Accuracy of continuous noninvasive arterial pressure monitoring in living-liver donors during transplantation.

Coskun Araz1, Pinar Zeyneloglu, Arash Pirat, Nukhet Veziroglu, Aynur Camkiran Firat, Gulnaz Arslan.   

Abstract

OBJECTIVES: Hemodynamic monitoring is vital during liver transplant surgeries because distinct hemodynamic changes are expected. The continuous noninvasive arterial pressure (CNAP) monitor is a noninvasive device for continuous arterial pressure measurement by a tonometric method. This study compared continuous noninvasive arterial pressure monitoring with invasive direct arterial pressure monitoring in living-liver donors during transplant.
MATERIALS AND METHODS: There were 40 patients analyzed while undergoing hepatic lobectomy for liver transplant. Invasive pressure monitoring was established at the radial artery and continuous noninvasive arterial pressure monitoring using a finger sensor was recorded simultaneously from the contralateral arm. Systolic, diastolic, and mean arterial pressures from the 2 methods were compared. Correlation between the 2 methods was calculated.
RESULTS: A total of 5433 simultaneous measurements were obtained. For systolic arterial blood pressure, 55% continuous noninvasive arterial pressure measurements were within 10% direct arterial measurement; the correlation was 0.479, continuous noninvasive arterial pressure bias was -0.3 mm Hg, and limits of agreement were 32.0 mm Hg. For diastolic arterial blood pressure, 50% continuous noninvasive arterial pressure measurements were within 10% direct arterial measurement; the correlation was 0.630, continuous noninvasive arterial pressure bias was -0.4 mm Hg, and limits of agreement were 21.1 mm Hg. For mean arterial blood pressure, 60% continuous noninvasive arterial pressure measurements were within 10% direct arterial measurement; the correlation was 0.692, continuous noninvasive arterial pressure bias was +0.4 mm Hg, and limits of agreement were 20.8 mm Hg.
CONCLUSIONS: The 2 monitoring techniques did not show acceptable agreement. Our results suggest that continuous noninvasive arterial pressure monitoring is not equivalent to invasive arterial pressure monitoring in donors during living-donor liver transplant.

Entities:  

Mesh:

Year:  2015        PMID: 25894178     DOI: 10.6002/ect.mesot2014.p140

Source DB:  PubMed          Journal:  Exp Clin Transplant        ISSN: 1304-0855            Impact factor:   0.945


  2 in total

1.  Approach to the High-Risk Cardiac Patient.

Authors:  Alice Luo; Christopher Abbott; Karyn Nunez
Journal:  J ECT       Date:  2022-03-01       Impact factor: 3.692

2.  Can Hypotension Episodes that were not Identified in the Non-Invasive Blood Pressure be Detected during Cesarean Section? A Randomized Controlled Trial

Authors:  Asude Ayhan; Nükhet Akovalı; Aynur Camkıran Fırat
Journal:  Balkan Med J       Date:  2022-08-23       Impact factor: 3.570

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.