Literature DB >> 25944843

A Nuclear Magnetic Resonance Spectroscopy as a Method for Evaluation of In Vivo Poly-l-Lactide Biodegradation Kinetics From Stent-Polymer Matrices: An Experimental Study Utilizing Porcine Model of In-Stent Restenosis.

Bartłomiej Orlik1, Piotr P Buszman2, Agata Krauze1, Paweł Gąsior1, Piotr Desperak1, Jacek Pająk3, Janusz Kasperczyk4, Adam Janas1, Michał Jelonek1, Gabriela Handzlik-Orlik1, Paweł E Buszman1, Krzysztof Milewski5.   

Abstract

OBJECTIVES: We aimed to comprehensively evaluate poly-lactide polymer degradation and sirolimus release kinetics from a drug-eluting stent matrix in the in vivo setting using a nuclear magnetic resonance (NMR) method.
METHODS: In 22 domestic swine, 18 biodegradable polymer-only coated stents (BPSs) and 36 biodegradable polymer-coated sirolimus-eluting stents (BP-SES) were implanted in coronary arteries with 115% overstretch. The animals were sacrificed at 1, 3, 7, 14, 28, and 56 days following baseline procedures. Vessel segments with BPS were harvested to evaluate polymer degradation with a NMR method, whereas BP-SES to analyze sirolimus tissue uptake and retention. Additionally, 8 BP-SES were implanted for histological analysis for 90 days of follow-up.
RESULTS: The NMR showed a gradual absorption of the polymer over the 6 consecutive time points, from 5.48 µg of the polymer on the stent at 1-day follow-up, through 4.33 µg at 3 days, 3.16 µg at 7 days, 2.42 µg at 14 days, 1.92 µg at 28 days to 1.24 µg in the last day of the study. The curve of polymer degradation corresponds well with the pharmacokinetic profile of sirolimus eluted from its surface and measured at identical time points. In histopathology, at 90 days, complete healing and biocompatibility were reported.
CONCLUSIONS: The utilization of NMR method for BP absorption kinetics evaluation is a useful tool, which may be widely adopted to test other biodegradable implants. Further, it may substantially improve their safety and efficacy by facilitating programmed polymer and drugs elution.
© The Author(s) 2015.

Entities:  

Keywords:  NMR spectroscopy; biodegradable polymer; coronary stents; porcine model; sirolimus-eluting stent

Mesh:

Substances:

Year:  2015        PMID: 25944843     DOI: 10.1177/1074248415583091

Source DB:  PubMed          Journal:  J Cardiovasc Pharmacol Ther        ISSN: 1074-2484            Impact factor:   2.457


  4 in total

1.  Polymer Coating Integrity, Thrombogenicity and Computational Fluid Dynamics Analysis of Provisional Stenting Technique in the Left Main Bifurcation Setting: Insights from an In-Vitro Model.

Authors:  Marek Milewski; Chen Koon Jaryl Ng; Pawel Gąsior; Shaoliang Shawn Lian; Su Xiao Qian; Shengjie Lu; Nicolas Foin; Elvin Kedhi; Wojciech Wojakowski; Hui Ying Ang
Journal:  Polymers (Basel)       Date:  2022-04-22       Impact factor: 4.967

2.  Research on a Nonwoven Fabric Made from Multi-Block Biodegradable Copolymer Based on l-Lactide, Glycolide, and Trimethylene Carbonate with Shape Memory.

Authors:  Joanna Walczak; Michał Chrzanowski; Izabella Krucińska
Journal:  Molecules       Date:  2017-08-10       Impact factor: 4.411

3.  Two-Step Geometry Design Method, Numerical Simulations and Experimental Studies of Bioresorbable Stents.

Authors:  Natalia Molęda; Grzegorz Kokot; Wacław Kuś; Michał Sobota; Jakub Włodarczyk; Mateusz Stojko
Journal:  Materials (Basel)       Date:  2022-03-24       Impact factor: 3.623

4.  Safety and efficacy of biodegradable polymer-coated thin strut sirolimus-eluting stent vs. durable polymer-coated everolimus-eluting stent in patients with acute myocardial infarction.

Authors:  Paweł Gąsior; Marek Gierlotka; Krzysztof Szczurek-Katanski; Marcin Osuch; Magda Roleder; Michal Hawranek; Mariusz Gasior; Wojciech Wojakowski; Lech Polonski
Journal:  Postepy Kardiol Interwencyjnej       Date:  2018-11-09       Impact factor: 1.426

  4 in total