Correlation between Central Corneal Thickness and Intraocular Pressure Measured by Goldmann Applanation Tonometry or Pascal Dynamic Contour Tonometry.

J M Katsimpris1, P E Theoulakis2, K Vasilopoulos1, G Skourtis1, G E Papadopoulos3, I K Petropoulos4.   

Abstract

PURPOSE: The aim of this study was to investigate the relationship between central corneal thickness and intraocular pressure measured by Goldmann applanation tonometry and Pascal dynamic contour tonometry. PATIENTS AND METHODS: The study included 45 persons (90 eyes), divided into 4 groups: a) 10 normal volunteers (20 eyes); b) 16 patients (32 eyes) with primary open-angle glaucoma; c) 8 patients (16 eyes) with normal-tension glaucoma; and d) 11 patients (22 eyes) with ocular hypertension. Intraocular pressure was measured by Goldmann applanation tonometry and Pascal dynamic contour tonometry, and central corneal thickness was measured by ultrasound pachymetry. The relationship between intraocular pressure and central corneal thickness was evaluated.
RESULTS: Intraocular pressure was correlated positively but not strongly enough with central corneal thickness when it was measured by Goldmann applanation tonometry. On the contrary, there was no correlation between intraocular pressure and central corneal thickness when intraocular pressure was measured by Pascal dynamic contour tonometry.
CONCLUSION: Central corneal thickness is an important variable in the evaluation of intraocular pressure by Goldmann applanation tonometry. This factor does not interfere with the intraocular pressure measurements taken by Pascal dynamic contour tonometry. Georg Thieme Verlag KG Stuttgart · New York.

Entities:  

Mesh:

Year:  2015        PMID: 25902089     DOI: 10.1055/s-0035-1545792

Source DB:  PubMed          Journal:  Klin Monbl Augenheilkd        ISSN: 0023-2165            Impact factor:   0.700


  2 in total

1.  Corrected Intraocular Pressure Variability with Central Corneal Thickness Measurement.

Authors:  Mohamed Attia Ali Ahmed; Ahmed Shawkat Abdelhalim
Journal:  Clin Ophthalmol       Date:  2020-12-24

2.  Intraocular Pressure and Its Relation to Ocular Geometry: Results From the Gutenberg Health Study.

Authors:  Esther M Hoffmann; Fidan Aghayeva; Felix M Wagner; Achim Fiess; Markus Nagler; Thomas Münzel; Philipp S Wild; Manfred E Beutel; Irene Schmidtmann; Karl J Lackner; Norbert Pfeiffer; Alexander K Schuster
Journal:  Invest Ophthalmol Vis Sci       Date:  2022-01-03       Impact factor: 4.799

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.