OBJECTIVES: Conversion of a previous endovascular aneurysm repair (EVAR) with suprarenal fixation is a challenging situation even in the elective setting. The outcomes of a technique based on preservation of the first proximal covered stent of the endograft, used as a "neo-neck" for proximal anastomosis, are presented.
METHODS: From 2001 to 2014, nine patients underwent elective conversion of a previous suprarenally fixed EVAR. After supraceliac clamping, the aneurysm sac was opened and the endograft identified; the fabric was cut beyond the first covered stent together with its native aortic wall in order to create a "neo-neck." An aortic balloon was inflated into the visceral aorta to avoid back bleeding. A Dacron bifurcated tube graft (Intergard, Maquet) was then sutured to the neo-neck mimicking endobanding, passing the stitches into the aortic wall and the first covered stent.
RESULTS: The mean age was 68 years (range, 52-84 years). The stent grafts removed were four Zenith (Cook Medical), three Endurant (Medtronic), and two E-vita (Jotec). The indication for conversion was type 1A (n = 2), type 2 (n = 2), and type 3 (n = 1) endoleak, complete endograft thrombosis (n = 2), and abdominal pain with sac enlargement with no radiological sign of endoleak (n = 2). Blood loss was 1,428 mL (range 500-3,000 mL); the visceral ischemic time to perform the proximal anastomosis was 23.5 min ± 2.3 min). The post-operative complication rate was 11% (n = 1/9) related to a case of sac wall bleeding requiring re-intervention; mortality at 30 days was 0%. At 22 months (range, 8-41) the computed tomography angiogram demonstrated no signs of leaks or anastomotic pseudoaneurysm.
CONCLUSION: Preservation of the proximal covered stent of an endograft with suprarenal fixation used as an infrarenal "neo-neck" with incorporation of the aorta to the suture line during elective surgical explantation simplifies the procedure, and can be achieved with very low early morbidity and mortality; furthermore, it seems to be durable over mid-term follow up.
Copyright © 2015 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  EVAR conversion; EVAR explantation; Failed EVAR; Hybrid aortic reconstruction

Mesh:

Year:  2015        PMID: 25920632     DOI: 10.1016/j.ejvs.2015.03.027

Source DB:  PubMed          Journal:  Eur J Vasc Endovasc Surg        ISSN: 1078-5884            Impact factor:   7.069


  4 in total

1.  Unique Technique for Open Surgical Repair after Failed Endovascular Aneurysm Repair with Proximal Anastomoses.

Authors:  Satoshi Takebayashi; Jun Hirota; Kazuki Mori; Takashi Shuto; Keitaro Okamoto; Aiko Sato; Tomoyuki Wada; Hirofumi Anai; Shinji Miyamoto
Journal:  Ann Vasc Dis       Date:  2016-04-25

2.  Outcomes of Late Open Conversion after Endovascular Abdominal Aneurysm Repair.

Authors:  Yoshikatsu Nomura; Kanetsugu Nagao; Shota Hasegawa; Motoharu Kawashima; Takanori Tsujimoto; So Izumi; Masamichi Matsumori; Hiroshi Tanaka; Hirohisa Murakami; Tasuku Honda; Ryota Kawasaki; Nobuhiko Mukohara
Journal:  Ann Vasc Dis       Date:  2019-09-25

3.  Technique of partial open surgical stent graft explantation with preservation of fenestrated stent graft component to treat recalcitrant type II endoleak.

Authors:  Jessica A Steadman; Bernardo C Mendes; Gustavo S Oderich
Journal:  J Vasc Surg Cases Innov Tech       Date:  2022-07-20

4.  A rapid aneurysmal formation after late open conversion of endovascular abdominal aortic repair with complete endograft explant.

Authors:  Yuki Tamagawa; Masashi Kawamura; Masahiro Ryugo; Osamu Monta; Yasushi Tsutsumi
Journal:  J Surg Case Rep       Date:  2021-06-30
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.