Proton beam deflection in MRI fields: Implications for MRI-guided proton therapy.

B M Oborn1, S Dowdell2, P E Metcalfe3, S Crozier4, R Mohan5, P J Keall6.   

Abstract

PURPOSE: This paper investigates, via magnetic modeling and Monte Carlo simulation, the ability to deliver proton beams to the treatment zone inside a split-bore MRI-guided proton therapy system.
METHODS: Field maps from a split-bore 1 T MRI-Linac system are used as input to geant4 Monte Carlo simulations which model the trajectory of proton beams during their paths to the isocenter of the treatment area. Both inline (along the MRI bore) and perpendicular (through the split-bore gap) orientations are simulated. Monoenergetic parallel and diverging beams of energy 90, 195, and 300 MeV starting from 1.5 and 5 m above isocenter are modeled. A phase space file detailing a 2D calibration pattern is used to set the particle starting positions, and their spatial location as they cross isocenter is recorded. No beam scattering, collimation, or modulation of the proton beams is modeled.
RESULTS: In the inline orientation, the radial symmetry of the solenoidal style fringe field acts to rotate the protons around the beam's central axis. For protons starting at 1.5 m from isocenter, this rotation is 19° (90 MeV) and 9.8° (300 MeV). A minor focusing toward the beam's central axis is also seen, but only significant, i.e., 2 mm shift at 150 mm off-axis, for 90 MeV protons. For the perpendicular orientation, the main MRI field and near fringe field act as the strongest to deflect the protons in a consistent direction. When starting from 1.5 m above isocenter shifts of 135 mm (90 MeV) and 65 mm (300 MeV) were observed. Further to this, off-axis protons are slightly deflected toward or away from the central axis in the direction perpendicular to the main deflection direction. This leads to a distortion of the phase space pattern, not just a shift. This distortion increases from zero at the central axis to 10 mm (90 MeV) and 5 mm (300 MeV) for a proton 150 mm off-axis. In both orientations, there is a small but subtle difference in the deflection and distortion pattern between protons fired parallel to the beam axis and those fired from a point source. This is indicative of the 3D spatially variant nature of the MRI fringe field.
CONCLUSIONS: For the first time, accurate magnetic and Monte Carlo modeling have been used to assess the transport of generic proton beams toward a 1 T split-bore MRI. Significant rotation is observed in the inline orientation, while more complex deflection and distortion are seen in the perpendicular orientation. The results of this study suggest that due to the complexity and energy-dependent nature of the magnetic deflection and distortion, the pencil beam scanning method will be the only choice for delivering a therapeutic proton beam inside a potential MRI-guided proton therapy system in either the inline or perpendicular orientation. Further to this, significant correction strategies will be required to account for the MRI fringe fields.

Mesh:

Substances:

Year:  2015        PMID: 25979006     DOI: 10.1118/1.4916661

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  13 in total

Review 1.  Proton beam therapy for tumors of the upper abdomen.

Authors:  Ann Raldow; James Lamb; Theodore Hong
Journal:  Br J Radiol       Date:  2019-08-23       Impact factor: 3.039

2.  Proton Therapy for Vaginal Reirradiation.

Authors:  Yun Rose Li; Maura Kirk; Lilie Lin
Journal:  Int J Part Ther       Date:  2016-12-30

3.  Analytical investigation of magnetic field effects on Proton lateral deflection and penetrating depth in the water phantom: A relativistic approach.

Authors:  Mohammad Javad Tahmasebi Birgani; Nahid Chegeni; Mansour Zabihzadeh; Marziyeh Tahmasbi
Journal:  Electron Physician       Date:  2017-12-25

4.  A pencil beam algorithm for magnetic resonance image-guided proton therapy.

Authors:  Fatima Padilla-Cabal; Dietmar Georg; Hermann Fuchs
Journal:  Med Phys       Date:  2018-03-30       Impact factor: 4.071

Review 5.  MR-guided proton therapy: a review and a preview.

Authors:  Aswin Hoffmann; Bradley Oborn; Maryam Moteabbed; Susu Yan; Thomas Bortfeld; Antje Knopf; Herman Fuchs; Dietmar Georg; Joao Seco; Maria Francesca Spadea; Oliver Jäkel; Christopher Kurz; Katia Parodi
Journal:  Radiat Oncol       Date:  2020-05-29       Impact factor: 3.481

Review 6.  Current and Future Perspectives of the Use of Organoids in Radiobiology.

Authors:  Peter W Nagle; Robert P Coppes
Journal:  Cells       Date:  2020-12-09       Impact factor: 6.600

7.  Technical Note: Design and commissioning of a water phantom for proton dosimetry in magnetic fields.

Authors:  Hermann Fuchs; Fatima Padilla-Cabal; Andreas Hummel; Dietmar Georg
Journal:  Med Phys       Date:  2020-12-08       Impact factor: 4.071

Review 8.  Roadmap: proton therapy physics and biology.

Authors:  Harald Paganetti; Chris Beltran; Stefan Both; Lei Dong; Jacob Flanz; Keith Furutani; Clemens Grassberger; David R Grosshans; Antje-Christin Knopf; Johannes A Langendijk; Hakan Nystrom; Katia Parodi; Bas W Raaymakers; Christian Richter; Gabriel O Sawakuchi; Marco Schippers; Simona F Shaitelman; B K Kevin Teo; Jan Unkelbach; Patrick Wohlfahrt; Tony Lomax
Journal:  Phys Med Biol       Date:  2021-02-26       Impact factor: 4.174

9.  Benchmarking a GATE/Geant4 Monte Carlo model for proton beams in magnetic fields.

Authors:  Fatima Padilla-Cabal; Jose Alejandro Fragoso; Andreas Franz Resch; Dietmar Georg; Hermann Fuchs
Journal:  Med Phys       Date:  2019-11-13       Impact factor: 4.071

10.  Characterization of EBT3 radiochromic films for dosimetry of proton beams in the presence of magnetic fields.

Authors:  Fatima Padilla-Cabal; Peter Kuess; Dietmar Georg; Hugo Palmans; Lukas Fetty; Hermann Fuchs
Journal:  Med Phys       Date:  2019-05-31       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.