Literature DB >> 25957512

Liquid based microbiological transport systems: conformity assessment of two commercial devices.

Manuela Avolio1, Alessandro Camporese2.   

Abstract

We compared two types of liquid-based microbiology devices for microorganism viability according to standardized quantitative elution method CLSI M40-A2. The eSwab® met CLSI acceptance criteria of viability maintenance for all microorganisms tested. The Σ-Transwab® failed to meet CLSI acceptance criteria for Peptostreptococcus anaerobius, Prevotella melaninogenica, Fusobacterium nucleatum and Haemophilus influenzae.
Copyright © 2015 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  CLSI M40-A2; Flocked swabs; Liquid-based microbiology

Mesh:

Substances:

Year:  2015        PMID: 25957512     DOI: 10.1016/j.mimet.2015.05.004