Abstract
OBJECTIVE: To evaluate the evolution of mammographic image quality in the state of Rio de Janeiro on the basis of parameters measured and analyzed during health surveillance inspections in the period from 2006 to 2011.Entities:
Keywords: Breast neoplasms; Health care; Mammography; Public health surveillance; Quality assurance; Quality control
Year: 2015 PMID: 25987749 PMCID: PMC4433297 DOI: 10.1590/0100-3984.2014.0047
Source DB: PubMed Journal: Radiol Bras ISSN: 0100-3984
Parameters and standards utilized in the study.
| Parameters | Standards |
|---|---|
| Monthly image quality testing | The facility performs monthly imaging quality testing and the results remain at the health surveillance disposal |
| Collimation evaluation | The x-rays must cover the whole film surface, without exceeding the compression paddle borders |
| Automatic exposure control | Variations should not be > 20% for breasts between 2 and 5 cm in thickness. It is desirable that the variation does not exceed 10% |
| Compression paddle force | The compression paddle force should be between 11 and 18 kg |
| Compression paddle alignment | The maximum acceptable deformation is 5 mm |
| Surface entrance dose in the phantom | The dose should be < 10 mGy |
| Films processing | Films processing must comply with the physical quality standards |
| Films response | The films must comply with minimum specifications for appropriate images generation |
| High-contrast details (microcalcifications) | Small-sized objects should be visualized up to sets with 0.25 mm in diameter |
| Low-contrast circular details (disks) | Low-contrast objects, polyester disks with 2 mm in diameter and thickness between 0.5 and 3.0 mm should be visualized up to the 7th disk |
| Low-contrast linear details (fibers) | Low-contrast linear objects of different diameters mimic fibrotic tissue into fat tissue and should be visualized up to 0.75 mm fiber |
| Visualization of tumor masses | Nylon spherical caps mimic tumor masses and should be visualized up to the 4th cap of 4.0 mm in diameter and 2.0 mm thickness |
| Background optical density (OD) | Measurement at a spot on the phantom image located at 6 cm from the chest wall and laterally centered on the film. Threshold between 1.10 and 1.80 |
| Contrast index | The image contrast affects the visualization of structures. Represented by optical density points on the film of the evaluated facility. Values between 0.55 and 0.67 |
| Absence of imaging artifacts | Imaging artifacts are defects in the image processing which may result either in loss or marking of information. They should not be present |
| Availability of mammography-specific film processor | The availability of mammography-specific film processor ensures the maintenance of proper adjustment tothe thickness of the mammographic film |
Criteria for parameters evolution analysis.
| Status | Description |
|---|---|
| Compliance stability | The compliance observed at the first inspection remained unchanged at the second inspection |
| Compliance rate decrease | The compliance observed at the first inspection was no longer observed at the second inspection |
| Compliance rate increase | The non-compliance observed at the first inspection evolved to a compliance situation at the second inspection |
| Stationary | The non-compliance observed at the first inspection remained unchanged at the second inspection |
Percentage of mammography apparatuses in compliance situation at the first and second inspections.
| Compliance (%) | ||
|---|---|---|
| Parameters | First inspection | Second inspection |
| Monthly image quality testing | 26.9 | 28.8 |
| Collimation evaluation | 64.6 | 62.5 |
| Automatic exposure control | 68 | 58 |
| Compression paddle force | 72.3 | 85.1 |
| Compression paddle alignment | 52.9 | 52.9 |
| Entrance dose on the phantom surface | 73.1 | 51.9 |
| Films processing | 54.5 | 72.7 |
| Facility film response | 47.7 | 72.7 |
| High-contrast details (microcalcifications) | 80.4 | 47.1 |
| Low-contrast circular details (disks) | 47.1 | 45.1 |
| Low-contrast linear details (fibers) | 96.1 | 92.2 |
| Tumor masses visualization | 60.8 | 76.5 |
| Background optical density | 68.6 | 62.7 |
| Contrast index | 66.0 | 64.0 |
| Absence of imaging artifacts | 64.7 | 94.1 |
| Availability of mammography-specific film processors | 88.5 | 88.2 |
Evolution of parameters related to mammography apparatuses and film processors.
| Status | ||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Stability | Compliance rate decrease | Compliance rate increase | Stationary | Total | ||||||||||
| Parameters | % | % | % | % | % | |||||||||
| Collimation evaluation | 17 | 32.7 | 13 | 25 | 12 | 23.1 | 6 | 11.5 | 48 | 92.3 | ||||
| Automatic exposure control | 25 | 48.1 | 9 | 17.3 | 4 | 7.7 | 12 | 23.1 | 50 | 96.2 | ||||
| Compression paddle force | 29 | 55.8 | 5 | 9.6 | 8 | 15.4 | 5 | 9.6 | 47 | 90.4 | ||||
| Compression paddle alignment | 18 | 34.6 | 9 | 17.3 | 9 | 17.3 | 15 | 28.8 | 51 | 98.1 | ||||
| Entrance dose on the phantom surface | 20 | 38.5 | 18 | 34.6 | 7 | 13.5 | 7 | 13.5 | 52 | 100 | ||||
| Films processing | 14 | 26.9 | 8 | 15.4 | 17 | 32.7 | 5 | 9.6 | 44 | 84.6 | ||||
| Facility film response | 13 | 25 | 5 | 9.6 | 19 | 36.5 | 7 | 13.5 | 44 | 84.6 | ||||
| Mammography-specific film processors | 36 | 69.2 | 3 | 5.8 | 3 | 5.8 | 3 | 5.8 | 52 | 100 | ||||
The difference for 100% corresponds to the absence of data in the inspection reports.
Evolution of parameters related to imaging quality.
| Status | ||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Stability | Compliance rate decrease | Compliance rate increase | Stationary | Total | ||||||||||
| Parameters | % | % | % | % | % | |||||||||
| Monthly imaging quality testing | 8 | 15.4 | 5 | 10.0 | 7 | 13.5 | 29 | 56 | 49 | 94.2 | ||||
| High-contrast details (microcalcifications) | 22 | 42.3 | 18 | 34.6 | 2 | 3.8 | 9 | 17.3 | 51 | 98.1 | ||||
| Low-contrast circular details (disks) | 12 | 23.1 | 12 | 23.1 | 11 | 21.2 | 16 | 30.8 | 51 | 98.1 | ||||
| Low-contrast linear details (fibers) | 47 | 90.4 | 1 | 1.9 | 0 | 0 | 3 | 5.8 | 51 | 98.1 | ||||
| Tumor masses visualization | 28 | 53.8 | 3 | 5.8 | 12 | 23.1 | 8 | 15.4 | 51 | 98.1 | ||||
| Background optical density | 23 | 44.2 | 11 | 21.2 | 9 | 17.3 | 8 | 15.4 | 51 | 98.1 | ||||
| Contrast index | 22 | 42.3 | 11 | 21.2 | 10 | 19.2 | 7 | 13.5 | 50 | 96.2 | ||||
| Absence of imaging artifacts | 31 | 59.6 | 2 | 3.8 | 16 | 30.8 | 2 | 3.8 | 51 | 98.1 | ||||
The difference for 100% corresponds to the absence of data in the inspection reports.