Abstract

PURPOSE: To establish and validate a rapid, cost-effective and accurate screening assay for the simultaneous testing of human naturally occurring anti-cytokine autoantibodies (c-aAb) targeting interleukin-1α (IL-1α), interleukin-6 (IL-6), interleukin-10 (IL-10), granulocyte-macrophage colony-stimulating factor (GM-CSF), and interferon α (IFNα). Because the c-aAbs can be transferred to patients through blood transfusion, the assay was used to assess c-aAb levels in a cohort of patients who were receiving blood transfusions and subsequently presented with or without febrile reactions.
MATERIALS AND METHODS: The microsphere-based Luminex platform was used. Recombinant forms of human IL-1α, IL-6, IL-10, GM-CSF, and IFNα were gently coupled to MAG-PLEX beads. Plasma IgG binding was measured with phycoerythrin (PE)-labeled secondary antibodies. Previously confirmed c-aAb positive and negative donor plasma samples and pooled normal immunoglobulin preparations were used to validate the assay. Plasma samples from 98 transfusion recipients, half of whom presented with febrile reactions, were tested by the assay.
RESULTS: The assay detected specific and saturable immunoglobulin G (IgG) binding to each of the tested cytokines in previously confirmed c-aAb positive plasmas and in preparations of pooled normal immunoglobulin. Confirmed c-aAb negative plasmas gave no saturable binding. The detection limit of the cytokine autoantibodies was estimated to be between 1 pM and 10 pM. The recovery of confirmed cytokine autoantibodies quantities in the negative plasma samples ranged between 80% and 125%. The analytical intra- and inter-assay variations were 4% and 11%, respectively. Varying c-aAb levels were detectable in the transfusion recipients. There was no difference in c-aAb frequency between the patients with or without febrile transfusion reactions. The c-aAb level before and after the blood transfusions varied only slightly and in an irregular manner.
CONCLUSION: This assay simultaneously detected up to five different c-aAbs in pooled human IgG and in plasma from individual blood donors, and it was deemed suitable for larger screenings. Based on confirmed antibody binding characteristics and the resultant reactivity in this multiplex assay, a classification of the c-aAb levels was suggested. The screening results of the recipients who received blood transfusions indicate that more studies are needed to clarify the role of antibodies, if any, in transfusion medicine and in high-dose immunoglobulin treatment.
Copyright © 2015 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Antibody detection; Autoantibodies; Blood transfusion; Cytokine; Multiplex assay

Mesh:

Substances:

Year:  2015        PMID: 26080061     DOI: 10.1016/j.jim.2015.06.010

Source DB:  PubMed          Journal:  J Immunol Methods        ISSN: 0022-1759            Impact factor:   2.303


  5 in total

1.  IL-6 Autoantibodies Predict Lower Platelet Counts and Altered Plasma Cytokine Profiles in Healthy Blood Donors: Results From the Danish Blood Donor Study.

Authors:  Jakob Hjorth von Stemann; Ole Birger Vesterager Pedersen; Henrik Hjalgrim; Christian Erikstrup; Henrik Ullum; Joseph Dowsett; Lise Wegner Thørner; Margit Anita Hørup Larsen; Erik Sørensen; Morten Bagge Hansen; Sisse Rye Ostrowski
Journal:  Front Med (Lausanne)       Date:  2022-06-24

2.  Prevalence and correlation of cytokine-specific autoantibodies with epidemiological factors and C-reactive protein in 8,972 healthy individuals: Results from the Danish Blood Donor Study.

Authors:  Jakob Hjorth von Stemann; Andreas Stribolt Rigas; Lise Wegner Thørner; Daniel Guldager Kring Rasmussen; Ole Birger Pedersen; Klaus Rostgaard; Christian Erikstrup; Henrik Ullum; Morten Bagge Hansen
Journal:  PLoS One       Date:  2017-06-30       Impact factor: 3.240

3.  Cytokine autoantibodies are stable throughout the haematopoietic stem cell transplantation course and are associated with distinct biomarker and blood cell profiles.

Authors:  Jakob Hjorth von Stemann; Lars Klingen Gjærde; Eva Kannik Haastrup; Lia Minculescu; Patrick Terrence Brooks; Henrik Sengeløv; Morten Bagge Hansen; Sisse Rye Ostrowski
Journal:  Sci Rep       Date:  2021-12-14       Impact factor: 4.379

4.  Cytokine Autoantibodies Are Associated with Infection Risk and Self-Perceived Health: Results from the Danish Blood Donor Study.

Authors:  Jakob H von Stemann; Ole B Pedersen; Henrik Hjalgrim; Christian Erikstrup; Henrik Ullum; Lise W Thørner; Margit Ah Larsen; Kristoffer S Burgdorf; Erik Sørensen; Morten B Hansen; Sisse R Ostrowski
Journal:  J Clin Immunol       Date:  2020-01-15       Impact factor: 8.317

5.  Chronic inflammation markers and cytokine-specific autoantibodies in Danish blood donors with restless legs syndrome.

Authors:  Joseph Dowsett; Maria Didriksen; Jakob Hjorth von Stemann; Margit Hørup Larsen; Lise Wegner Thørner; Erik Sørensen; Christian Erikstrup; Ole Birger Pedersen; Morten Bagge Hansen; Jesper Eugen-Olsen; Karina Banasik; Sisse Rye Ostrowski
Journal:  Sci Rep       Date:  2022-01-31       Impact factor: 4.379

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.