Improvement of the polarized neutron interferometer setup demonstrating violation of a Bell-like inequality.

H Geppert1, T Denkmayr1, S Sponar1, H Lemmel2, Y Hasegawa1.   

Abstract

For precise measurements with polarised neutrons high efficient spin-manipulation is required. We developed several neutron optical elements suitable for a new sophisticated setup, i.e., DC spin-turners and Larmor-accelerators which diminish thermal disturbances and depolarisation considerably. The gain in performance is exploited demonstrating violation of a Bell-like inequality for a spin-path entangled single-neutron state. The obtained value of [Formula: see text], which is much higher than previous measurements by neutron interferometry, is [Formula: see text] above the limit of S=2 predicted by contextual hidden variable theories. The new setup is more flexible referring to state preparation and analysis, therefore new, more precise measurements can be carried out.

Entities:  

Keywords:  Bell inequality; Neutron interferometry; Polarised neutrons; Quantum physics

Year:  2014        PMID: 26089579      PMCID: PMC4375610          DOI: 10.1016/j.nima.2014.06.080

Source DB:  PubMed          Journal:  Nucl Instrum Methods Phys Res A        ISSN: 0168-9002            Impact factor:   1.455


Introduction

Perfect crystal neutron interferometry was first demonstrated in 1974 at the 250 kW Triga MARK-II reactor in Vienna [1]. Ever since neutron optical experiments, based on interference of matter waves, have provided a power full means of demonstrating effects related to fundamental aspects of quantum physics [2], [3], [4], such as measuring the 4π-periodicity of fermions [5], gravitational effects on the neutron [6], spin superposition [7], [8] and topological phases [9], [10], [11]. Entanglement between different degrees of freedom like the neutron׳s spin, energy and path have been accomplished [12] and used for testing Bell׳s inequality [13], [14] or measuring the influence of geometric phases [15]. Such an entanglement is achieved within single particles. Further demonstrations of the contextual nature of quantum mechanics (QM) have been performed successfully using neutron interferometry [16], [17], [18]. The violation of the Bell inequality can only be shown with high interference contrast and high spin polarisation. In the first experiment [13] a Mu-metal sheet was used as a spin turner, which induced dephasing due to small angle scattering and thereby reducing the interference contrast. The next setup [14] solved the problem of dephasing but the degree of polarisation became problematic. In this paper we report a significantly improved experimental setup. We designed new DC spin-turners and Larmor-accelerators which allow for very high contrast of the interference fringes and high temperature stability during long measurements. Temperature fluctuations below 0.1 °C over several days are achieved. They also enable high degrees of polarisation and high efficiency spin manipulation. This setup allows a large variety of state preparations and therefore provides capability for many future experiments [18], [19]. We performed a test of Bell׳s inequality using this new setup. The results reveal the substantial improvements achieved by the newly designed setup.

Improvement of the polarised interferometer setup

Overview of the polarised interferometer setup

In our setup high degrees of polarisation, thermal stability, efficient spin-manipulation and spin-analysis are required. Former setups had drawbacks that degrade the quality of the measurement results. Such setups were used for spin-superposition, geometric phase and entanglement measurements [7], [15], [17]. Spin-turners, which are realized in a way so they put materials in the beam inside the interferometer (IFM) such as Mu-metal sheets [13], anodized aluminum [20] or magnetic foils [21], cause dephasing and therefore loss in contrast. To avoid depolarization the spin-turners need to provide a homogeneous magnetic field over the whole beam cross section. Devices like those used in Refs. [6], [14] would cause too much depolarization. For earlier Bell-measurement using single-neutron interferometry two different setups were realized [13], [14]. In both setups the spin manipulation in the IFM was problematic: the contrast and the degree of polarisation were reduced. These two setups are shown in Fig. 1.
Fig. 1

(a) Setup with Mu-metal inserted in the beam to turn the spin causing dephasing. (b) Setup using a Mu-metal ring to turn the spin causing depolarisation due to inhomogeneity of the field.

