Literature DB >> 26058389

Implantation of an Iliac Branch Device After EVAR via a Femoral Approach Using a Steerable Sheath.

Alexander Oberhuber1, Mansur Duran2, Neslihan Ertaş2, Florian Simon2, Hubert Schelzig2.   

Abstract

PURPOSE: To describe a contralateral femoral approach for iliac branch device implantation using a steerable sheath in the setting of an existing bifurcated stent-graft. TECHNIQUE: The method is demonstrated in an 80-year-old man who developed a 4-cm iliac aneurysm 3 years after implantation of an Endurant bifurcated stent-graft. Both femoral arteries were cannulated after surgical cutdown. The steerable sheath was advanced from the contralateral side over the neobifurcation of the bifurcated stent-graft. A 0.014-inch Roadrunner wire was used as a through-and-through wire to stabilize the curve of the sheath and to get proper push. The bridging stent-graft for the iliac branch was advanced over this sheath to seal the iliac aneurysm. During the entire procedure, the sheath was stable over the neobifurcation without pulling it down.
CONCLUSION: The contralateral femoral approach for iliac branch graft implantation is feasible in cases with an extant bifurcated stent-graft using a steerable sheath and a through-and-through wire.
© The Author(s) 2015.

Entities:  

Keywords:  abdominal aortic aneurysm; endovascular aneurysm repair; femoral approach; iliac aneurysm; iliac branch; steerable sheath; stent-graft

Mesh:

Year:  2015        PMID: 26058389     DOI: 10.1177/1526602815590972

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  3 in total

1.  A novel all-retrograde approach for t-Branch implantation in ruptured thoracoabdominal aneurysm.

Authors:  A Claire Watkins; Alla Avramenko; Raphael Soler; Dominique Fabre; Stephan Haulon
Journal:  J Vasc Surg Cases Innov Tech       Date:  2018-12-04

2.  Endovascular repair of a dissecting pararenal abdominal aortic aneurysm in a patient with type III Loeys-Dietz syndrome.

Authors:  Kirthi S Bellamkonda; Alan Dardik; Naiem Nassiri
Journal:  J Vasc Surg Cases Innov Tech       Date:  2020-10-26

3.  Steerable sheath for exclusively femoral bilateral extension of previous fenestrated endovascular aneurysm repair with iliac branch devices.

Authors:  Roberta Vaccarino; Angelos Karelis; Björn Sonesson; Nuno V Dias
Journal:  J Vasc Surg Cases Innov Tech       Date:  2021-04-24
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.