A retrospective dosimetric comparison of TG43 and a commercially available MBDCA for an APBI brachytherapy patient cohort.

K Zourari1, T Major2, A Herein2, V Peppa1, C Polgár2, P Papagiannis3.   

Abstract

PURPOSE: To compare dosimetry using a contemporary model based dose calculation algorithm (MBDCA) following TG186 recommendations, and the conventional TG43 method in an (192)Ir high dose rate (HDR) accelerated partial breast irradiation (APBI) patient cohort.
METHODS: Data of 38 APBI patients were studied. Dosimetry for the treatment plans was performed using both the TG43 and TG186 dose calculation methods of the Oncentra Brachy v4.4 treatment planning system (TPS). Analysis included indices of clinical interest for the planning target volume (PTV coverage, dose homogeneity, conformity) as well as dose volume histograms (DVH) for the breast, lung, heart, rib and skin. Significance testing of observed differences between TG43 and TG186 results was carried out and the effect of target location to these differences was studied.
RESULTS: Statistically significant differences were observed in the values of clinically relevant DVH parameters for the PTV and the organs at risk (OAR), except for the heart. Differences for the PTV are relatively small (<1% for coverage, on the order of 2% for homogeneity and conformity) with a slight TG43 overestimation except for the dose homogeneity. Percentage differences are larger for the rib and lung (on the order of 4% for Dmax and 5% for V10Gy, respectively) and maximum for the skin (on the order of 6% for D10cc), with a correlation of the observed differences with target location.
CONCLUSION: While the MBDCA option of the TPS appears to improve dosimetric accuracy, differences from TG43 do not appear to warrant dose prescription changes or treatment protocol amendment..
Copyright © 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Accelerated partial breast irradiation; DVH analysis; Dose homogeneity and conformity indices; Dosimetry; Model-based dose calculation algorithms; Treatment planning

Mesh:

Year:  2015        PMID: 26055951     DOI: 10.1016/j.ejmp.2015.05.010

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  8 in total

Review 1.  A review of dosimetric impact of implementation of model-based dose calculation algorithms (MBDCAs) for HDR brachytherapy.

Authors:  Yousif A M Yousif; Alexander F I Osman; Mohammed A Halato
Journal:  Phys Eng Sci Med       Date:  2021-06-17

Review 2.  A brief look at model-based dose calculation principles, practicalities, and promise.

Authors:  Ron S Sloboda; Hali Morrison; Brie Cawston-Grant; Geetha V Menon
Journal:  J Contemp Brachytherapy       Date:  2017-02-08

3.  Individualized mould-based high-dose-rate brachytherapy for perinasal skin tumors: technique evaluation from a dosimetric point of view.

Authors:  Christian Scherf; Jörg Licher; Christina Mletzko; Martin Trommel; Nikolaos Tselis; Georgios Chatzikonstantinou; Markus Diefenhardt; Claus Rödel; Janett Köhn; Ulla Ramm
Journal:  J Contemp Brachytherapy       Date:  2021-04-14

4.  Impact of heterogeneity-corrected dose calculation using a grid-based Boltzmann solver on breast and cervix cancer brachytherapy.

Authors:  Julia Hofbauer; Christian Kirisits; Alexandra Resch; Yingjie Xu; Alina Sturdza; Richard Pötter; Nicole Nesvacil
Journal:  J Contemp Brachytherapy       Date:  2016-04-19

5.  Impact of a commercially available model-based dose calculation algorithm on treatment planning of high-dose-rate brachytherapy in patients with cervical cancer.

Authors:  Kota Abe; Noriyuki Kadoya; Shinya Sato; Shimpei Hashimoto; Yujiro Nakajima; Yuya Miyasaka; Kengo Ito; Rei Umezawa; Takaya Yamamoto; Noriyoshi Takahashi; Ken Takeda; Keiichi Jingu
Journal:  J Radiat Res       Date:  2018-03-01       Impact factor: 2.724

6.  A comparative assessment of inhomogeneity and finite patient dimension effects in 60Co and 192Ir high-dose-rate brachytherapy.

Authors:  Irina Fotina; Kyveli Zourari; Vasileios Lahanas; Evaggelos Pantelis; Panagiotis Papagiannis
Journal:  J Contemp Brachytherapy       Date:  2018-02-28

7.  The dosimetric impact of replacing the TG-43 algorithm by model based dose calculation for liver brachytherapy.

Authors:  Anna Sophie Duque; Stefanie Corradini; Florian Kamp; Max Seidensticker; Florian Streitparth; Christopher Kurz; Franziska Walter; Katia Parodi; Frank Verhaegen; Jens Ricke; Claus Belka; Gabriel Paiva Fonseca; Guillaume Landry
Journal:  Radiat Oncol       Date:  2020-03-09       Impact factor: 3.481

8.  Pre-clinical dosimetry of a new six-channel applicator for high-dose-rate treatment of esophageal cancer.

Authors:  Anzi Zhao; Shengqiang Gao; John Greskovich; Douglas Allan Wilkinson
Journal:  J Contemp Brachytherapy       Date:  2021-05-12
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.