(a) Setup with Mu-metal inserted in the beam to turn the spin causing dephasing. (b) Setup using a Mu-metal ring to turn the spin causing depolarisation due to inhomogeneity of the field. Fig, 1(a) shows the IFM with inserted soft magnetic Mu-metal foil as a spin turner. This is achieved by a magnetic field induced into the Mu-metal by a DC-coil outside of the IFM. The Mu-metal foil considerably reduced the contrast of the IFM due to dephasing. To overcome this problem another setup was designed, which does not need any material in the neutron beam in the IFM [14], shown in Fig. 1(b). In one path of the IFM the beam passes a tube of Mu-metal which reduces the strength of the magnetic guide field and thereby inducing a relative spin rotation by different Larmor precessions in the two IFM paths. Since the guide field leaks into the cylinder at its open ends, the field homogeneity is compromised which causes depolarisation of the neutron beam. This setup also requires a spin turner in front of the IFM which additionally reduces the degree of polarisation as described below. A schematic view of the new setup is shown in Fig. 2. The beam is monochromatised to have a mean wave length of  Å by a silicon channel-cut perfect-crystal monochromator. The beam profile is set to 3× 3 mm2 by an aperture. The incoming neutron beam is polarised by two birefringent magnetic prisms which deflect beams of up- and down-spin neutrons in different directions. The angle between these two beams is . Since the acceptance width of the interferometer crystal for Laue diffraction is even smaller, we can select one of the spin components (spin-up) by adjusting the rotation angle of the IFM accordingly. Neutrons with spin-down pass the IFM without being reflected and are blocked by a beam stopper afterwards. To avoid depolarisation of the beam a guide field is applied over the entire setup. In front of the IFM the spin is rotated by a DC spin-turner into the xy-plane. Within this plane we can adjust the spin by utilizing Larmor precession without putting any material into the beam. This is important to avoid loss of interference contrast due to dephasing. A sapphire phase shifter of 5 mm thickness between second and third plates of the IFM tunes the relative phase χ between the beams in path I and path II. Behind the IFM the spin analysis is carried out using a DC-coil on a translation stage together with a Co–Ti super-mirror array. The neutrons are detected in 3He counters with more than 99% efficiency [22].
Fig. 2

Experimental setup for measuring Bell inequalities using a triple Laue interferometer. Magnetic prisms are used to polarise the incoming beam. To avoid depolarisation a magnetic guide field B is applied around the hole setup. A spin turner before the IFM rotates the spin into the xy-plane. The first plate of the interferometer splits the beam. In each path a Larmor accelerator turns the spin by . With a phase shifter the relative phase χ can be tuned. The two exit beams are monitored by the O- and H-detectors. The beam arriving at the O-detector is filtered by a spin analyser.

Experimental setup for measuring Bell inequalities using a triple Laue interferometer. Magnetic prisms are used to polarise the incoming beam. To avoid depolarisation a magnetic guide field B is applied around the hole setup. A spin turner before the IFM rotates the spin into the xy-plane. The first plate of the interferometer splits the beam. In each path a Larmor accelerator turns the spin by . With a phase shifter the relative phase χ can be tuned. The two exit beams are monitored by the O- and H-detectors. The beam arriving at the O-detector is filtered by a spin analyser.

π/2-Spin turner

The π/2-spin turner is placed between the magnetic prisms and the IFM. Due to the small separation of spin-up and spin-down beams by the magnetic prisms and the fact that the selection of the peak takes place at the first plate of the IFM, wider peaks of the IMF׳s rocking curves degrade the degree of polarisation of the neutron beam. The peak width at the first IFM plate is determined by the monochromator and the properties of the π/2-spin turner regarding small-angle scattering. In contrast to earlier experiments where a single reflecting monochromator was used, we used a three-fold channel-cut monochromator. A comparison of rocking curves by using one- and three-fold reflections is shown in Fig. 3.
Fig. 3

Rocking curves for one- and three-fold reflecting monochromator crystals. The FWHM׳s are  rad for the 1-fold and rad 3-fold monochromator crystals.

Rocking curves for one- and three-fold reflecting monochromator crystals. The FWHM׳s are  rad for the 1-fold and rad 3-fold monochromator crystals. The full width at half maximum σ (FWHM) of the rocking peak of the single reflecting crystal has a , whereas the triple reflecting crystal has a : the former is 43% wider than the triple reflecting crystal. The peak broadening was measured for different coils made of copper ribbon (0.1× 3 mm2 and 0.1× 4 mm2 in profile), aluminium ribbon (1× 4 mm2 in profile), and aluminium wire (0.5 mm in diameter). In Fig. 4 the aluminium ribbon coil and the copper ribbon coil are shown next to an empty coil frame. Fig. 5 shows the rocking curves for different coils in comparison to the empty coil frame. In Table 1 the peak height and the width of the rocking curves for different coils are given with respect to the empty beam line. One can see that despite the small absorption and the scattering cross-section of aluminium, the wire coil enlarges the width of peak and lowers the peak intensity because of small-angular scattering, whereas within the error there is no evident difference between the three ribbon coils. The rectangular profiled ribbons do not produce significant small-angle scattering and therefore more neutrons fulfil Bragg׳s law. For our measurements we used 3 mm wide copper ribbon for both DC spin-turners. In Fig. 6 one can see the separation of the up-spin peak and the down-spin peak with the π/2-spin turner in the beam line. Between the two main peaks one can see a small peak produced by neutrons that fulfil higher orders of the Bragg condition. The degree of polarisation and the efficiencies of the two DC spin turners can be evaluated using the two-flipper method corresponding to Ref. [23]. Two spin-flippers, which rotate the spin by π when turned on, are necessary, in our case the DC-spin turner in front and behind the IFM are used. Assuming that the polariser and the analyser have the same efficiency, which is justified in our setup, it is possible to calculate the efficiencies of the two spin-flippers (DC 1 and DC 2) and the degree of polarization using
Fig. 4

