Foveal pit morphology evaluation during optical biometry measurements using a full-eye-length swept-source OCT scan biometer prototype.

Thomas Bertelmann1, Marcus Blum2, Kathleen Kunert2, Wolfgang Haigis3, Anke Messerschmidt-Roth1, Monika Peter4, Tobias Buehren4, Walter Sekundo1.   

Abstract

PURPOSE: To assess the scan quality of foveal pit morphology (FPM) and to quantify central retinal thickness (CRT) during routine optical biometry measurements with a full-eye-length swept-source optical coherence tomography (SS-OCT) scan biometer prototype (IOLMaster700) and to compare these results with standard examinations using spectral-domain optical coherence tomography (SD-OCT) technique (Cirrus4000 SD-OCT).
METHODS: As part of a SS-OCT scan protocol to measure biometric parameters for intraocular lens power calculation, central horizontal 1 mm retinal B scans were taken from phakic (group I) and pseudophakic (group II) nonvitrectomized eyes. To evaluate FPM, macular scans of either examination technique were subjectively analyzed and compared. Repeated CRT measurements were performed to analyze repeatability and consistency of IOLMaster700 recordings. These results were compared with CRT evaluations using SD-OCT.
RESULTS: Overall, 146 eyes of 146 patients were included in this series. The subjective assessments of FPM are illustrated. Repeated CRT measurements (repeatability) with the IOLMaster700 disclosed an overall intraclass correlation of 0.57 (group I: 0.48; group II: 0.89). Overall coefficient of variation (accuracy) was calculated to be 12.43% (group I: 14.21%; group II: 5.66%). The comparison of CRT measurements between both devices showed significant differences in group I (p = 0.006).
CONCLUSIONS: Compared with SD-OCT, resolution of the 1 mm retinal B scan of SS-OCT scan biometry was lower. However, advanced pathologic characteristics were clearly discernible. Repeatability and accuracy of CRT measurements were acceptable though lower than with the standard SD-OCT technique. The CRT differed significantly in eyes of particular interest (group I) between both devices. The new scan could provide useful information for subsequent patient examination and further treatment planning for cataract surgery.

Entities:  

Mesh:

Year:  2015        PMID: 26044376     DOI: 10.5301/ejo.5000630

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   2.597


  2 in total

1.  Swept-Source Optical Coherence Tomography Biometer as Screening Strategy for Macular Disease in Patients Scheduled for Cataract Surgery.

Authors:  Daniele Tognetto; Marco R Pastore; Chiara De Giacinto; Riccardo Merli; Marco Franzon; Rossella D'Aloisio; Lorenzo Belfanti; Rosa Giglio; Gabriella Cirigliano
Journal:  Sci Rep       Date:  2019-07-09       Impact factor: 4.379

2.  Impact of lens opacity and axial length on concomitant screening of maculopathy by swept-source optical coherence tomography-based optical biometer.

Authors:  Yingyan Qin; Shaobi Ye; Liangping Liu; Mingxing Wu
Journal:  Ann Transl Med       Date:  2022-08
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.