Endovascular coil detachment causing EMG artefact in BIS: a mechanistic exploration.

Dhritiman Chakrabarti1, Venkatapura J Ramesh2, Hima Pendharkar3.   

Abstract

Deployment of endovascular coils used in interventional neuroradiology commonly involves electrolytic detachment of the coil from the pusher catheter. This report describes a case of artefactual increase in electromyography (EMG) values of bispectral index (BIS) monitor during coil detachment. An explanation of this event is provided connecting mechanism of coil detachment and derivation of EMG values in a BIS monitor. While rising EMG values are thought to arise from frontalis contraction, they may as well be an unrecognized electrical artefact, especially in context of undistorted electroencephalography waveform.

Keywords:  Artefact; Bispectral index; Electromyography; Endovascular coil detachment

Mesh:

Year:  2015        PMID: 25948093     DOI: 10.1007/s10877-015-9702-y

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  3 in total

1.  Diagnostic and prognostic use of bispectral index in coma, vegetative state and related disorders.

Authors:  C Schnakers; D Ledoux; S Majerus; P Damas; F Damas; B Lambermont; M Lamy; M Boly; A Vanhaudenhuyse; G Moonen; S Laureys
Journal:  Brain Inj       Date:  2008-11       Impact factor: 2.311

2.  EMG contamination of EEG: spectral and topographical characteristics.

Authors:  I I Goncharova; D J McFarland; T M Vaughan; J R Wolpaw
Journal:  Clin Neurophysiol       Date:  2003-09       Impact factor: 3.708

Review 3.  A primer for EEG signal processing in anesthesia.

Authors:  I J Rampil
Journal:  Anesthesiology       Date:  1998-10       Impact factor: 7.892

  3 in total
  3 in total

Review 1.  Journal of clinical monitoring and computing 2016 end of year summary: anesthesia.

Authors:  Jan F A Hendrickx; Andre M De Wolf
Journal:  J Clin Monit Comput       Date:  2017-01-07       Impact factor: 2.502

2.  Artifactual increase in bispectral index caused by connector contaminated with povidone-iodine.

Authors:  Hee Yong Kang; Ann Hee You; Youngsoon Kim; Jeong-Hyun Choi
Journal:  J Clin Monit Comput       Date:  2018-11-12       Impact factor: 2.502

3.  Variations in Values of State, Response Entropy and Haemodynamic Parameters Associated with Development of Different Epileptiform Patterns during Volatile Induction of General Anaesthesia with Two Different Anaesthetic Regimens Using Sevoflurane in Comparison with Intravenous Induct: A Comparative Study.

Authors:  Michał Stasiowski; Anna Duława; Izabela Szumera; Radosław Marciniak; Ewa Niewiadomska; Wojciech Kaspera; Lech Krawczyk; Piotr Ładziński; Beniamin Oskar Grabarek; Przemysław Jałowiecki
Journal:  Brain Sci       Date:  2020-06-12
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.