Literature DB >> 26031443

Temperature rises in the round window caused by various light sources during insertion of rigid endoscopes: an experimental animal study.

R Dundar1, H Bulut2, A Yükkaldiran1, O K Güler1, Y Demirtaş2, I Iynen1, F Bozkuş1, F K Soy1.   

Abstract

OBJECTIVES: The instruments used in endoscopic surgery include rigid endoscopes of different diameters and angles, sources of light and monitors. In this study, we explored temperature rises in the round window caused by insertion of rigid endoscopes of different diameters into the middle ear; the endoscopes were fitted with different light sources.
DESIGN: An experimental animal study.
SETTING: We subjected seven guinea pigs to simulate otological surgery at room temperature. We monitored body temperatures, along with temperature rises caused by the use of 0° rigid endoscopes of diameters 3 or 4 mm, fitted with light sources including a halogen light, a light-emitting diode (LED) and a xenon light. Data were collected every second from a sensor placed in the round window. PARTICIPANTS: An experimental study on 7 guinea pigs. MAIN OUTCOME MEASURES: Rise of the temperature on round window.
RESULTS: Rigid endoscopes caused the temperature of the tympanum to rise when xenon and halogen light sources were used, regardless of endoscope diameter. However, the temperature rise was less when a LED light source was employed.
CONCLUSION: The endoscopic instruments used in middle ear surgery caused the temperature of the round window to rise. The rise varied with endoscope diameter and the type of light source used.
© 2015 John Wiley & Sons Ltd.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 26031443     DOI: 10.1111/coa.12468

Source DB:  PubMed          Journal:  Clin Otolaryngol        ISSN: 1749-4478            Impact factor:   2.597


  3 in total

1.  Thermal Safety of Endoscopic Usage in Robot-Assisted Middle Ear Surgery: An Experimental Study.

Authors:  Jinxi Pan; Haoyue Tan; Jun Shi; Zhaoyan Wang; Olivier Sterkers; Huan Jia; Hao Wu
Journal:  Front Surg       Date:  2021-05-14

2.  The Effect of Endoscopic Tympanoplasty on Cochlear Function.

Authors:  Isa Kaya; Goksel Turhal; Arin Ozturk; Sercan Gode; Cem Bilgen; Tayfun Kirazli
Journal:  Clin Exp Otorhinolaryngol       Date:  2017-11-28       Impact factor: 3.372

3.  Comparison of the anatomical and functional success of fascia and perichondrium grafts in transcanal endoscopic type 1 tympanoplasty.

Authors:  Kadir Özdamar; Alper Sen
Journal:  J Otolaryngol Head Neck Surg       Date:  2019-11-27
  3 in total