| Literature DB >> 26019609 |
Chan-Myeong Han1, Joon-Ho Park2, Chu-Seok Chang2, Myung-Chun Ryoo2.
Abstract
A new method for spot matching in two-dimensional gel electrophoresis images using a cumulative frequency matrix is proposed. The method improves on the weak points of the previous method called 'spot matching by topological patterns of neighbour spots'. It accumulates the frequencies of neighbour spot pairs produced through the entire matching process and determines spot pairs one by one in order of higher frequency. Spot matching by frequencies of neighbour spot pairs shows a fairly better performance. However, it can give researchers a hint for whether the matching results can be trustworthy or not, which can save researchers a lot of effort for verification of the results.Entities:
Keywords: cumulative frequency matrix; matching probability; neighbour spot; spot matching; topological pattern
Year: 2014 PMID: 26019609 PMCID: PMC4434135 DOI: 10.1080/13102818.2014.949036
Source DB: PubMed Journal: Biotechnol Biotechnol Equip ISSN: 1310-2818 Impact factor: 1.632
Figure 1. Examples of neighbour spots by Gabriel graph (left) and relative neighbour graph (right).
Figure 2. Spot matching by topological patterns of neighbour spots.
Figure 3. Process of matching neighbour spots.
Figure 4. Cumulative frequency matrix.
Figure 5. Format of landmark.tbl.
Spot-matching information for gel no. 34.
| xRsample | yRsample | xSample | ySample | lmNbr |
|---|---|---|---|---|
| 207 | 190 | 197 | 206 | 1 |
| 176 | 151 | 166 | 161 | 2 |
| 158 | 190 | 149 | 205 | 3 |
| 183 | 203 | 177 | 222 | 4 |
| 186 | 225 | 180 | 244 | 5 |
| 127 | 227 | 132 | 250 | 6 |
| 144 | 241 | 144 | 263 | 7 |
| 107 | 265 | 113 | 287 | 8 |
| 179 | 295 | 176 | 321 | 9 |
| 234 | 232 | 221 | 251 | 10 |
| 251 | 250 | 237 | 265 | 11 |
| 270 | 183 | 252 | 198 | 12 |
| 237 | 166 | 221 | 177 | 13 |
| 295 | 218 | 279 | 235 | 14 |
| 304 | 285 | 292 | 305 | 15 |
| 247 | 313 | 239 | 337 | 16 |
| 198 | 325 | 195 | 351 | 17 |
| 95 | 350 | 110 | 372 | 18 |
| 104 | 386 | 120 | 415 | 19 |
| 156 | 381 | 166 | 412 | 20 |
| 204 | 460 | 209 | 496 | 21 |
| 285 | 417 | 289 | 451 | 22 |
Figure 6. Topological pattern of gel number 34.
Figure 7. 5-NNG graph of reference (left) and target gel (right).
Figure 8. Part of result obtained by TPNS.
Figure 9. Cumulative frequency matrix of gel no. 34.
Result from experiment.
| Item | Values |
|---|---|
| Total number of gel pairs | 128 |
| Total number of spot pairs | 2763 |
| Number of correctly matched spot pairs | 2761 |
| Detection rate | 100% |
| Matching accuracy | 99.93% |