Ultrashort echo-time MRI versus CT for skull aberration correction in MR-guided transcranial focused ultrasound: In vitro comparison on human calvaria.

G Wilson Miller1, Matthew Eames2, John Snell3, Jean-François Aubry4.   

Abstract

PURPOSE: Transcranial magnetic resonance-guided focused ultrasound (TcMRgFUS) brain treatment systems compensate for skull-induced beam aberrations by adjusting the phase and amplitude of individual ultrasound transducer elements. These corrections are currently calculated based on a preacquired computed tomography (CT) scan of the patient's head. The purpose of the work presented here is to demonstrate the feasibility of using ultrashort echo-time magnetic resonance imaging (UTE MRI) instead of CT to calculate and apply aberration corrections on a clinical TcMRgFUS system.
METHODS: Phantom experiments were performed in three ex-vivo human skulls filled with tissue-mimicking hydrogel. Each skull phantom was imaged with both CT and UTE MRI. The MR images were then segmented into "skull" and "not-skull" pixels using a computationally efficient, threshold-based algorithm, and the resulting 3D binary skull map was converted into a series of 2D virtual CT images. Each skull was mounted in the head transducer of a clinical TcMRgFUS system (ExAblate Neuro, Insightec, Israel), and transcranial sonications were performed using a power setting of approximately 750 acoustic watts at several different target locations within the electronic steering range of the transducer. Each target location was sonicated three times: once using aberration corrections calculated from the actual CT scan, once using corrections calculated from the MRI-derived virtual CT scan, and once without applying any aberration correction. MR thermometry was performed in conjunction with each 10-s sonication, and the highest single-pixel temperature rise and surrounding-pixel mean were recorded for each sonication.
RESULTS: The measured temperature rises were ∼ 45% larger for aberration-corrected sonications than for noncorrected sonications. This improvement was highly significant (p < 10(-4)). The difference between the single-pixel peak temperature rise and the surrounding-pixel mean, which reflects the sharpness of the thermal focus, was also significantly larger for aberration-corrected sonications. There was no significant difference between the sonication results achieved using CT-based and MR-based aberration correction.
CONCLUSIONS: The authors have demonstrated that transcranial focal heating can be significantly improved in vitro by using UTE MRI to compute skull-induced ultrasound aberration corrections. Their results suggest that UTE MRI could be used instead of CT to implement such corrections on current 0.7 MHz clinical TcMRgFUS devices. The MR image acquisition and segmentation procedure demonstrated here would add less than 15 min to a clinical MRgFUS treatment session.

Entities:  

Mesh:

Year:  2015        PMID: 25979016     DOI: 10.1118/1.4916656

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  15 in total

1.  Comparison of analytical and numerical approaches for CT-based aberration correction in transcranial passive acoustic imaging.

Authors:  Ryan M Jones; Kullervo Hynynen
Journal:  Phys Med Biol       Date:  2015-11-25       Impact factor: 3.609

2.  Rapid dual-RF, dual-echo, 3D ultrashort echo time craniofacial imaging: A feasibility study.

Authors:  Hyunyeol Lee; Xia Zhao; Hee Kwon Song; Rosaline Zhang; Scott P Bartlett; Felix W Wehrli
Journal:  Magn Reson Med       Date:  2018-12-18       Impact factor: 4.668

Review 3.  Image-guided ultrasound phased arrays are a disruptive technology for non-invasive therapy.

Authors:  Kullervo Hynynen; Ryan M Jones
Journal:  Phys Med Biol       Date:  2016-08-05       Impact factor: 3.609

Review 4.  Ultrasound treatment of neurological diseases--current and emerging applications.

Authors:  Gerhard Leinenga; Christian Langton; Rebecca Nisbet; Jürgen Götz
Journal:  Nat Rev Neurol       Date:  2016-02-19       Impact factor: 42.937

5.  Adaptive Ultrasound Focusing Through the Cranial Bone for Non-invasive Treatment of Brain Disorders.

Authors:  Thomas Bancel; Thomas Tiennot; Jean-François Aubry
Journal:  Adv Exp Med Biol       Date:  2022       Impact factor: 2.622

6.  Acoustic properties across the human skull.

Authors:  Thomas S Riis; Taylor D Webb; Jan Kubanek
Journal:  Ultrasonics       Date:  2021-10-21       Impact factor: 2.890

7.  Measurements of the Relationship Between CT Hounsfield Units and Acoustic Velocity and How It Changes With Photon Energy and Reconstruction Method.

Authors:  Taylor D Webb; Steven A Leung; Jarrett Rosenberg; Pejman Ghanouni; Jeremy J Dahl; Norbert J Pelc; Kim Butts Pauly
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2018-07       Impact factor: 2.725

8.  Transcranial MR-Guided Histotripsy System.

Authors:  Ning Lu; Timothy L Hall; Dave Choi; Dinank Gupta; Badih Junior Daou; Jonathan R Sukovich; Adam Fox; Tyler I Gerhardson; Aditya S Pandey; Douglas C Noll; Zhen Xu
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2021-08-27       Impact factor: 3.267

9.  Acoustic Attenuation: Multifrequency Measurement and Relationship to CT and MR Imaging.

Authors:  Taylor D Webb; Steven A Leung; Pejman Ghanouni; Jeremy J Dahl; Norbert J Pelc; Kim Butts Pauly
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2021-04-26       Impact factor: 2.725

10.  MR thermometry for focused ultrasound monitoring utilizing model predictive filtering and ultrasound beam modeling.

Authors:  Henrik Odéen; Scott Almquist; Joshua de Bever; Douglas A Christensen; Dennis L Parker
Journal:  J Ther Ultrasound       Date:  2016-09-22
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.