Andrew Naylor1, Martin C Bone2, Anthony Unsworth3, Sumedh C Talwalkar4, Ian A Trail4, Thomas J Joyce2.
Abstract
Clinical results of the PyroCarbon proximal interphalangeal joint replacement are inconsistent with various complications reported. To address this, in vitro testing was conducted using finger joint simulators. Two PyroCarbon proximal interphalangeal prostheses were tested in a lubricant of dilute bovine serum to 5 × 10(6) cycles of flexion-extension (90°-30°) with dynamic forces of 10 N applied. At intervals of 3000 cycles testing ceased and a static load of 100 N was applied to simulate gripping. In addition, two 'control' prostheses were immersed alongside the test prostheses to account for lubricant absorption. Wear and roughness averages (Ra) were measured every 1 × 10(6) cycles. Minimal wear for all of the components was measured with a negligible increase in Ra for most of the components. One condyle of one component increased in Ra over the 5 × 10(6) cycles with a value above the recommended 50 nm. Unidirectional marks were visible on the condyle from micrographs, consistent with an abrasive wear mode. © IMechE 2015.Entities:
Keywords: Wear analysis/testing; biomechanical testing/analysis; joint simulators; orthopaedic tribology; surfaces