+ | 0.45 ± 0.01E |
| (I) Cardiolife AED 2100 | 55 ± 1A,+ | 41 ± 2A,D,+ | 7 ± 1C,D,H,I | 48 ± 2+ | 126 ± 1E,G,H | 21 ± 2H | 0.37 ± 0.01 |
| (K) BLS without AED | 25 ± 2 | 0.26 ± 0.01 |
Values are the mean ± standard deviation. The Philips HS1 asks the rescuer to press an “info button” for detailed information about chest compressions and ventilations after each shock. The scenario was run 3 times when pressing the info button§ and 3 times without pressing the button*. A lack of statistical significance (p > 0.01) in comparison with a specific model is indicated by the identifying letter (A) – (I). +p < 0.01 against the respective parameter in the scenarios with an optimized sequence of actions
BLS with the use of an AED in scenarios of persistent asystole
| Time until 1st chest compression [seconds] | Time until completion of 1st rhythm analysis [seconds] | Duration of rhythm analysis [seconds] | Interval between completion of 1st rhythm analysis and start of 2nd rhythm analysis [seconds] | No-flow fraction | |
|---|---|---|---|---|---|
| (A) Heartsave PAD | 58+ ± 2 D,E,H,I,+ | 50 ± 3D,H,I | 20 ± 1 | 140 ± 1C | 0.52 ± 0.03C |
| (B) Philips HS1§ | 112 ± 2G | 77 ± 3C,G | 11 ± 1C,D,E,I | 200 ± 1 | 0.71 ± 0F |
| (C) Philips HS1* | 92 ± 2G,+ | 78 ± 2B,G,+ | 12 ± 0B,D,E,I | 140 ± 0A | 0.50 ± 0A,+ |
| (D) Lifeline VIEW | 64 ± 1A,E,H,I,+ | 53 ± 2A,H | 12 ± 1B,C,F,I | 154 ± 1 | 0.58 ± 0.01+ |
| (E) Lifepak CR Plus | 67 ± 1 A,D,H,I,+ | 61 ± 1+ | 10 ± 1B,C,H,I | 125 ± 0G,H,I | 0.42 ± 0.01G,H,I |
| (F) PowerHeart G5 | 148 ± 13+ | 96 ± 2+ | 15 ± 1D,G | 183 ± 1 | 0.72 ± 0B,+ |
| (G) Fred Easy Life | 97 ± 1 B,C,+ | 74 ± 1B,C,+ | 15 ± 0F | 123 ± 1E | 0.45 ± 0E,H,+ |
| (H) AED Plus | 57 ± 1A,D,E,I,+ | 53 ± 2A,D,+ | 8 ± 0E,I | 126 ± 1E,I | 0.41 ± 0.01E,G,I,+ |
| (I) Cardiolife AED 2100 | 52 ± 1A,D,E,H,+ | 45 ± 1A,+ | 10 ± 0B,C,D,E,H | 127 ± 0E,H | 0.40 ± 0.01E,H |
| (K) BLS without AED | 25 ± 2 | 0.26 ± 0.01 |
The duration of the scenario was 300 seconds. Values are the mean ± standard deviation. The Philips HS1 asks the rescuer to press an “info button” for detailed information about chest compressions and ventilations after each rhythm analysis. The scenario was run 3 times when pressing the info button§ and 3 times without pressing the button*. A lack of statistical significance (p > 0.01) in comparison with a specific model is indicated by the identifying letter (A) – (I). +p < 0.01 against the respective parameter in the scenarios with an optimized sequence of actions.
