Literature DB >> 26044897

Effects of prolonged ethanol lock exposure to carbothane- and silicone-based hemodialysis catheters: a 26-week study.

Daniel L Landry1, Randa A Jaber, Nandheesha Hanumanthappa, George S Lipkowitz, Michael H O'Shea, Harry Bermudez, Adam P Hathorne, Gregory L Braden.   

Abstract

PURPOSE: Antibiotic locks in catheter-dependent chronic hemodialysis patients reduce the rate of catheter-related bloodstream infections (CRBSIs), but may be associated with the development of resistant bacteria. Ethanol-based catheter locks may provide a better alternative; however, there are limited data on the long-term integrity of dialysis catheters exposed to ethanol.
METHODS: We performed in vitro testing of two types of hemodialysis catheters—silicone (SLC) and carbothane (CBT) based—with a 70% ethanol lock (EL) versus heparin lock (HL) for 26 weeks. Lock solutions were changed thrice weekly to mimic a conventional hemodialysis schedule. We tested mechanical properties of the catheters at 0, 13 and 26 weeks by examining stress/strain relationships (SS400%) and modulus of elasticity (ME). Electron microscopy was performed to examine catheter ultrastructure at 0 and 26 weeks.
RESULTS: Catheter integrity for HL versus EL in SLC (SS400%: 4.5 vs. 4.5 MPa, p = NS; ME: 4.6 vs. 4.7 MPa, p = NS) or CBT-based catheters (SS400%: 7.6 vs. 8.9 MPa, p = NS; ME: 9.6 vs. 12.2 MPa, p = NS) were all similar at 13 and 26 weeks. Scanning electron microscopy revealed no structural changes in the central and luminal wall internal surfaces of EL- versus HL-treated catheters.
CONCLUSIONS: There were no significant differences in catheter integrity between SLC or CBT catheters exposed to a 70% EL for 26 weeks. Given its low cost, potential to avoid antibiotic resistance and structural integrity after 6 months of high-dose ethanol, ELs should be studied prospectively against antibiotic locks to assess the efficacy and safety in hemodialysis patients.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26044897     DOI: 10.5301/jva.5000397

Source DB:  PubMed          Journal:  J Vasc Access        ISSN: 1129-7298            Impact factor:   2.283


  3 in total

1.  A material stress test study on occurrence of leakage and material failure of peritoneal dialysis (PD) catheters.

Authors:  Matthias Klingele; Martin Carstens; Lea Baerens; Matthias W Laschke; Wolfgang Metzger; Danilo Fliser; Clemens M Meier
Journal:  Sci Rep       Date:  2021-05-13       Impact factor: 4.379

2.  Preemptive Ethanol Lock Therapy in Pediatric Hematology/Oncology Patients With Catheter-Associated Bloodstream Infection: Impact on Length of Stay, Cost, and Catheter Salvage.

Authors:  Eric McGrath; Wei Du; Madhvi Rajpurkar
Journal:  Clin Pediatr (Phila)       Date:  2017-06-30       Impact factor: 1.168

3.  Evaluation of the efficacy of an interdialytic "ethanol 40% v/v - enoxaparin 1000 U/mL" lock solution to prevent tunnelled catheter infections in chronic hemodialysis patients: a multi-centre, randomized, single blind, parallel group study.

Authors:  Julien Aniort; Aurélien Piraud; Mireille Adda; Bruno Perreira; Marc Bouiller; Jacques Fourcade; Abdallah Guerraoui; Emilie Kalbacher; Thierry Krumel; Hélène Leray Moragues; Damien Thibaudin; Carlos Gustavo Vela; Guillaume Vernin; Hugo Weclawiak; Lise Bernard; Anne Elisabeth Heng; Bertrand Souweine
Journal:  BMC Nephrol       Date:  2019-04-30       Impact factor: 2.388

  3 in total