Isak Silander, Thomas Hausmaninger, Weiguang Ma, Patrick Ehlers, Ove Axner.
Abstract
We report on the realization of a Doppler-broadened noise-immune cavity-enhanced optical heterodyne molecular spectrometry (NICE-OHMS) instrumentation based on a high-finesse (50,000) cavity with a detection sensitivity of 4×10(-13) cm(-1) Hz(-1/2). For the P(e)(11) transition targeted at 1.5316 μm, this corresponds to a C(2)H(2) concentration of 240 ppq (parts-per-quadrillion) detected at 100 Torr. The setup was originally affected by recurrent dips in the cavity transmission, which were attributed to excitation of high-order transverse mode by scattering from the mirrors. The effect of these was reduced by insertion of a small pinhole in the cavity.Year: 2015 PMID: 25927769 DOI: 10.1364/OL.40.002004