| Literature DB >> 26014285 |
Leonardo Di Donna1, Hicham Benabdelkamel, Domenico Taverna, Serena Indelicato, Donatella Aiello, Anna Napoli, Giovanni Sindona, Fabio Mazzotti.
Abstract
A method for the determination and quantification of ketosteroid hormones in meat by mass spectrometry, based on the derivatization of the carbonyl moiety of steroids by O-methylhydroxylamine, is presented. The quantitative assay is performed by means of multiple-reaction-monitoring (MRM) scan mode and using the corresponding labelled species, obtained by reaction with d 3-methoxylamine, as internal standard. The accuracy of the method was established by evaluating artificially spiked samples, obtaining values in the range 90-110%. Recovery tests were performed on blank matrix samples spiked with non-natural steroids including trenbolone and melengestrol acetate. The latter experiment revealed that the yield of the extraction processes was approximately 60%. Good values of LOQ and LOD were achieved, making this method competitive with current hormone assay methods.Entities:
Mesh:
Substances:
Year: 2015 PMID: 26014285 DOI: 10.1007/s00216-015-8772-5
Source DB: PubMed Journal: Anal Bioanal Chem ISSN: 1618-2642 Impact factor: 4.142