Literature DB >> 25931061

The X-ray Correlation Spectroscopy instrument at the Linac Coherent Light Source.

Roberto Alonso-Mori1, Chiara Caronna1, Matthieu Chollet1, Robin Curtis1, Daniel S Damiani1, Jim Defever1, Yiping Feng1, Daniel L Flath1, James M Glownia1, Sooheyong Lee1, Henrik T Lemke1, Silke Nelson1, Eric Bong1, Marcin Sikorski1, Sanghoon Song1, Venkat Srinivasan1, Daniel Stefanescu1, Diling Zhu1, Aymeric Robert1.   

Abstract

The X-ray Correlation Spectroscopy instrument is dedicated to the study of dynamics in condensed matter systems using the unique coherence properties of free-electron lasers. It covers a photon energy range of 4-25 keV. The intrinsic temporal characteristics of the Linac Coherent Light Source, in particular the 120 Hz repetition rate, allow for the investigation of slow dynamics (milliseconds) by means of X-ray photon correlation spectroscopy. Double-pulse schemes could probe dynamics on the picosecond timescale. A description of the instrument capabilities and recent achievements is presented.

Entities:  

Keywords:  FEL; XPCS; coherent scattering; hard X-ray

Year:  2015        PMID: 25931061      PMCID: PMC4416668          DOI: 10.1107/S1600577515004397

Source DB:  PubMed          Journal:  J Synchrotron Radiat        ISSN: 0909-0495            Impact factor:   2.616


Introduction

The Linac Coherent Light Source (LCLS), a US Department of Energy Office of Science user facility operated by Stanford University, achieved first light in 2009 (Emma et al., 2010 ▶). The X-ray Correlation Spectroscopy (XCS) instrument began operations in 2011; located in the Far Experimental Hall (FEH), XCS was the fifth LCLS instrument to become operational. In contrast with synchrotron storage ring sources, which provide high-brilliance beam with partial coherence (Grübel et al., 2008 ▶), LCLS provides pulsed transversely coherent hard X-rays with unprecedented flux and short pulse duration. These characteristics enable the investigation of dynamics in condensed matter by measuring time-resolved coherent scattering patterns (i.e. time-resolved speckles). This can be achieved by means of X-ray Photon Correlation Spectroscopy (XPCS) (Grübel et al., 2008 ▶), where the temporal evolution of speckle patterns can be quantified by calculating intensity autocorrelation functions and extracting typical relaxation times for specific length scales (Robert, 2007 ▶). The LCLS repetition rate, 120 Hz, limits the fastest measurable time scales to several tens of milliseconds (Stephenson et al., 2009 ▶). More elaborate schemes, requiring the measurement of the sum of two speckle patterns originating from a sequence of two X-ray pulses separated in time (Gutt et al., 2009 ▶), allow the measurement of much faster dynamics. In that case the accessible time scales are limited by the ability to generate two pulses separated in time by and can potentially reach the femtosecond regime. The XCS instrument was designed to provide an optimum platform for taking advantage of the unique coherence properties of the LCLS (e.g. the possibility to access very large sample–detector distances) while still accommodating experiments using different X-ray techniques (small-angle X-ray scattering, diffraction, X-ray spectroscopy and imaging) (Robert et al., 2013 ▶). The addition of an ultrafast laser system will not only allow ‘standard’ optical pump/X-ray probe experiments but will open new opportunities to use coherent scattering techniques in combination with ultrafast optical excitation.

Instrument overview

The XCS instrument operates in the hard X-ray regime (above 4 keV). A set of silicon mirrors (coated with silicon carbide), with an incidence angle of 1.32 mrad, are located in the front-end enclosure of the Near Experimental Hall (NEH), which feeds the hard X-ray beam to all LCLS hard X-ray hutches. These limit the maximum photon energy that can be delivered to the LCLS hard X-ray hutches to 25 keV. The first components of the XCS instrument are located in the X-ray Transport Tunnel, a 200 m-long tunnel connecting the Near and Far Experimental Halls. These include slits and diagnostics located 196 m upstream from the sample, as indicated in Fig. 1 ▶.
Figure 1

