| Literature DB >> 26032991 |
Tomoya Segawa1, Shokichi Teramoto2, Kenji Omi3, Osamu Miyauchi4, Yoshiaki Watanabe5, Hisao Osada6.
Abstract
BACKGROUND: The improved reagent for measuring estradiol (E2), the ST AIA-PACK iE2 reagent, has a higher specificity for the measurement of E2 levels than the original ST AIA-PACK E2 reagent, because of its lower cross-reactivity with estrone (E1). As we had E2 data obtained with either of the reagents, we analyzed changes in E1 and E2 levels during follicle development.Entities:
Mesh:
Substances:
Year: 2015 PMID: 26032991 PMCID: PMC4461898 DOI: 10.1186/s12958-015-0051-y
Source DB: PubMed Journal: Reprod Biol Endocrinol ISSN: 1477-7827 Impact factor: 5.211
Cross-reactivity of ST AIA-PACK E2 and ST AIA-PACK iE2 reagents
| Cross-reactivity (mol %) | ||
|---|---|---|
| Substance | ST AIA-PACK E2 reagent | ST AIA-PACK iE2 reagent |
| Estradiol (E2) | 100 | 100 |
| Estrone (E1) | 7.2 | 1.1 |
| Estriol (E3) | 4.9 | 0.25 |
Please refer to the following websites for additional data. (The text is in Japanese but the data referred to above is in English)
http://www.info.pmda.go.jp/downfiles/ivd/PDF/480201_13A2X00174000017_B_02_01.pdf
http://www.info.pmda.go.jp/downfiles/ivd/PDF/480201_13A2X00174000006_A_04_01.pdf
E2 concentrations measured with the E2 reagent and iE2 reagent according to largest follicle size
| Diameter of the largest follicle | Serum E2 (pg/mL) | |||||
|---|---|---|---|---|---|---|
| Age <38 year ( | Age 38–40 year ( | Age >40 year ( | ||||
| 11 mm | E2 reagent ( | 99.9 ± 32.6a | E2 reagent ( | 101.5 ± 30.9a | E2 reagent ( | 113.4 ± 45.3a |
| iE2 reagent ( | 68.0 ± 34.7a | iE2 reagent ( | 71.0 ± 40.4a | iE2 reagent ( | 86.8 ± 51.8a | |
| 12 mm | E2 reagent ( | 105.8 ± 35.6a | E2 reagent ( | 113.5 ± 40.9a | E2 reagent ( | 125.3 ± 48.2a |
| iE2 reagent ( | 80.4 ± 41.8a | iE2 reagent ( | 84.7 ± 42.9a | iE2 reagent ( | 102.8 ± 53.5a | |
| 13 mm | E2 reagent ( | 122.2 ± 44.6a | E2 reagent ( | 135.9 ± 49.0a | E2 reagent ( | 142.2 ± 49.1a |
| iE2 reagent ( | 87.3 ± 42.6a | iE2 reagent ( | 96.7 ± 44.6a | iE2 reagent ( | 116.1 ± 61.2a | |
| 14 mm | E2 reagent ( | 130.7 ± 38.8a | E2 reagent ( | 155.2 ± 57.4a | E2 reagent ( | 166.6 ± 59.8a |
| iE2 reagent ( | 100.3 ± 51.7a | iE2 reagent ( | 115.8 ± 58.6a | iE2 reagent ( | 126.1 ± 64.0a | |
| 15 mm | E2 reagent ( | 168.0 ± 57.0a | E2 reagent ( | 168.8 ± 60.2a | E2 reagent ( | 187.0 ± 70.6a |
| iE2 reagent ( | 132.6 ± 64.2a | iE2 reagent ( | 150.2 ± 66.5a | iE2 reagent ( | 170.6 ± 76.5a | |
| 16 mm | E2 reagent ( | 186.7 ± 67.8a | E2 reagent ( | 195.4 ± 60.0a | E2 reagent ( | 210.4 ± 64.6 |
| iE2 reagent ( | 160.0 ± 77.9a | iE2 reagent ( | 177.0 ± 78.2a | iE2 reagent ( | 197.4 ± 79.5 | |
| 17 mm | E2 reagent ( | 228.5 ± 66.2a | E2 reagent ( | 236.6 ± 64.0a | E2 reagent ( | 231.8 ± 61.1 |
| iE2 reagent ( | 200.2 ± 85.7a | iE2 reagent ( | 216.5 ± 78.5a | iE2 reagent ( | 222.3 ± 86.6 | |
| 18 mm | E2 reagent ( | 241.0 ± 75.0 | E2 reagent ( | 257.3 ± 60.9a | E2 reagent ( | 255.9 ± 59.2 |
| iE2 reagent ( | 229.2 ± 85.4 | iE2 reagent ( | 233.5 ± 88.9a | iE2 reagent ( | 247.0 ± 82.2 | |
| 19 mm | E2 reagent ( | 253.4 ± 66.1 | E2 reagent ( | 260.8 ± 68.2 | E2 reagent ( | 259.2 ± 64.6 |
