BACKGROUND: An artificial object that imitates human brain activity is called "phantom" and is used for evaluation of magnetoencephalography (MEG) systems. The accuracy of the phantom itself had not been guaranteed in the previous studies, although role of the phantom is to evaluate the accuracy of MEG measurement. The purposes of this paper are to develop a novel MEG phantom that can be calibrated and to demonstrate the advantages of the calibrated phantoms. NEW
METHOD: We proposed and fabricated a practical dry phantom that is composed of 50 isosceles-triangle coils based on Ilmoniemi's model. This phantom was calibrated based on three-dimensional measurement of the current paths in the phantom and on numerical calculations.
RESULTS: The calibrated positions of the equivalent current dipoles (ECDs) shifted 0.83mm, on average, from the designed positions. The uncertainties of the calibrated ECDs were also evaluated, by combining the uncertainties which could reasonably be attributed to them. COMPARISON WITH EXISTING METHOD(S): Furthermore, we demonstrated performance of the developed phantom through experimental evaluation of an MEG system. The results of this evaluation differed from those obtained using an uncalibrated phantom. Moreover, the calibrated phantom can provide detailed information regarding the uncertainty of the measurement and also the uncertainty of the phantom itself.
CONCLUSIONS: A more appropriate evaluation of MEG measurements can be achieved using a calibrated phantom.
Copyright © 2015 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Calibration; Magnetoencephalography; Measurement uncertainty; Phantom

Mesh:

Year:  2015        PMID: 25985908     DOI: 10.1016/j.jneumeth.2015.05.004

Source DB:  PubMed          Journal:  J Neurosci Methods        ISSN: 0165-0270            Impact factor:   2.390


  3 in total

1.  Magnetoencephalography Phantom Comparison and Validation: Hospital Universiti Sains Malaysia (HUSM) Requisite.

Authors:  Hazim Omar; Alwani Liyan Ahmad; Noburo Hayashi; Zamzuri Idris; Jafri Malin Abdullah
Journal:  Malays J Med Sci       Date:  2015-12

2.  Beamspace dual signal space projection (bDSSP): a method for selective detection of deep sources in MEG measurements.

Authors:  Kensuke Sekihara; Yoshiaki Adachi; Hiroshi K Kubota; Chang Cai; Srikantan S Nagarajan
Journal:  J Neural Eng       Date:  2018-03-12       Impact factor: 5.379

3.  A study of scalar optically-pumped magnetometers for use in magnetoencephalography without shielding.

Authors:  Richard J Clancy; Vladislav Gerginov; Orang Alem; Stephen Becker; Svenja Knappe
Journal:  Phys Med Biol       Date:  2021-09-03       Impact factor: 4.174

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.