Comparison of the endotracheal tube cuff pressure between a tapered- versus a cylindrical-shaped cuff after changing from the supine to the lateral flank position.

Hyun-Chang Kim1, Yong-Hun Lee, Eugene Kim, Eun-Ah Oh, Young-Tae Jeon, Hee-Pyoung Park.   

Abstract

PURPOSE: Positional change can displace an endotracheal tube (ETT) and change the ETT cuff pressure in a tracheally intubated patient. Endotracheal tubes with different cuff shapes may lead to different cuff pressures after positional change. We hypothesized that the intracuff pressure in the TaperGuard™ ETT with a tapered-shaped cuff would be higher than that in the conventional ETT with a cylindrical-shaped cuff after a change from the supine to the lateral flank position.
METHODS: Fifty-eight patients scheduled for open urological procedures in the lateral flank position were randomly allocated to receive either a TaperGuard ETT (group T) or conventional ETT (group C). The ETT cuff pressure was initially set at 20 cm H2O in the supine position and was measured after the change to the lateral flank position. The distance from the ETT tip to the carina was measured in both the supine and the lateral flank positions.
RESULTS: Two patients, one from each group, were excluded from the data analysis. The mean (SD) ETT cuff pressure was significantly higher in group T (n = 28) than in group C (n = 28) after the change in position [31 (7) cm H2O vs 25 (4) cm H2O, respectively; mean difference, 6 cm; 95% confidence intervals [CI], 3 to 9; P < 0.001]. The mean (SD) proximal migration of the ETT tip was comparable between the two groups [8 (18) mm vs 4 (14) mm, respectively; P = 0.367].
CONCLUSIONS: After the change from the supine to the lateral flank position, the ETT cuff pressure was significantly higher in the TaperGuard ETT than in the conventional ETT, although the extent of cephalad displacement of the ETT was comparable between the two groups. This trial was registered at Clinicaltrials.gov: NCT02165319.

Entities:  

Mesh:

Year:  2015        PMID: 25894912     DOI: 10.1007/s12630-015-0394-z

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  7 in total

1.  Changes in tapered endotracheal tube cuff pressure after changing position to hyperextension of neck: A randomized clinical trial.

Authors:  Ji-Hoon Park; Hyo-Jin Lee; Sou Hyun Lee; Ji Seob Kim
Journal:  Medicine (Baltimore)       Date:  2021-07-23       Impact factor: 1.817

2.  A Tapered Cuff Tracheal Tube Decreases the Need for Cuff Pressure Adjustment After Surgical Retraction During Anterior Cervical Spine Surgery: A Randomized Controlled, Double-Blind Trial.

Authors:  Yi-Shiuan Li; Elise Chia-Hui Tan; Yueh-Ju Tsai; Mercedes Susan Mandell; Shiang-Suo Huang; Ting-Yun Chiang; Wen-Cheng Huang; Wen-Kuei Chang; Ya-Chun Chu
Journal:  Front Med (Lausanne)       Date:  2022-06-29

3.  Comparison of the cuff pressure of a TaperGuard endotracheal tube and a cylindrical endotracheal tube after lateral rotation of head during middle ear surgery: A single-blind, randomized clinical consort study.

Authors:  Eunkyung Choi; Yongmin Park; Younghoon Jeon
Journal:  Medicine (Baltimore)       Date:  2017-03       Impact factor: 1.889

4.  Effect of lateral positioning on the bronchial cuff pressure of a left-sided double-lumen endotracheal tube during thoracic surgery: study protocol for a prospective observational study.

Authors:  Eugene Kim; In-Young Kim; Sung-Hye Byun
Journal:  BMJ Open       Date:  2019-03-30       Impact factor: 2.692

5.  The influence of high-dose intraoperative remifentanil on postoperative sore throat: a prospective randomized study: A CONSORT compliant article.

Authors:  Ji-Hoon Park; Yong-Cheol Lee; Jiwon Lee; Hyunjae Kim; Hyun-Chang Kim
Journal:  Medicine (Baltimore)       Date:  2018-12       Impact factor: 1.817

6.  Influence of two-handed jaw thrust during tracheal intubation on postoperative sore throat: a prospective randomised study.

Authors:  Hyub Huh; Doo Yeon Go; Jang Eun Cho; Jihoon Park; Jiwon Lee; Hyun-Chang Kim
Journal:  J Int Med Res       Date:  2021-02       Impact factor: 1.671

7.  The effect of different inflating volume on the measurement accuracy of the modified cuff pressure measurement method.

Authors:  Lijun Xiang; Meng Cao; Xuemei Song; Miaoqin Tan; Taosheng Deng; Yuan Wang; Xiaomei Zhang
Journal:  J Clin Monit Comput       Date:  2021-03-11       Impact factor: 1.977

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.