Overview of the XCS instrument layout. Distances are indicated in meters from the center of the diffractometer. S&D: slits and non-destructive intensity diagnostics; DCM: large-offset double-crystal monochromator; CCM: channel-cut monochromator; TT: time-tool measuring the arrival time of the optical laser with reference to the X-rays; M1/M2: silicon mirrors that can be used to deflect the beam in the vertical direction and can also provide harmonic rejection; L-IN: laser in-coupling for the optical laser. Components located downstream of the dashed line can be translated into the main LCLS line and allow the XCS instrument to take advantage of the full power and properties of the fundamental. The sample at the XCS instrument is located approximately 420 m from the source.

In the following, we describe the specifications of key components of the XCS instrument. Monochromators. The XCS instrument has two distinct configurations. It can operate in the LCLS main line and take full advantage of the first-harmonic properties of LCLS and its various operation modes (SASE, two-color, etc.). To operate in this ‘pink beam’ configuration, all components located downstream of the dashed line in Fig. 1 ▶ are translated to the LCLS main line. For experiments that require a monochromatic beam [i.e. a larger longitudinal coherence length = = ] but do not intend to scan broadly the incident energy, a custom-built (JJ X-ray, Denmark) large-offset Double-Crystal Monochromator (DCM) (Zhu et al., 2014 ▶) is located 44 m upstream of the sample. In this configuration all components located downstream of the dashed line are translated 600 mm horizontally as displayed in Fig. 1 ▶. The DCM operates with Si(111) crystals and provides a monochromaticity of = 1.4 × 10−4 at 8 keV. If more monochromaticity is required, a Si(511) Channel-Cut Monochromator (CCM) (Narayanan et al., 2008 ▶) operating in the vertical scattering geometry provides an energy resolution of = 8.9 × 10−6. Future plans include an upgrade of the DCM crystals to (111), which not only provides better energy resolution ( = 5.3 × 10−5) but most importantly allows multiplexing with another instrument located downstream in the FEH (Feng et al., 2013 ▶, 2015 ▶; Zhu et al., 2014 ▶). A detailed discussion concerning the effect of the monochromaticity on the intensity and longitudinal coherence properties of a FEL SASE beam is provided by Lee et al. (2012) ▶. X-ray focusing lenses. The unfocused beam at XCS is typically 0.75 mm × 0.75 mm FWHM in size. Beryllium compound refractive lenses (RWTH, Aachen, Germany) allow focusing of the beam, and provide beam size control in one or two dimensions (Snigirev et al., 1996 ▶). The lenses are located 3.3 m upstream of the sample and can be adjusted ±150 mm along the incident beam axis. The minimum spot size was measured to be ∼2 µm × 2 µm (limited by the finite SASE bandwidth and imperfections in the optics). A one-dimensional focusing option allows delivery of a line focus for elongated samples or for applications such as grazing-incidence experiments. Pulse picker. A fast shutter is available for selecting single X-ray pulses from the LCLS pulse train on demand, as well as for reducing the repetition rate. It consists of a channel which is rotated back and forth in order to create a brief opening time. It can be used to create arbitrary pulse train time patterns provided that the pulse train structure has an average rate of equal to or less than 10 Hz. Mirrors. Two silicon mirrors located 1.5 and 2 m upstream of the sample allow delivery of the beam with a vertical grazing angle (as, for example, is required for liquid surfaces). These can also reduce the third-harmonic content of the LCLS beam by operating above the critical angle for these energies. Diffractometer. A custom-built horizontal geometry