First-in-man evaluation of the novel balloon delivery system STENTYS Xposition S for the self-apposing coronary artery stent: impact on longitudinal geographic miss during stenting.

Huangling Lu1, Alexander J J IJsselmuiden, Maik J Grundeken, Martina Nassif, Arie G de Vries, Auke Weevers, Martijn Scholte, René Spaargaren, Joanna J Wykrzykowska, Jan G P Tijssen, Robbert J de Winter, Karel T Koch.   

Abstract

AIMS: A novel balloon delivery system (BDS) for the self-apposing STENTYS sirolimus-eluting stent (SES) has been developed for highly precise longitudinal stent positioning and deployment. The aim of this first-in-man study is to report the quantitative coronary analysis (QCA) angiography and optical coherence tomography (OCT) results as well as the 30-day clinical outcomes of the STENTYS Xposition S SES. METHODS AND
RESULTS: We included 25 patients (mean age 66.1±10.7 years) with stable coronary artery disease (24%) or acute coronary syndrome (including STEMI in 40%). The device was successfully placed at the intended site in all cases (100%), without procedural complications. Longitudinal geographic miss (entire lesion length [on QCA] not completely covered by the stent) was not observed. Pre-procedural MLD on QCA was 1.30±0.74 mm and post-procedural MLD was 2.74±0.44 mm, p<0.001 (acute gain 1.44±0.70 mm). OCT analyses showed a low percentage of malapposed stent struts directly post stent placement (2.4%), which further decreased after post-dilatation (0.6%, p=0.013), while mean stent area increased (from 9.7 mm2 to 10.5 mm2, p<0.001). At 30-day clinical follow-up, one (4%) major adverse cardiac event (MACE) was observed. One acute stent thrombosis (ST) occurred immediately post procedure in a STEMI patient which was related to inadequate medication therapy.
CONCLUSIONS: This first-in-man study demonstrated that the use of the novel STENTYS Xposition S balloon delivery system is feasible with a high technical success rate without longitudinal geographical miss. Stent strut malapposition rate directly after STENTYS placement was low and improved further after post-dilatation.

Entities:  

Mesh:

Year:  2016        PMID: 25983027     DOI: 10.4244/EIJY15M05_07

Source DB:  PubMed          Journal:  EuroIntervention        ISSN: 1774-024X            Impact factor:   6.534


  2 in total

1.  Comparison between the STENTYS self-apposing bare metal and paclitaxel-eluting coronary stents for the treatment of saphenous vein grafts (ADEPT trial).

Authors:  A J J IJsselmuiden; C Simsek; A G van Driel; D Bouchez; G Amoroso; P Vermeersch; P P Karjalainen
Journal:  Neth Heart J       Date:  2018-02       Impact factor: 2.380

2.  Five-year clinical follow-up of the STENTYS self-apposing stent in complex coronary anatomy: a single-centre experience with report of specific angiographic indications.

Authors:  H Lu; R J Bekker; M J Grundeken; P Woudstra; J J Wykrzykowska; J G P Tijssen; R J de Winter; K T Koch
Journal:  Neth Heart J       Date:  2018-05       Impact factor: 2.380

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.