| iE2 reagent ( | 249.6 ± 78.0 | iE2 reagent ( | 256.0 ± 79.5 | iE2 reagent ( | 261.0 ± 83.5 | |
| 20 mm | E2 reagent ( | 275.2 ± 72.2 | E2 reagent ( | 279.9 ± 75.1 | E2 reagent ( | 281.5 ± 74.2 |
| iE2 reagent ( | 275.6 ± 95.2 | iE2 reagent ( | 285.1 ± 88.0 | iE2 reagent ( | 280.6 ± 92.0 | |
Data are presented as mean ± standard deviation
aSignificant differences in serum E2 concentration were found between measurements performed using the ST AIA-PACK E2 reagent and the ST AIA-PACK iE2 reagent (P <0.01)
E2 estradiol, E2 reagent ST AIA-PACK E2 reagent, iE2 reagent ST AIA-PACK iE2 reagent
E2 and FSH concentrations in the three age groups, classified according to follicle size
| Diameter of the largest follicle | Serum E2 (pg/mL) measured by iE2 reagent | Serum FSH (mIU/mL) | ||||
|---|---|---|---|---|---|---|
| ( | ( | |||||
| Age <38 year | Age 38–40 year | Age >40 year | Age <38 year | Age 38–40 year | Age >40 year | |
| 11 mm | 68.0 ± 34.7a | 71.0 ± 40.4a | 86.8 ± 51.8a | 8.6 ± 2.4b | 9.5 ± 4.6 | 10.9 ± 6.8b |
| 12 mm | 80.4 ± 41.8a | 84.7 ± 42.9a | 102.8 ± 53.5a | 8.4 ± 2.4a | 8.8 ± 3.0a | 11.1 ± 7.1a |
| 13 mm | 87.3 ± 42.6a | 96.7 ± 44.6a | 116.1 ± 61.2a | 7.9 ± 2.3a | 8.3 ± 3.2a | 10.3 ± 7.0a |
| 14 mm | 100.3 ± 51.7c | 115.8 ± 58.6c | 126.1 ± 64.0c | 7.4 ± 1.9c | 8.5 ± 3.5c | 9.9 ± 8.5c |
| 15 mm | 132.6 ± 64.2d | 150.2 ± 66.5d | 170.6 ± 76.5d | 7.3 ± 2.3c | 8.3 ± 3.7c | 8.8 ± 4.3c |
| 16 mm | 160.0 ± 77.9a | 177.0 ± 78.2a | 197.4 ± 79.5a | 7.1 ± 2.5a | 7.7 ± 3.6a | 9.0 ± 5.2a |
| 17 mm | 200.2 ± 85.7b | 216.5 ± 78.5 | 222.3 ± 86.6b | 6.8 ± 2.2d | 7.7 ± 3.1d | 8.8 ± 6.0d |
| 18 mm | 229.2 ± 85.4b | 233.5 ± 88.9 | 247.0 ± 82.2b | 6.6 ± 2.3d | 7.2 ± 2.8d | 8.5 ± 5.1d |
| 19 mm | 249.6 ± 78.0 | 256.0 ± 79.5 | 261.0 ± 83.5 | 6.3 ± 2.3a | 6.8 ± 2.3a | 8.3 ± 6.4a |
| 20 mm | 275.6 ± 95.2 | 285.1 ± 88.0 | 280.6 ± 92.0 | 6.5 ± 2.2a | 6.8 ± 2.3a | 7.8 ± 4.0a |
Data are presented as mean ± standard deviation
aSignificant differences were found between the <38 year group and the >40 year group, as well as between the 38–40 year group and the >40 year group (P < 0.01)
bSignificant differences were found between the <38 year group and the >40 year group (P < 0.01)
cSignificant differences were found between the <38 year group and the 38–40 year group, as well as between the <38 year group and the >40 year group (P < 0.01)
dSignificant differences were found among all three age groups (P < 0.01)
E2 estradiol, FSH follicle-stimulating hormone, iE2 reagent ST AIA-PACK iE2 reagent
Comparison of estrogen measurements in the early follicular phase and the ovulatory phase
| Early follicular phase | Ovulatory phase | |
|---|---|---|
| ( | ( | |
| Serum E2 (pg/mL) | ||
| Measured with ST AIA-PACK E2 reagent (A) | 58.7 ± 7.9 | 245.4 ± 50.5 |
| Measured with ST AIA-PACK iE2 reagent (B) | 21.3 ± 4.7 | 262.9 ± 54.2 |
| (A)/(B) | 2.9 ± 0.6a | 0.9 ± 0.1a |
| Serum E1 (pg/mL) | 68.0 ± 13.9 | 191.1 ± 51.9 |
| E1/E2 (measured with A) | 3.4 ± 1.1a | 0.7 ± 0.1a |
| Serum E3 (pg/mL)b | <5.0 | <5.0 |
Data are presented as mean ± standard deviation
aSignificant differences were found between the early follicular phase and the ovulatory phase groups (P <0.01)
bSerum E3 concentrations were below the detection range for both groups
E1 estrone, E2 estradiol, E3 estriol