Literature DB >> 26009905

In vitro fatigue and fracture resistance of one- and two-piece CAD/CAM zirconia implant abutments.

Peter Gehrke, Dirk Johannson, Carsten Fischer, Bogna Stawarczyk, Florian Beuer.   

Abstract

PURPOSE: All-ceramic abutments are employed increasingly often in implant dentistry for esthetic reasons. In vitro stress testing is required to evaluate the suitability of these constructions, especially in load-bearing posterior regions. The purpose of the study was to assess and compare the fatigue and fracture resistance of one- and two-piece computer-aided design/computer-assisted manufacture (CAD/CAM) zirconia implant abutments with an internal-hex connection and prefabricated commercially available zirconia stock abutments.
MATERIALS AND METHODS: Twenty-one abutment-crown specimens were prepared for three test groups. Control group 1 (SZ) included specimens with unprepared stock zirconia abutments, test group 2 (OP) included one-piece CAD/CAM zirconia abutments, and test group 3 (TP) included two-piece CAD/CAM zirconia abutments. All 21 specimens underwent thermocycling and fatigue testing. Finally, all specimens were tested for fracture resistance with a universal testing machine. The maximum load was applied to the tapered occlusal area of each crown at a 30-degree angle and a crosshead speed of 0.5 mm/min until the implant-abutment connection failed. Kolmogorov-Smirnov, Shapiro-Wilk, and post-hoc Scheffé tests were used for statistical analysis.
RESULTS: All abutments in groups SZ and OP fractured into two or more pieces after fracture resistance testing. None of the TP abutments displayed apparent disintegration, but failure was evidenced by bending of the retention screw. OP abutments (232.1 ± 29.8 N) and SZ abutments (251.8 ± 23.2 N) showed lower fracture loads than the TP abutments (291.4 ± 27.8 N). However, only the difference between the OP and TP groups was statistically significant. Further load-displacement analyses corroborated the higher mechanical stability of the TP abutments.
CONCLUSION: Superior resistance was achieved for two-piece hybrid CAD/CAM zirconia abutments. These abutments might be clinically beneficial in high-load areas, such as premolar and molar regions.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26009905     DOI: 10.11607/jomi.3942

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  14 in total

1.  Mechanical stability of angulated zirconia abutments supporting maxillary anterior single crowns on narrow-diameter implants.

Authors:  Ruiyang Ma; Ping Yu; Yuqiang Zhang; Chenyang Xie; Xin Tan; Jikui Sun; Haiyang Yu
Journal:  Clin Oral Investig       Date:  2022-09-26       Impact factor: 3.606

2.  In vitro performance and fracture resistance of CAD/CAM-fabricated implant supported molar crowns.

Authors:  Martin Rosentritt; Sebastian Hahnel; Frank Engelhardt; Michael Behr; Verena Preis
Journal:  Clin Oral Investig       Date:  2016-07-01       Impact factor: 3.573

3.  Stock Versus CAD/CAM Customized Zirconia Implant Abutments - Clinical and Patient-Based Outcomes in a Randomized Controlled Clinical Trial.

Authors:  Ulf Schepke; Henny J A Meijer; Wouter Kerdijk; Gerry M Raghoebar; Marco Cune
Journal:  Clin Implant Dent Relat Res       Date:  2016-07-31       Impact factor: 3.932

4.  The biocompatibility and mechanical properties of plasma sprayed zirconia coated abutment.

Authors:  Zhengfei Huang; Zhifeng Wang; Kaifeng Yin; Chuanhua Li; Meihua Guo; Jing Lan
Journal:  J Adv Prosthodont       Date:  2020-06-18       Impact factor: 1.904

5.  Tensile Bond Strength of Three Custom-made Tooth-Colored Implant Superstructures to Titanium Inserts.

Authors:  Mitra Zirak; Mahroo Vojdani; Amir Ali Reza Khaledi; Mitra Farzin
Journal:  J Dent (Shiraz)       Date:  2019-06

6.  Effects of Liner-Bonding of Implant-Supported Glass-Ceramic Crown to Zirconia Abutment on Bond Strength and Fracture Resistance.

Authors:  Yong-Seok Jang; Sang-Hoon Oh; Won-Suck Oh; Min-Ho Lee; Jung-Jin Lee; Tae-Sung Bae
Journal:  Materials (Basel)       Date:  2019-08-30       Impact factor: 3.623

7.  Reverse torque evaluation in indexed and nonindexed abutments of Morse Taper implants in a mechanical fatigue test.

Authors:  Caio Marques Martins; Elimário Venturin Ramos; Simone Kreve; Geraldo Alberto Pinheiro de Carvalho; Aline Batista Gonçalves Franco; Luís Guilherme Scavone de Macedo; Alecsandro de Moura Silva; Sérgio Candido Dias
Journal:  Dent Res J (Isfahan)       Date:  2019 Mar-Apr

8.  Fracture resistance of different implant supported ceramic abutment/crown systems.

Authors:  Merve Bankoglu Gungor; Secil Karakoca Nemli; Handan Yilmaz; Cemal Aydin
Journal:  Eur Oral Res       Date:  2019-05-01

9.  A Comparative Study on the Mechanical Properties of a Polymer-Infiltrated Ceramic-Network Material Used for the Fabrication of Hybrid Abutment.

Authors:  Salim Ongun; Sevcan Kurtulmus-Yilmaz; Gökçe Meriç; Mutahhar Ulusoy
Journal:  Materials (Basel)       Date:  2018-09-11       Impact factor: 3.623

10.  Influence of scaling procedures on the integrity of titanium nitride coated CAD/CAM abutments.

Authors:  Peter Gehrke; Emmanouil Spanos; Carsten Fischer; Helmut Storck; Florian Tebbel; Dirk Duddeck
Journal:  J Adv Prosthodont       Date:  2018-06-12       Impact factor: 1.904

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.