Twice cutting method reduces tibial cutting error in unicompartmental knee arthroplasty.

Hiroshi Inui1, Shuji Taketomi2, Ryota Yamagami2, Takaki Sanada2, Sakae Tanaka2.   

Abstract

BACKGROUND: Bone cutting error can be one of the causes of malalignment in unicompartmental knee arthroplasty (UKA). The amount of cutting error in total knee arthroplasty has been reported. However, none have investigated cutting error in UKA. The purpose of this study was to reveal the amount of cutting error in UKA when open cutting guide was used and clarify whether cutting the tibia horizontally twice using the same cutting guide reduced the cutting errors in UKA.
METHODS: We measured the alignment of the tibial cutting guides, the first-cut cutting surfaces and the second cut cutting surfaces using the navigation system in 50 UKAs. Cutting error was defined as the angular difference between the cutting guide and cutting surface.
RESULTS: The mean absolute first-cut cutting error was 1.9° (1.1° varus) in the coronal plane and 1.1° (0.6° anterior slope) in the sagittal plane, whereas the mean absolute second-cut cutting error was 1.1° (0.6° varus) in the coronal plane and 1.1° (0.4° anterior slope) in the sagittal plane. Cutting the tibia horizontally twice reduced the cutting errors in the coronal plane significantly (P<0.05).
CONCLUSION: Our study demonstrated that in UKA, cutting the tibia horizontally twice using the same cutting guide reduced cutting error in the coronal plane.
Copyright © 2014 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Computer-assisted navigation system; Cutting error; Tibial alignment; Twice cutting method; Unicompartmental knee arthroplasty

Mesh:

Year:  2014        PMID: 25982299     DOI: 10.1016/j.knee.2014.11.015

Source DB:  PubMed          Journal:  Knee        ISSN: 0968-0160            Impact factor:   2.199


  4 in total

1.  Case-related factors affecting cutting errors of the proximal tibia in total knee arthroplasty assessed by computer navigation.

Authors:  Tadashi Tsukeoka; Yoshikazu Tsuneizumi; Kensuke Yoshino; Mashiko Suzuki
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-12-22       Impact factor: 4.342

2.  A modified technique to reduce tibial keel cutting errors during an Oxford unicompartmental knee arthroplasty.

Authors:  Hiroshi Inui; Shuji Taketomi; Keitarou Tahara; Ryota Yamagami; Takaki Sanada; Sakae Tanaka
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-05-02       Impact factor: 4.342

3.  Effects of a 1-mm difference in bearing thickness on intraoperative bearing movement and kinematics in Oxford unicompartmental knee arthroplasty.

Authors:  Kohei Kawaguchi; Hiroshi Inui; Shuji Taketomi; Ryota Yamagami; Kenichi Kono; Shin Sameshima; Tomofumi Kage; Sakae Tanaka
Journal:  BMC Musculoskelet Disord       Date:  2022-04-09       Impact factor: 2.362

4.  Intraoperative rotational kinematics and its influence on postoperative clinical outcomes differ according to age in Unicompartmental knee Arthroplasty.

Authors:  Kohei Kawaguchi; Hiroshi Inui; Shuji Taketomi; Ryota Yamagami; Kenichi Kono; Shin Sameshima; Tomofumi Kage; Sakae Tanaka
Journal:  BMC Musculoskelet Disord       Date:  2021-06-01       Impact factor: 2.362