four-circle diffractometer (Huber, Germany) is available and enables precise orientation of samples or sample environments such as vacuum chambers, gas/liquid injectors, etc. Its sphere of confusion is better than 20 µm. It can be used in conjunction with the auxiliary 2θ arm (FMB Oxford, UK) providing large sample–detector distances (4 and 7.5 m) and covering scattering angles up to 55° in the horizontal plane. Additional diagnostics. The SASE process induces pulse-to-pulse fluctuations of the beam properties, such as pulse energy, duration, spatial profile, wavefront, temporal profile and spectral content. In situ pulse property monitoring is thus crucial for data interpretation. Multiple intensity monitors are installed at various locations along the instrument for pulse-to-pulse intensity normalization (Feng et al., 2011 ▶). The spatial profile of the beam can also be measured at various locations along the instrument using scintillating screens with high-resolution camera–lens combinations. Detectors. Several X-ray detectors are available and integrated with the XCS data acquisition system. These have various characteristics (pixel size, number of pixel, noise, frame rate and dynamic range) which are evaluated in order to identify the most suitable detector for specific experimental needs. Coherent X-ray experiments, for example, typically require small pixel size, very low noise, moderate dynamic range and a large number of pixels. This can be achieved with the 20 µm pixel size direct illumination Princeton CDD, but at a very low frame rate. A new detector is currently being developed at LCLS with 55 µm pixels and low noise, but running at the 120 Hz full repetition rate (Dragone et al., 2014 ▶). For more information about the LCLS detectors, see Blaj et al. (2015) ▶. Split and delay. The XCS instrument has space allocated for instrumentation to generate double-pulse X-ray patterns with a controlled typically below 1 ns. A split and delay prototype built by DESY (Hamburg, Germany) is currently installed and its performances are described elsewhere (Roseker et al., 2009 ▶, 2012 ▶). Other prototypes offering different beam properties are being evaluated or tested. Ultrafast laser capabilities will be added to the XCS instrument in 2015. These include the construction of a dedicated laser hutch in close proximity to XCS, a standard ultrafast laser system, and timing diagnostics; the specifications of each are listed below: Optical laser system. Core laser systems at the LCLS consist of an ultrashort-pulse Ti:sapphire oscillator synchronized to the FEL seeding a commercially available chirped pulse amplifier producing 4 mJ at 40 fs. An additional four-pass amplifier, developed in-house, can boost the pulse energy to over 30 mJ. Wavelength conversion can cover a broad spectral range from 200 nm to 150 µm (∼1500 to 2 THz). A more thorough description of the optical laser capabilities at LCLS can be found by Minitti et al. (2015 ▶). Timing diagnostics. Typical phase locking between the accelerator and the laser system only holds the temporal jitter between the two sources to about 200 fs FWHM. In order to take full advantage of the short pulses and reach pulse-length-limited time resolution, diagnostics to measure the relative arrival time between laser and X-ray pulses have been developed. These are based on the X-ray induced change in refractive index of a thin target probed by a chirped broadband white-light continuum pulse derived from the optical laser. The optical light transmission change is resolved by an optical spectrometer for each pulse. Typical target materials are silicon nitride (Si3N4) or Ce:YAG crystals of different thicknesses to accommodate different beam conditions (Lemke et al., 2013 ▶). A summary of the XCS instrument parameters is given in Table 1 ▶.
Table 1