Photographs of -spin turner coils made out of aluminium, copper and an empty frame for a coil (right to left).

Fig. 5

Rocking curves without coil, with the copper ribbon coil, the aluminium ribbon coil and the aluminium wire coil inserted in the neutron beam.

Table 1

Rocking curve comparison with spin turning coils made of different materials, normalised to the empty setup.

Material of the coilPeakFWHM
No coil1.0001.000
Al wire0.56(1)1.68(4)
Al ribbon0.80(1)1.16(2)
Cu ribbon 3 mm width0.84(1)1.11(2)
Cu ribbon 4 mm width0.85(1)1.16(2)
Fig. 6

Rocking curve showing the up-spin and the down-spin peak produced by the birefringent magnetic prisms.

Photographs of -spin turner coils made out of aluminium, copper and an empty frame for a coil (right to left). Rocking curves without coil, with the copper ribbon coil, the aluminium ribbon coil and the aluminium wire coil inserted in the neutron beam. Rocking curve comparison with spin turning coils made of different materials, normalised to the empty setup. Rocking curve showing the up-spin and the down-spin peak produced by the birefringent magnetic prisms. Therefor four intensities are measured, first no spin-flipper turned on , one of the spin-flippers turned on and , and both turned on . We measured the degree of polarisation to be , the efficiency of the first DC-coil to be and efficiency of the second coil to be . The efficiency of the π/2-spin turners is reduced by the stray fields of the magnetic prisms in front of the IFM and of the super-mirror spin analyser behind it. Therefore the coils were placed as far from super mirror and magnetic prisms as possible.

Larmor-spin rotator

The state preparation requires two Larmor accelerator coils placed in the IFM as shown in Fig. 2. These coils in Helmholtz geometry apply an additional parallel or an anti-parallel field to the guide field in z-direction locally and thereby changing the spin precession in the xy-plane. Since the rotation angle is given by , where μ is the magnetic moment of the neutron, l is the length of the coils and v is the velocity of the neutrons, a magnetic field of about 0.33 mT is required for a spin rotation of π/2. For the fabricated coil a current of about 0.7 A is required for a spin rotation of π/2. Since the coils produce heat and due to the high sensitivity of the IFM to thermal influences, the coils need to be cooled down. The coils are placed in small boxes which are completely flooded with temperature controlled water. The coil wire is in direct contact with the water, insulated only by lacquer. The boxes are made of acrylic glass, which is a thermal and electrical insulator. Length, width and height of the boxes amount to 22, 26 and 26 mm respectively. Fig. 7(a) shows a schematic view of a box. The boxes have a straight passage for the neutron beam (green arrows), so that the beam does not pass any material and therefore no dephasing occurs. The figure also shows the magnetic field in z-direction (red arrow) applied by the coils in Helmholtz configuration (yellow). The flux of the cooling water is depicted by blue arrows. Fig. 7(b) and (c) shows the box without top and finished with the connectors for the water cooling respectively. In Fig. 7(d) one can see the boxes placed in the IFM. In this picture the mountings for beam stoppers between second and third plates of the IFM are depicted. The beam stoppers are used to calibrate the Larmor accelerators one at a time by blocking the other path. The beam stoppers used here are 1 mm thick cadmium plates.
Fig. 7

(a) Schematic view of the Larmor accelerator boxes with coils in Helmholtz geometry (yellow), magnetic field (red), neutron beam (green) and water flux (blue), (b) Larmor acceleration coil with Helmholtz geometry in box, (c) closed box with connectors for water cooling, and (d) boxes in the interferometer with connected cooling system and absorber holder for adjustment of the Larmor accelerators. (For interpretation of the references to colour in this figure caption, the reader is referred to the web version of this paper.)

(a) Schematic view of the Larmor accelerator boxes with coils in Helmholtz geometry (yellow), magnetic field (red), neutron beam (green) and water flux (blue), (b) Larmor acceleration coil with Helmholtz geometry in box, (c) closed box with connectors for water cooling, and (d) boxes in the interferometer with connected cooling system and absorber holder for adjustment of the Larmor accelerators. (For interpretation of the references to colour in this figure caption, the reader is referred to the web version of this paper.)

Violation of Bell-like inequalities

Quantum mechanics (QM) is one of the most successful physical theories and its predictions have been proven accurately in many experiments using various kinds of systems. Einstein, Podolsky and Rosen (EPR) [24] argued that QM is not a complete theory since it only gives probabilistic predictions and that there must be an underlying deterministic theory to QM. In 1964 Bell [25] showed that local hidden variable theories satisfy some inequalities that are violated by QM. Shortly after Bell published his well known paper Clauser, Horne, Shimony, and Holt (CSHS) reformulated Bell׳s inequalities suitable for the first experimental test of quantum non-locality [26], [27]. In the case of neutrons not two particles are entangled but two different degrees of freedom within one particle [28], [29]. Non-contextual hidden variables theories (NCHVTs) state that the outcome of a measurement is independent of previous or simultaneous measurements on any set of commuting observables.

Theory

In our single neutron interferometer we entangle two different degrees of freedom (spatial and spin) of a single neutron [30]. The neutron is described by a tensor product Hilbert space , where H corresponds to the spatial wave function and H to the spinor wave function. Since observables of the spacial part commute with those of the spinor part one can derive a Bell-like state. The normalized wave function is given bywhere and correspond to the up- and the down-spin and and represent the two paths in the IFM. The expectation value of the joint spin and the path measurement can be written asThe observables for spin and path can be decomposed by projection operators and , which project onto orthogonal spin states and orthogonal path states , respectively. The expectation value given in Eq. (3) and the projection operators correspond to , and in the conventional EPR argument [31]. In the experiment the parameter α can be varied by polarisation measurement of the Bell-like state, χ is tuned by an auxiliary phase shifter. For single-neutron interferometry a Bell-like inequality can be expressed using the expectation values as , withIn our experiment the expectation values are determined by a combination of count rates of a single detector donated to appropriated settings of α and χ. This givesThe count rates are given by according to quantum mechanical predictions. This leads to a sinusoidal behaviour of the expectation values . Bell׳s inequality are violated for various sets of polarisation analyses and phase shifts , but the largest violation is expected for , , and , which gives the value . By measuring S one can test whether or not NCHVTs can describe nature correctly.

Measurement

For optimisation of a measurement every part of the setup should be adjusted in a systematic way. First of all the point of highest interference contrast on the IFM, i.e. the sweet spot is looked for. The interferometer crystal is mapped by moving an aperture vertically and horizontally in the xz-plane in front of the IFM and measuring the contrast on each position. The result of such a raster scan is shown in Fig. 8. This scan is performed with a beam cross-section of 3× 3 mm2 and a step increment of 1 mm. One can see that only a small part of the IFM provides highest contrast, which reaches at C=0.82.
Fig. 8

Raster scan showing the contrast of the IFM with respect to position in xz-plane.

Raster scan showing the contrast of the IFM with respect to position in xz-plane. The contrast of the IFM is very sensitive to temperature fluctuations. Even temperature changes by 0.1 °C disturb the interference pattern. Since the guide field and the Larmor-accelerators produce heat, both elements are water cooled. This is done by two temperature stabilised water pumps. To optimise the temperature of the cooling water for the guide field and the boxes, temperature scans are performed. In Fig. 9 the contrast of the IFM is plotted for different temperatures of the cooling water in the boxes. For 25.2 °C an average contrast of is achieved. After stabilisation contrast up to C=0.91 can be observed as seen in Fig. 10. When the temperature is raised up to 26.8 °C the contrast drops to . A raise by 1 °C in temperature results in a decrease in contrast of C=0.60.
Fig. 9

The contrast of the interferometer as a function of temperature. At 25.2 °C a contrast can be achieved, at 1 °C higher temperature only can be reached.

Fig. 10

Interference pattern with a contrast of 0.91.

