Comparison of four commercial viral load techniques in an area of non-B HIV-1 subtypes circulation.

Ousseynou Ndiaye1, Halimatou Diop-Ndiaye2, Abdou Salam Ouedraogo3, Fatim Zahra Fall-Malick1, Amina Sow-Sall1, Moussa Thiam1, Abou Abadallah Malick Diouara1, Cheikh Tidiane Ndour1, Aïssatou Gaye-Diallo1, Souleymane Mboup1, Coumba Toure-Kane1.   

Abstract

The aim of this study was to compare four HIV-1 viral quantitation platforms, Nuclisens EasyQ v2.0(®) (EQ), COBAS AmpliPreP/Cobas Taqman(®) HIV-1 test v 2.0 (CTM), GENERIC HIV CHARGE VIRALE(®) (GEN), with Abbott Real Time HIV-1(®) (m2000sp/rt) as reference technique. The study had first evaluated m2000sp/rt performances and then compared quantitation between techniques. Discordant samples were genotyped on gag and pol gene and sequences were analyzed using Sequence locator and SeqPublish to detect eventual mismatches. Performance analysis of m2000sp/rt showed good results with coefficients of variation values (CV) of 1.35%, 0.65%, and 0.54% for repeatability testing of low, intermediate and high concentrations, respectively. Reproducibility tests showed low CV values with 2.36% and 1.42% for low and high concentration levels, respectively and contamination test was very low value with 0.94%. Correlation and concordance between techniques ranged from r(2)=0.98 and bias=-0.00185 (for m2000sp/rt vs CTM) to r(2)=0.90 and bias=-0.135 (for EQ vs GEN). Discrepancies were observed on 37 samples mostly CRF02_AG but despite some mismatches, sequence analysis (26/37) did not show any remarkable differences between CRF02_AG queries and references. This study showed good correlation and good concordance between techniques. However, EQ yielded under-quantitation of CRF02_AG.
Copyright © 2015 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Comparison; HIV-1; Viral load

Mesh:

Substances:

Year:  2015        PMID: 26068392     DOI: 10.1016/j.jviromet.2015.06.002

Source DB:  PubMed          Journal:  J Virol Methods        ISSN: 0166-0934            Impact factor:   2.014


  4 in total

1.  Evaluation of the RealTime HIV-1, Xpert HIV-1, and Aptima HIV-1 Quant Dx Assays in Comparison to the NucliSens EasyQ HIV-1 v2.0 Assay for Quantification of HIV-1 Viral Load.

Authors:  Orna Mor; Yael Gozlan; Marina Wax; Fernando Mileguir; Avia Rakovsky; Bina Noy; Ella Mendelson; Itzchak Levy
Journal:  J Clin Microbiol       Date:  2015-08-19       Impact factor: 5.948

2.  Field evaluation of an open and polyvalent universal HIV-1/SIVcpz/SIVgor quantitative RT-PCR assay for HIV-1 viral load monitoring in comparison to Abbott RealTime HIV-1 in Cameroon.

Authors:  Emilande Guichet; Avelin Aghokeng; Sabrina Eymard-Duvernay; Nicole Vidal; Ahidjo Ayouba; Eitel Mpoudi Ngole; Eric Delaporte; Laura Ciaffi; Martine Peeters
Journal:  J Virol Methods       Date:  2016-09-05       Impact factor: 2.014

3.  Evaluation of Hologic Aptima HIV-1 Quant Dx Assay on the Panther System on HIV Subtypes.

Authors:  Mark M Manak; Holly R Hack; Sangeetha V Nair; Andrew Worlock; Jennifer A Malia; Sheila A Peel; Linda L Jagodzinski
Journal:  J Clin Microbiol       Date:  2016-08-10       Impact factor: 5.948

4.  Analytical Performances of Human Immunodeficiency Virus Type 1 RNA-Based Amplix® Real-Time PCR Platform for HIV-1 RNA Quantification.

Authors:  Christian Diamant Mossoro-Kpinde; Ralph-Sydney Mboumba Bouassa; Mohammad-Ali Jenabian; Serge Tonen Wolyec; Leman Robin; Mathieu Matta; Jean de Dieu Longo; Gérard Grésenguet; Laurent Andreoletti; Laurent Bélec
Journal:  AIDS Res Treat       Date:  2016-12-05
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.