| Literature DB >> 25909025 |
Dominik Spether1, Marcus Scharpf2, Jörg Hennenlotter3, Christian Schwentner3, Alexander Neugebauer4, Daniela Nüßle4, Klaus Fischer4, Hans Zappe1, Arnulf Stenzl3, Falko Fend2, Andreas Seifert1, Markus Enderle4.
Abstract
Complete surgical removal of cancer tissue with effective preservation of healthy tissue is one of the most important challenges in modern oncology. We present a method for real-time, in situ differentiation of tissue based on optical emission spectroscopy (OES) performed during electrosurgery not requiring any biomarkers, additional light sources or other excitation processes. The analysis of the optical emission spectra, enables the differentiation of healthy and tumorous tissue. By using multi-class support vector machine (SVM) algorithms, distinguishing between tumor types also seems to be possible. Due to its fast reaction time (0.05s) the method can be used for real-time navigation helping the surgeon achieve complete resection. The system's easy realization has been proven by successful integration in a commercial electro surgical unit (ESU). In a first step the method was verified by using ex vivo tissue samples. The histological analysis confirmed, 95% of correctly classified tissue types.Entities:
Keywords: (170.1610) Clinical applications; (170.3890) Medical optics instrumentation; (170.6510) Spectroscopy, tissue diagnostics
Year: 2015 PMID: 25909025 PMCID: PMC4399680 DOI: 10.1364/BOE.6.001419
Source DB: PubMed Journal: Biomed Opt Express ISSN: 2156-7085 Impact factor: 3.732