The contrast of the interferometer as a function of temperature. At 25.2 °C a contrast can be achieved, at 1 °C higher temperature only can be reached. Interference pattern with a contrast of 0.91. Thermal stability is important for another crucial point, since not only loss in contrast but also considerable phase drifts occur. A temperature change of 1 °C in the boxes results in 1.92 rad phase shift. Fig. 11 shows this situation. The large error bars at high temperatures arises from the low contrast obtained at this temperatures as seen in Fig. 9. Both the loss in contrast as well as the phase shift resulting from thermal instability degrade the quality of measurement results. To increase the stability of the setup the IFM and the Larmor accelerators are placed in a box to avoid air convection and therefore temperature fluctuations over long periods of time. We measured temperature fluctuations less than 0.1 °C over several days. The temperature stability is limited by the thermal stabilisation at the beam line׳s housing, which is passive.
Fig. 11

The phase shift induced by temperature change of the Larmor accelerators.

The phase shift induced by temperature change of the Larmor accelerators. To determine the expectation values occurring in the Bell-inequalities four spin-directions , , , and four phase-shifts , , , need to be measured. The spin-directions are selected by the DC spin-turner behind the IFM, while the phase shifts are tuned by the sapphire phase shifter inside the IFM. A set of this measurements is shown in Fig. 12. The data is fitted to a sinusoidal function using the least squares fit method. The error given in the final S-value results from statistical fluctuations in count rates. Systematic errors due to imperfect spin manipulation, finite contrast of the IFM and phase instabilities during the measurement lead to a smaller value for S than the theoretical predicted value . We repeated the measurements twice, in order to reduce statistical errors. By evaluating the Bell inequality the S-value in equality (4) is calculated to bewhich implies a violation by more than 28σ.
Fig. 12

Intensity oscillations obtained by tuning the phase shift χ for four different spin analyses α and C, give the measured contrast.

Intensity oscillations obtained by tuning the phase shift χ for four different spin analyses α and C, give the measured contrast.

Discussion and conclusion

We have presented a new design of the neutron-interferometer setup and new devices for spin manipulation, which considerably improve the abilities of the polarised neutron IFM. Using a new coil design for the DC spin-turners and a channel-cut monochromator the degree of polarisation of the incoming beam of is achieved. New Larmor accelerators allow the reduction of thermal disturbances on the IFM and dephasing since no material is put in the beam path inside the IFM. This enables high contrasts up to C=0.91. The newly designed spin manipulators allow easy and precise manipulation of the neutron׳s spin and enable various applications for future experiments. The temperature was stable within 0.1 °C over several days. A sufficient active thermal stabilization is required to improve the setup further on. An upgrade of the beam-line housing including active thermal stabilisation is planed. With this setup we obtained the value of for Bell-like inequality measurements, which is above the boarder of 2 and as a consequence disproves NCHVT clearly. This result is not only clearly better than previous measurements [13], [14] but it was also achieved using less repetitions and therefore less beam time. This was feasible through higher stability of the setup and better manipulations of the neutron-spin. Also previous setups were designed solely for measuring Bell-like inequalities. The new setup is much more applicable for other experiments such as measuring a quantum cheshire cat [18] or performing weak measurements [19].
  7 in total

1.  Violation of a Bell-like inequality in single-neutron interferometry.

Authors:  Yuji Hasegawa; Rudolf Loidl; Gerald Badurek; Matthias Baron; Helmut Rauch
Journal:  Nature       Date:  2003-09-04       Impact factor: 49.962

2.  Observation of the topological Aharonov-Casher phase shift by neutron interferometry.

Authors: 
Journal:  Phys Rev Lett       Date:  1989-07-24       Impact factor: 9.161

3.  Simple unified form for the major no-hidden-variables theorems.

Authors: 
Journal:  Phys Rev Lett       Date:  1990-12-31       Impact factor: 9.161

4.  Quantum contextuality in a single-neutron optical experiment.

Authors:  Yuji Hasegawa; Rudolf Loidl; Gerald Badurek; Matthias Baron; Helmut Rauch
Journal:  Phys Rev Lett       Date:  2006-12-06       Impact factor: 9.161

5.  Experimental test of quantum contextuality in neutron interferometry.

Authors:  H Bartosik; J Klepp; C Schmitzer; S Sponar; A Cabello; H Rauch; Y Hasegawa
Journal:  Phys Rev Lett       Date:  2009-07-23       Impact factor: 9.161

6.  Geometric phase in coupled neutron interference loops.

Authors: 
Journal:  Phys Rev A       Date:  1996-04       Impact factor: 3.140

7.  Observation of a quantum Cheshire Cat in a matter-wave interferometer experiment.

Authors:  Tobias Denkmayr; Hermann Geppert; Stephan Sponar; Hartmut Lemmel; Alexandre Matzkin; Jeff Tollaksen; Yuji Hasegawa
Journal:  Nat Commun       Date:  2014-07-29       Impact factor: 14.919