The optimal tomotherapy treatment planning parameters for extremity soft tissue sarcomas.

Philippe Meyer1, Hugo Bouhours2, Nicolas Dehaynin2, Delphine Jarnet2, Matthieu Gantier2, Diran Karamanoukian2, Claudine Niederst2.   

Abstract

BACKGROUND AND
PURPOSE: To determine the optimum combination of treatment parameters between pitch, field width (FW) and modulation factor (MF) for extremity sarcomas in tomotherapy.
MATERIALS AND METHODS: Six patients previously treated for extremity sarcomas (3 arms and 3 legs) with tomotherapy were included in this study. 288 treatment plans were recalculated, corresponding to all combinations between 2 FW (2.5 and 5 cm), 4 MF (1.5, 2, 2.5 and 3) and 6 pitches (0.215, 0.287, 0.43 and 3 off-axis pitches). The treatment parameters (MF, FW or pitch) are modified between each plan, and the calculation is relaunched for 400 iterations, without modifying the optimisation constraints of the plan under which the patient has been treated.
RESULTS: We suggest eliminating the 0.43 pitch and never combining a 0.215 pitch with an MF ≤ 2. We also do not recommend using an MF = 1.5 unless treatment time is an absolute priority over plan quality. We did not see any advantage in using Chen off-axis pitches, except for targets far from the axis (>15 cm) treated with a high pitch. A combination of MF = 2/FW = 5 cm/pitch = 0.287 gives plans of acceptable quality, combined with reduced treatment times. These conclusions are true only for extremity sarcomas treated in 2 Gy/fraction.
CONCLUSIONS: We have shown that the choice of pitch/MF/FW combination is crucial for the treatment of extremity sarcomas in tomotherapy: some produce good dosimetric quality with a reduced irradiation time, while others may increase the time without improving the quality.
Copyright © 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Field width; Modulation factor; Pitch; Sarcomas; Tomotherapy planning

Mesh:

Year:  2015        PMID: 26032005     DOI: 10.1016/j.ejmp.2015.05.005

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  2 in total

1.  Influence of Modulation Factor on Treatment Plan Quality and Irradiation Time in Hippocampus-Sparing Whole-Brain Radiotherapy Using Tomotherapy.

Authors:  Akihiko Ishibashi; Hiromasa Kurosaki; Kosei Miura; Nobuko Utsumi; Hideyuki Sakurai
Journal:  Technol Cancer Res Treat       Date:  2021 Jan-Dec

2.  Interfacility variation in treatment planning parameters in tomotherapy: field width, pitch, and modulation factor.

Authors:  Hidetoshi Shimizu; Koji Sasaki; Takashi Kubota; Hiroshi Fukuma; Takahiro Aoyama; Tohru Iwata; Hiroyuki Tachibana; Takeshi Kodaira
Journal:  J Radiat Res       Date:  2018-09-01       Impact factor: 2.724

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.