| Literature DB >> 25992000 |
Regina Tangerino de Souza Jacob1, Ticiana Cristina de Souza Zambonato2, Maria Fernanda Capoani Garcia Mondelli1.
Abstract
INTRODUCTION: Description for using stethoscopes adapted to hearing aids. AIM: To describe the adaptation of HAs to stethoscopes used by 2 students in the health field with bilateral hearing impairment. CASE REPORTS: Two subjects with hearing loss had their stethoscopes coupled to HAs because of the individual requirements of their professions (healthcare) to perform auscultation.Entities:
Keywords: Hearing Aids; Hearing Loss; Stethoscopes
Year: 2013 PMID: 25992000 PMCID: PMC4423277 DOI: 10.7162/S1809-97772013000100017
Source DB: PubMed Journal: Int Arch Otorhinolaryngol ISSN: 1809-4864
Audiometric thresholds of Case 1.
| 250 Hz | 500 Hz | 1 Hz | 2 Hz | 3 Hz | 4 Hz | 6 Hz | 8 Hz | ||
|---|---|---|---|---|---|---|---|---|---|
| AWRE | 65 | 65 | 50 | 35 | 40 | 45 | 35 | 45 | dB |
| BRE | 15 | 20 | 15 | 10 | 15 | dB | |||
| AWLE | AUS | AUS | AUS | AUS | AUS | AUS | AUS | AUS | |
| BLE | AUS | AUS | AUS | AUS | AUS | AUS |
Legend: AWRE: Airway right ear; BRE: bone right ear; AWLE: Airway left ear; BLE: bone left ear; AUS: absent; Hz: Hertz; dB: decibel.
Audiometric thresholds of Case 2.
| 250 Hz | 500 Hz | 1 Hz | 2 Hz | 3 Hz | 4 Hz | 6 Hz | 8 Hz | ||
|---|---|---|---|---|---|---|---|---|---|
| AWRE | 30 | 35 | 50 | 65 | 70 | 70 | 75 | 80 | dB |
| BRE | 35 | 50 | 60 | AUS | AUS | dB | |||
| AWLE | 35 | 40 | 55 | 65 | 70 | 75 | 80 | 85 | dB |
| BLE | 35 | 50 | 60 | AUS | AUS | dB |
Legend: AWRE: Airway right ear; BRE: bone right ear; AWLE: Airway left ear; BLE: bone left ear; AUS: absent; Hz: Hertz; dB: decibel.
Figure 1.HA coupled to the stethoscope.
Results of the Real Ear Insertion Gain (REIG).
| Frequencies | 250Hz | 500Hz | 1KHz | 2KHz | 3KHz | 4KHz | |
|---|---|---|---|---|---|---|---|
| REIG (Case1) | 26.2 | 31,6 | 43,6 | 39,4 | 30,1 | 34,6 | dB |
| Target (Case 1) | 8,6 | 19,1 | 31,2 | 19,9 | 20,4 | 20,4 | dB |
| REIG (Case 2) | 14,3 | 23,4 | 34,8 | 30,3 | 33,6 | 32,1 | dB |
| Target (Case 2) | 15,1 | 24,3 | 36,7 | 33,1 | 32,2 | 35,2 | dB |
Figure 2.Results of the questionnaire used to verify the satisfaction of subjects.