X-ray parameters and capabilities of the XCS instrument

Instrument nameXCS
Mirrors, incidence angle2 SiC on Si, 1.32mrad
Monochromaticity () 1 103 (SASE), 2 104 (seeding)
Energy range (keV)4 to 11 (fundamental)
Unfocused beam size (m)750 at 8.3keV
Focused beam size (m)2750
Focusing opticsBe lenses, 1D and 2D focusing
Flux (photons pulse1)1 1012 (fundamental)
Pulse length (fs)5200
Repetition rate (Hz)120, 60, 30, 10, 5, 1, on demand
Optical laser pulse energy (mJ)§ 20 (800nm), 45 (400nm), 1 (266nm)
Optical laser pulse length (fs)§ 10150
Standard detectorsCSPAD, CSPAD-140k, ePix
 Princeton
Sample environmentHuber horizontal four-circle diffractometer, general purpose vacuum, liquid jet, He enclosure, Oxford LN2 cryojet down to 100K
2 arm capabilitiesLarge sampledetector distance arm (4 and 7.5m) covering 0 2 55

Typical single-shot value.

Excluding beamline and instrument transmission.

To be installed in 2015.

Highlights

The XCS instrument focuses on measuring time-resolved coherent scattering patterns from condensed matter systems from which typical relaxation rates are deduced. These measurements take full advantage of the transverse coherence properties of the LCLS beam. Fig. 2 ▶ displays a typical single-shot speckle pattern from a static sample consisting of dried silica colloidal 150 nm diameter spheres (Kisker GmbH) with 8.3 keV X-rays. The detector is a CSPAD (Blaj et al., 2015 ▶) located 7.5 m downstream of the sample. The concentric excess scattering signals observed at Q ≃ 0.005 Å−1 and Q ≃ 0.01 Å−1 are typical small-angle scattering form-factor features related to the size and shape of the colloidal particles. As observed, the speckles (coherent scattering pattern appearing as the grainy features decorating the structure rings) are well developed.
Figure 2

Single-shot speckle pattern measured at 8.3 keV from 150 nm colloidal spheres. The dark blue areas are gaps between the CSPAD tiled sensors. The central aperture allows the transmitted beam to pass through, and therefore does not require a beamstop.

In order to characterize the transverse coherence of the beam, a series of speckle patterns are measured with and without the XCS monochromator, in the SASE operation mode, and analyzed in terms of photon statistics to determine their contrast, i.e. a measure of the transverse coherence in the small-angle regime (Gutt et al., 2012 ▶; Hruszkewycz et al., 2012 ▶; Lee et al., 2013 ▶; Lehmkühler et al., 2014 ▶). The intensities from a narrow iso-Q area consisting of an annulus centered at Q = 0 with a radius Q are histogrammed as displayed in Fig. 3 ▶ for Q = 0.0067 Å−1. These are then modeled using the negative-binomial distribution function (Mandel, 1959 ▶; Goodman, 2007 ▶)where I is the number of photons, is the mean number of photons in that area and M is the number of modes. M is related to the contrast of the speckle pattern C = . An example of the result of such analysis is displayed in Fig. 3 ▶ (inset). For that specific shot the fit to the experimental data yields a mean number ≃ 5.1 photons, mode number = 2.75 corresponding to a contrast = 0.6. The fit reproduces well the experimentally measured single-shot intensity distribution.
Figure 3

Single-shot speckle contrast measured at Q = 0.0067 Å−1 for various consecutive shots. Inset: probability density function of intensity within part of the speckle pattern corresponding to a wavevector Q = 0.0067 Å−1. The solid line represents the gamma distribution with number of modes M = 2.75 and average count rate ≃ 5.1 photons.

Because the mean number of photons is large one can also use a simpler formulation of the speckle contrast C = , where σ is the standard deviation of the measured intensities. This analysis was performed on successive shots at the same Q = 0.0067 Å−1. The results are displayed in Fig. 3 ▶. The observed contrast fluctuates between 0.6 and 0.8 with a mean contrast of 0.69. The shot-to-shot fluctuations of the contrast originate from the fine structures in the energy spectrum of the SASE beam, as simulated and described by Lee et al. (2012) ▶. The large degree of coherence of the LCLS beam makes it a suitable source to perform XPCS, measuring time-resolved speckle patterns from a sample. As the sample presents some dynamics, speckle patterns fluctuate in time. By means of time-autocorrelation functions calculated from the speckle patterns, information on the characteristic relaxation times can be obtained (Robert, 2007 ▶), a measure of the normalized intermediate scattering function of the system. In some cases the sample can also undergo non-equilibrium dynamical phenomena, referred to as aging (Robert et al., 2006 ▶), from which clear signatures can be observed in a time-dependent analysis of the degree of correlation of speckle patterns. This was recently investigated at XCS by Carnis et al. (2014) ▶, where the relaxation and aging dynamics of thin polymer films were investigated. One should also note that with a FEL the spatial position of the beam jitters shot-to-shot. Because of this effect the contrast β, usually referred to in second-order correlation functions (Robert, 2007 ▶), will be smaller than the single-shot speckle contrast, as discussed by Carnis et al. (2014 ▶). Time-resolved speckle spectroscopy at an FEL is intrinsically limited by the repetition rate of the source, as it relies on the time-correlation of two recorded coherently scattered patterns originating from two pulses separated in time. For the LCLS, the fastest dynamics that can be reached is of the order of 8.3 ms (i.e. 120 Hz repetition rate). A natural way of accessing faster timescales is to increase the repetition rate of the source. This would require a completely different accelerator technology such as the one planned to be used at the European XFEL (Grübel et al., 2007 ▶), where series of X-ray bunches separated by a minimum of 220 ns will be generated. To reach even faster timescales down to the picosecond regime and below, other possibilities involve generating two sub-pulses with a time separation between tens of femtoseconds up to a nanosecond as illustrated in Fig. 4 ▶. Each sub-pulse generates a speckle pattern. Current X-ray detectors are, however, only capable of measuring the sum of the scattered signal of both sub-pulses. If dynamics occurs on timescales of , the summed speckle patterns will present a decay of contrast, as compared with the contrast of a single speckle pattern. By adjusting the time delay between the two sub-pulses, one can extract the time evolution of the summed speckle pattern contrast, from which the normalized intermediate scattering function is obtained (Gutt et al., 2009 ▶), and therefore gain information about the underlying dynamics of the system.
Figure 4

