| Literature DB >> 25969234 |
Abstract
Illumination-based differential phase contrast (DPC) is a phase imaging method that uses a pair of images with asymmetric illumination patterns. Distinct from coherent techniques, DPC relies on spatially partially coherent light, providing 2× better lateral resolution, better optical sectioning and immunity to speckle noise. In this paper, we derive the 2D weak object transfer function (WOTF) and develop a quantitative phase reconstruction method that is robust to noise. The effect of spatial coherence is studied experimentally, and multiple-angle DPC is shown to provide improved frequency coverage for more stable phase recovery. Our method uses an LED array microscope to achieve real-time (10 Hz) quantitative phase imaging with in vitro live cell samples.Year: 2015 PMID: 25969234 DOI: 10.1364/OE.23.011394
Source DB: PubMed Journal: Opt Express ISSN: 1094-4087 Impact factor: 3.894