BLS with the use of an AED in scenarios of persistent ventricular fibrillation: optimized sequence of actionsThe duration of the scenario was 300 seconds
| Time until 1st chest compression [seconds] | Time until completion of 1st rhythm analysis [seconds] | Duration of rhythm analysis [seconds] | Time until 1st shock [seconds] | Interval between 1st shock and start of 2nd rhythm analysis [seconds] | Perishock pause [seconds] | No-flow fraction | |
|---|---|---|---|---|---|---|---|
| (A) Heartsave PAD | 35 ± 2B,C,E,G | 34 ± 2 B,C,E | 14 ± 1 B,C,E,F | 47 ± 2 F | 137 ± 1 B | 3 ± 1 D,+ | 0.36 ± 0.01B,C,D,E,F,G,H,+ |
| (B) Philips HS1 | 31 ± 2A,D,E,G,H,K,+ | 28 ± 2 A,D,E,G,H,+ | 14 ± 2 A,C,E,F,G | 30 ± 2 E,G,H,+ | 140 ± 0 A | 10 ± 1 E,H,+ | 0.34 ± 0.01A,C,D,E,F,G,H,+ |
| (C) Lifeline VIEW | 41 ± 1 A,+ | 35 ± 1 A,E,+ | 11 ± 2 A,B,D,E,G,H | 39 ± 1 D,+ | 154 ± 1 | 18 ± 1 F,G,+ | 0.36 ± 0.01A,B,D,E,F,G,H,+ |
| (D) Lifepak CR Plus | 28 ± 2B,E,G,H,K,+ | 27 ± 2 B,E,G,H,+ | 10 ± 1 B,C,E,G,H | 38 ± 3 C,E,+ | 125 ± 1 F,G,H | 1 ± 1 A,+ | 0.32 ± 0.01A,B,C,E,F,H,+ |
| (E) PowerHeart G5 | 34 ± 1A,B,D,G,H,+ | 30 ± 2 A,B,C,D,G,+ | 12 ± 1 A,B,C,D,F,G,H | 32 ± 1 B,D,G,H,+ | 185 ± 2 | 13 ± 2 B,G,H,+ | 0.33 ± 0.01A,B,C,D,F,G,H,+ |
| (F) Fred Easy Life | 17 ± 2+ | 49 ± 1+ | 15 ± 1 A,B,E | 51 ± 2 A,+ | 124 ± 1 D,G,H | 19 ± 1 C,G,+ | 0.35 ± 0.01A,B,C,D,E,G,H,+ |
| (G) AED Plus | 33 ± 2A,B,D,E,H,+ | 27 ± 2 B,D,E,H,+ | 10 ± 1 B,C,D,E,H | 31 ± 2 B,E,H,+ | 125 ± 1 D,F,H | 18 ± 1 C,E,F,+ | 0.37 ± 0.02A,B,C,E,F,H |
| (H) Cardiolife AED 2100 | 27 ± 1B,D,E,G,K,+ | 23 ± 0 B,D,G,+ | 8 ± 1 C,D,E,G | 26 ± 1 B,E,G,+ | 127 ± 2 D,F,G | 11 ± 1 B,E | 0.35 ± 0.02A,B,C,D,E,F,G |
| (K) BLS without AED | 25 ± 2B,D,H | 0.26 ± 0.01 |
Values are the mean ± standard deviation. A lack of statistical significance (p > 0.01) in comparison with a specific model is indicated by the identifying letter (A) – (I). +p < 0.01 against the respective parameter in the scenarios with a standard sequence of actions
BLS with the use of an AED in scenarios with persistent asystole: optimized sequence of actions
| Time until 1st chest compression [seconds] | Time until completion of 1st rhythm analysis [seconds] | Duration of rhythm analysis [seconds] | Interval between completion of 1st rhythm analysis and start of 2nd rhythm analysis [seconds] | No-flow fraction | |
|---|---|---|---|---|---|
| (A) Heartsave PAD | 40 ± 2C,E,+ | 39 ± 3 | 19 ± 2 | 140 ± 1B | 0.41 ± 0.01 |
| (B) Philips HS1 | 26 ± 1D,G,H,K,+ | 24 ± 1D,G,H,+ | 10 ± 1C,D,G,H | 142 ± 1A | 0.33 ± 0.02C,D,E,F,G,H,+ |
| (C) Lifeline VIEW | 33 ± 2A,D,E,G,H,+ | 32 ± 2D,E,+ | 12 ± 1B,D,E,F | 153 ± 0 | 0.36 ± 0.02B,D,E,F,G,H,+ |
| (D) Lifepak CR Plus | 28 ± 2B,C,E,G,H,K,+ | 26 ± 1B,C,E,G,H,+ | 10 ± 1B,C,G,H | 125 ± 0G | 0.33 ± 0.01B,C,E,F,G,H |
| (E) PowerHeart G5 | 34 ± 0A,C,D,G,+ | 32 ± 1C,D,+ | 14 ± 1C,F | 184 ± 1 | 0.34 ± 0.01B,C,D,F,G,H,+ |
| (F) Fred Easy Life | 16 ± 2+ | 48 ± 1+ | 15 ± 1C,E | 122 ± 1G | 0.33 ± 0.01B,C,D,E,G,H,+ |
| (G) AED Plus | 27 ± 2B,C,D,E,H,K,+ | 26 ± 1B,D,H,+ | 9 ± 0B,D,H | 125 ± 1D,F | 0.35 ± 0B,C,D,E,F,H,+ |
| (H) Cardiolife AED 2100 | 27 ± 3B,C,D,G,K,+ | 25 ± 3B,D,G,+ | 9 ± 0B,D,G | 128 ± 1 | 0.35 ± 0B,C,D,E,F,G |
| (K) BLS without AED | 25 ± 2B,D,G,H | 0.26 ± 0.01 |
The duration of the scenario was 300 seconds. Values are the mean ± standard deviation. A lack of statistical significance (p > 0.01) in comparison with a specific model is indicated by the identifying letter (A) – (I). +p < 0.01 against the respective parameter in the scenarios with a standard sequence of actions.