Schematic of the double-pulse scheme. A single pulse is split into two sub-pulses; one is delayed relative to the other by an increase in its pathlength. The sub-pulses, separated in time by , are then redirected on a common trajectory to the sample. A typical pathlength difference of 1 mm corresponds to a time delay of 3 ps.

Recent results from the commissioning of a split and delay prototype (Roseker et al., 2012 ▶) based on eight perfect Bragg crystals [including two thin crystals that can act as a beam splitter/recombiner (Roseker et al., 2009 ▶)] have shown the possibility to reach time delays of the order of a nanosecond. Other technical developments involving split and delayed beams have also been investigated. This is particularly the case with a single-shot split and multiple-delay system, offering the possibility to probe ultrafast dynamics by means of X-ray pump/X-ray probe experiments, as recently demonstrated by David et al. (2015) ▶.

Conclusion

The XCS instrument takes full advantage of the large number of transversely coherent photons per pulse at the LCLS. XCS is a versatile tool for performing time-resolved coherent scattering experiments in the hard X-ray regime from which fundamental dynamics in condensed matter ordered and disordered systems can be explored. This can be achieved by means of XPCS for slow dynamics and will further be extended to ultrafast timescales by double-pulse experiments. More details about the XCS instrument can be found on the following website: http://lcls.slac.stanford.edu/xcs.

Facility access

LCLS instruments are open to academia, industry, government agencies and research institutes worldwide for scientific investigations. There are two calls for proposals per year and an external peer-review committee evaluates proposals based on scientific merit and instrument suitability. Access is without charge for users who intend to publish their results. Prospective users are encouraged to contact instrument staff members to learn more about the science and capabilities of the facility, and opportunities for collaboration.
  14 in total

1.  High wavevector temporal speckle correlations at the Linac Coherent Light Source.

Authors:  Sooheyong Lee; Wojciech Roseker; Christian Gutt; Zhirong Huang; Yuantao Ding; Gerhard Grübel; Aymeric Robert
Journal:  Opt Express       Date:  2012-04-23       Impact factor: 3.894

2.  Single shot spatial and temporal coherence properties of the SLAC Linac Coherent Light Source in the hard x-ray regime.

Authors:  C Gutt; P Wochner; B Fischer; H Conrad; M Castro-Colin; S Lee; F Lehmkühler; I Steinke; M Sprung; W Roseker; D Zhu; H Lemke; S Bogle; P H Fuoss; G B Stephenson; M Cammarata; D M Fritz; A Robert; G Grübel
Journal:  Phys Rev Lett       Date:  2012-01-11       Impact factor: 9.161

3.  Design and performance of an ultra-high-vacuum-compatible artificial channel-cut monochromator.

Authors:  Suresh Narayanan; Alec Sandy; Deming Shu; Michael Sprung; Curt Preissner; Joseph Sullivan
Journal:  J Synchrotron Radiat       Date:  2007-12-18       Impact factor: 2.616

