We report a case where migrated Onyx could be removed safely with a mechanical thrombectomy device after withdrawal of the damaged microcatheter. A microcatheter was retained in the superior cerebellar artery with attached Onyx fragments on its tip during embolization of a bled cerebellar AVM. During retrieval maneuvers microcatheter shaft was ruptured and Onyx fragments were released into the vertebrobasilar system causing occlusion. Finally, Onyx fragment on tip of the microcatheter could be detached with dimethyl sulfoxide injection and the microcatheter could be withdrawn. A Merci retriever device (Concentric Medical, Mountain View, CA, USA) was successfully used to retrieve most of the dislodged Onyx material and recanalize the vertebrobasilar circulation.Entities: