Literature DB >> 25917552

A randomized multi-institutional crossover comparison of the GlideScope® Cobalt Video laryngoscope to the flexible fiberoptic bronchoscope in a Pierre Robin manikin.

John E Fiadjoe1, Matthew Hirschfeld1, Stephan Wu1, James Markley1, Harshad Gurnaney1, Abbas F Jawad1, Paul Stricker1, Todd Kilbaugh1, Patrick Ross2, Pete Kovatsis3.   

Abstract

BACKGROUND: The GlideScope Cobalt Video laryngoscope is being used more often in children with challenging laryngoscopy. There are, however, no pediatric trials comparing it to flexible fiberoptic bronchoscopy, the current accepted gold standard. This preliminary manikin study compares the first-attempt intubation success of the GlideScope Cobalt video laryngoscope to the flexible fiberoptic bronchoscope when performed by attending pediatric anesthesiologists at two major pediatric centers.
METHODS: This prospective randomized, crossover study evaluated 120 attempts (60 with each study device) to intubate the AirSim Pierre Robin manikin (PRM) with fiberoptic bronchoscopy and video laryngoscopy (VL). Attending pediatric anesthesiologists from two quaternary pediatric centers were eligible to participate. Each attending anesthesiologist randomly performed a single tracheal intubation attempt with one of the study devices followed by the alternate method. The primary outcome was the first-attempt success rate of tracheal intubation. Blinding was not feasible. We hypothesized that first-attempt success would be higher with fiberoptic bronchoscopy.
RESULTS: Thirty anesthesiologists from each center were randomized to use one of the study devices followed by the alternate method. We analyzed all participants' data. There was no overall difference in first-attempt success between VL and fiberoptic bronchoscopy (88.3% vs 85% respectively, P = 0.59). There were significant institutional differences in first-attempt success using VL (76.7% vs 100%).
CONCLUSIONS: There was no difference in first-attempt success of tracheal intubation using VL vs fiberoptic bronchoscopy when performed by attending anesthesiologists at two large pediatric centers. However, institutional differences exist in success rates with VL across the two centers. Results from single-center device evaluations should be verified by multi-center evaluations. A significant proportion of attending anesthesiologists lack experience with advanced airway devices; targeted education may enhance intubation success and patient safety.
© 2015 John Wiley & Sons Ltd.

Entities:  

Keywords:  difficult airway; difficult laryngoscopy; flexible fiberoptic intubation; tracheal intubation infant; video laryngoscopy

Mesh:

Year:  2015        PMID: 25917552     DOI: 10.1111/pan.12668

Source DB:  PubMed          Journal:  Paediatr Anaesth        ISSN: 1155-5645            Impact factor:   2.556


  1 in total

1.  Comparison of Macintosh Laryngoscope and GlideScope® for Orotracheal Intubation in Children Older Than One Year.

Authors:  Leyla Kılınç; Ayşe Surhan Çınar
Journal:  Sisli Etfal Hastan Tip Bul       Date:  2019-06-24
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.