4.  Single shot speckle and coherence analysis of the hard X-ray free electron laser LCLS.

Authors:  Sooheyong Lee; W Roseker; C Gutt; B Fischer; H Conrad; F Lehmkühler; I Steinke; D Zhu; H Lemke; M Cammarata; D M Fritz; P Wochner; M Castro-Colin; S O Hruszkewycz; P H Fuoss; G B Stephenson; G Grübel; A Robert
Journal:  Opt Express       Date:  2013-10-21       Impact factor: 3.894

5.  X-ray spectroscopy: Revealing the atomic dance.

Authors:  G Brian Stephenson; Aymeric Robert; Gerhard Grübel
Journal:  Nat Mater       Date:  2009-09       Impact factor: 43.841

6.  Performance of a picosecond x-ray delay line unit at 8.39 keV.

Authors:  Wojciech Roseker; Hermann Franz; Horst Schulte-Schrepping; Anita Ehnes; Olaf Leupold; Federico Zontone; Aymeric Robert; Gerhard Grübel
Journal:  Opt Lett       Date:  2009-06-15       Impact factor: 3.776

7.  Demonstration of simultaneous experiments using thin crystal multiplexing at the Linac Coherent Light Source.

Authors:  Y Feng; R Alonso-Mori; T R M Barends; V D Blank; S Botha; M Chollet; D S Damiani; R B Doak; J M Glownia; J M Koglin; H T Lemke; M Messerschmidt; K Nass; S Nelson; I Schlichting; R L Shoeman; Yu V Shvyd'ko; M Sikorski; S Song; S Stoupin; S Terentyev; G J Williams; D Zhu; A Robert; S Boutet
Journal:  J Synchrotron Radiat       Date:  2015-04-10       Impact factor: 2.616

Review 8.  Optical laser systems at the Linac Coherent Light Source.

Authors:  Michael P Minitti; Joseph S Robinson; Ryan N Coffee; Steve Edstrom; Sasha Gilevich; James M Glownia; Eduardo Granados; Philippe Hering; Matthias C Hoffmann; Alan Miahnahri; Despina Milathianaki; Wayne Polzin; Daniel Ratner; Franz Tavella; Sharon Vetter; Marc Welch; William E White; Alan R Fry
Journal:  J Synchrotron Radiat       Date:  2015-04-22       Impact factor: 2.616

9.  X-ray detectors at the Linac Coherent Light Source.

Authors:  Gabriel Blaj; Pietro Caragiulo; Gabriella Carini; Sebastian Carron; Angelo Dragone; Dietrich Freytag; Gunther Haller; Philip Hart; Jasmine Hasi; Ryan Herbst; Sven Herrmann; Chris Kenney; Bojan Markovic; Kurtis Nishimura; Shawn Osier; Jack Pines; Benjamin Reese; Julie Segal; Astrid Tomada; Matt Weaver
Journal:  J Synchrotron Radiat       Date:  2015-04-21       Impact factor: 2.616

10.  Single shot coherence properties of the free-electron laser SACLA in the hard X-ray regime.

Authors:  Felix Lehmkühler; Christian Gutt; Birgit Fischer; Martin A Schroer; Marcin Sikorski; Sanghoon Song; Wojciech Roseker; James Glownia; Mathieu Chollet; Silke Nelson; Kensuke Tono; Tetsuo Katayama; Makina Yabashi; Tetsuya Ishikawa; Aymeric Robert; Gerhard Grübel
Journal:  Sci Rep       Date:  2014-06-10       Impact factor: 4.379

View more
  12 in total

1.  Ideal charge-density-wave order in the high-field state of superconducting YBCO.

Authors:  H Jang; W-S Lee; H Nojiri; S Matsuzawa; H Yasumura; L Nie; A V Maharaj; S Gerber; Y-J Liu; A Mehta; D A Bonn; R Liang; W N Hardy; C A Burns; Z Islam; S Song; J Hastings; T P Devereaux; Z-X Shen; S A Kivelson; C-C Kao; D Zhu; J-S Lee
Journal:  Proc Natl Acad Sci U S A       Date:  2016-12-05       Impact factor: 11.205

