A Huda1, W J Garzón2, G C L Filho3, B Vieira3, R Kramer2, X G Xu4, Y Gao4, H J Khoury5.
Abstract
The use of endoscopic retrograde cholangiopancreatography (ERCP) in pregnant patients is not rare. Most studies on the safety and efficacy of these procedures report short- and long-term pregnancy outcomes and but not foetal absorbed doses. This investigation reports on an ERCP procedure for a 40-y-old woman who was 32-34 weeks pregnant. Thermoluminescent dosemeters (TLD 100) were used to measure doses received by the patient and the staff. Additionally, Monte Carlo calculations were performed using a 3D computational phantom representing a 9-month pregnant patient to estimate the foetal absorbed dose. The results show that the spleen of the mother received the largest absorbed dose of 12.18 mGy since it was closer to the source than other internal organs. For the foetus and uterus, the lowest absorbed dose was found to be 0.01 mGy to the foetal brain, while the largest absorbed dose was estimated to be 0.13 mGy to the placenta.Entities:
Mesh:
Year: 2015 PMID: 26084305 DOI: 10.1093/rpd/ncv354