Hypoglycemic Accuracy and Improved Low Glucose Alerts of the Latest Dexcom G4 Platinum Continuous Glucose Monitoring System.

Thomas A Peyser1, Katherine Nakamura2, David Price2, Lucas C Bohnett2, Irl B Hirsch3, Andrew Balo2.   

Abstract

OBJECTIVE: Accuracy of continuous glucose monitoring (CGM) devices in hypoglycemia has been a widely reported shortcoming of this technology. We report the accuracy in hypoglycemia of a new version of the Dexcom (San Diego, CA) G4 Platinum CGM system (software 505) and present results regarding the optimum setting of CGM hypoglycemic alerts.
MATERIALS AND METHODS: CGM values were compared with YSI analyzer (YSI Life Sciences, Yellow Springs, OH) measurements every 15 min. We reviewed the accuracy of the CGM system in the hypoglycemic range using standard metrics. We analyzed the time required for the CGM system to detect biochemical hypoglycemia (70 mg/dL) compared with the YSI with alert settings at 70 mg/dL and 80 mg/dL. We also analyzed the time between the YSI value crossing 55 mg/dL, defined as the threshold for cognitive impairment due to hypoglycemia, and when the CGM system alerted for hypoglycemia.
RESULTS: The mean absolute difference for a glucose level of less than 70 mg/dL was 6 mg/dL. Ninety-six percent of CGM values were within 20 mg/dL of the YSI values between 40 and 80 mg/dL. When the CGM hypoglycemic alert was set at 80 mg/dL, the device provided an alert for biochemical hypoglycemia within 10 min in 95% of instances and at least a 10-min advance warning before the cognitive impairment threshold in 91% of instances in the study.
CONCLUSIONS: Use of an 80 mg/dL threshold setting for hypoglycemic alerts on the G4 Platinum (software 505) may provide patients with timely warning of hypoglycemia before the onset of cognitive impairment, enabling them to treat themselves for hypoglycemia with fast-acting carbohydrates and prevent neuroglycopenia associated with very low glucose levels.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25961446     DOI: 10.1089/dia.2014.0415

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  11 in total

1.  Hypoglycemia Reduction and Accuracy of Continuous Glucose Monitoring.

Authors:  Boris P Kovatchev
Journal:  Diabetes Technol Ther       Date:  2015-05-15       Impact factor: 6.118

2.  Analysis of the Accuracy and Performance of a Continuous Glucose Monitoring Sensor Prototype: An In-Silico Study Using the UVA/PADOVA Type 1 Diabetes Simulator.

Authors:  Marc D Breton; Rolf Hinzmann; Enrique Campos-Nañez; Susan Riddle; Michael Schoemaker; Guenther Schmelzeisen-Redeker
Journal:  J Diabetes Sci Technol       Date:  2016-12-13

Review 3.  Future of Automated Insulin Delivery Systems.

Authors:  Jessica R Castle; J Hans DeVries; Boris Kovatchev
Journal:  Diabetes Technol Ther       Date:  2017-06       Impact factor: 6.118

4.  Continuous Glucose Monitoring Use in Clinical Trials for On-Market Diabetes Drugs.

Authors:  Bradley Q Fox; Peninah F Benjamin; Ammara Aqeel; Emily Fitts; Spencer Flynn; Brian Levine; Elizaveta Maslak; Rebecca L Milner; Benjamin Ose; Michael Poeschla; Meghna Ray; Maeve Serino; Sahaj S Shah; Kelly L Close
Journal:  Clin Diabetes       Date:  2021-04

5.  Continuous Glucose Monitoring: A Perspective on Its Past, Present, and Future Applications for Diabetes Management.

Authors:  Hanna S Mariani; Brian T Layden; Grazia Aleppo
Journal:  Clin Diabetes       Date:  2017-01

Review 6.  Selecting the Appropriate Continuous Glucose Monitoring System - a Practical Approach.

Authors:  Peter Adolfsson; Christopher G Parkin; Andreas Thomas; Lars G Krinelke
Journal:  Eur Endocrinol       Date:  2018-04-18

7.  Real-World Hypoglycemia Avoidance with a Continuous Glucose Monitoring System's Predictive Low Glucose Alert.

Authors:  Sarah Puhr; Mark Derdzinski; John B Welsh; Andrew Scott Parker; Tomas Walker; David A Price
Journal:  Diabetes Technol Ther       Date:  2019-03-22       Impact factor: 6.118

8.  Hypoglycemic symptoms in the absence of diabetes: Pilot evidence of clinical hypoglycemia in young women.

Authors:  Morgana Mongraw-Chaffin; Daniel P Beavers; Donald A McClain
Journal:  J Clin Transl Endocrinol       Date:  2019-07-24

Review 9.  Outcome Measures for Artificial Pancreas Clinical Trials: A Consensus Report.

Authors:  David M Maahs; Bruce A Buckingham; Jessica R Castle; Ali Cinar; Edward R Damiano; Eyal Dassau; J Hans DeVries; Francis J Doyle; Steven C Griffen; Ahmad Haidar; Lutz Heinemann; Roman Hovorka; Timothy W Jones; Craig Kollman; Boris Kovatchev; Brian L Levy; Revital Nimri; David N O'Neal; Moshe Philip; Eric Renard; Steven J Russell; Stuart A Weinzimer; Howard Zisser; John W Lum
Journal:  Diabetes Care       Date:  2016-07       Impact factor: 19.112

Review 10.  The Rationale for Continuous Glucose Monitoring-based Diabetes Treatment Decisions and Non-adjunctive Continuous Glucose Monitoring Use.

Authors:  David Price; Tomas Walker
Journal:  Eur Endocrinol       Date:  2016-03-15
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.