2.  Demonstration of simultaneous experiments using thin crystal multiplexing at the Linac Coherent Light Source.

Authors:  Y Feng; R Alonso-Mori; T R M Barends; V D Blank; S Botha; M Chollet; D S Damiani; R B Doak; J M Glownia; J M Koglin; H T Lemke; M Messerschmidt; K Nass; S Nelson; I Schlichting; R L Shoeman; Yu V Shvyd'ko; M Sikorski; S Song; S Stoupin; S Terentyev; G J Williams; D Zhu; A Robert; S Boutet
Journal:  J Synchrotron Radiat       Date:  2015-04-10       Impact factor: 2.616

Review 3.  Optical laser systems at the Linac Coherent Light Source.

Authors:  Michael P Minitti; Joseph S Robinson; Ryan N Coffee; Steve Edstrom; Sasha Gilevich; James M Glownia; Eduardo Granados; Philippe Hering; Matthias C Hoffmann; Alan Miahnahri; Despina Milathianaki; Wayne Polzin; Daniel Ratner; Franz Tavella; Sharon Vetter; Marc Welch; William E White; Alan R Fry
Journal:  J Synchrotron Radiat       Date:  2015-04-22       Impact factor: 2.616

4.  X-ray detectors at the Linac Coherent Light Source.

Authors:  Gabriel Blaj; Pietro Caragiulo; Gabriella Carini; Sebastian Carron; Angelo Dragone; Dietrich Freytag; Gunther Haller; Philip Hart; Jasmine Hasi; Ryan Herbst; Sven Herrmann; Chris Kenney; Bojan Markovic; Kurtis Nishimura; Shawn Osier; Jack Pines; Benjamin Reese; Julie Segal; Astrid Tomada; Matt Weaver
Journal:  J Synchrotron Radiat       Date:  2015-04-21       Impact factor: 2.616

5.  The Linac Coherent Light Source.

Authors:  William E White; Aymeric Robert; Mike Dunne
Journal:  J Synchrotron Radiat       Date:  2015-04-21       Impact factor: 2.616

6.  Direct experimental observation of the gas density depression effect using a two-bunch X-ray FEL beam.

Authors:  Y Feng; D W Schafer; S Song; Y Sun; D Zhu; J Krzywinski; A Robert; J Wu; F J Decker
Journal:  J Synchrotron Radiat       Date:  2018-01-01       Impact factor: 2.616

7.  Hard X-rays as pump and probe of atomic motion in oxide glasses.

Authors:  B Ruta; F Zontone; Y Chushkin; G Baldi; G Pintori; G Monaco; B Rufflé; W Kob
Journal:  Sci Rep       Date:  2017-06-21       Impact factor: 4.379

8.  Towards ultrafast dynamics with split-pulse X-ray photon correlation spectroscopy at free electron laser sources.

Authors:  W Roseker; S O Hruszkewycz; F Lehmkühler; M Walther; H Schulte-Schrepping; S Lee; T Osaka; L Strüder; R Hartmann; M Sikorski; S Song; A Robert; P H Fuoss; M Sutton; G B Stephenson; G Grübel
Journal:  Nat Commun       Date:  2018-04-27       Impact factor: 14.919

9.  Sequential Single Shot X-ray Photon Correlation Spectroscopy at the SACLA Free Electron Laser.

Authors:  Felix Lehmkühler; Paweł Kwaśniewski; Wojciech Roseker; Birgit Fischer; Martin A Schroer; Kensuke Tono; Tetsuo Katayama; Michael Sprung; Marcin Sikorski; Sanghoon Song; James Glownia; Matthieu Chollet; Silke Nelson; Aymeric Robert; Christian Gutt; Makina Yabashi; Tetsuya Ishikawa; Gerhard Grübel
Journal:  Sci Rep       Date:  2015-11-27       Impact factor: 4.379

10.  Pulse intensity characterization of the LCLS nanosecond double-bunch mode of operation.

Authors:  Yanwen Sun; Franz Josef Decker; James Turner; Sanghoon Song; Aymeric Robert; Diling Zhu
Journal:  J Synchrotron Radiat       Date:  2018-03-27       Impact factor: 2.616

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.