Joanne McCormack1, Vincent Casey2, Richard Conway3, Jean Saunders4, Alison Perry5.
Abstract
BACKGROUND: Commercially available tools for measuring oro-lingual pressures during swallowing or isometric (tongue 'pushing') tasks have either poor, or unknown, psychometric properties (stability, reliability) which means their validity in a clinical setting is unknown. A new wireless tool, OroPress, has been designed to address the shortcomings of existing devices. In this pilot cohort study of normal adults (i.e., people without dysphagia), the face validity of OroPress was examined when it was used to measure oro-lingual pressures during (i) isometric tongue strength (ITS) tasks and (ii) isometric tongue endurance (ITE) tasks. The effects of gender on isometric oro-lingual data, captured using OroPress, were compared to published oro-lingual pressure data recorded using either the Kay Swallowing Workstation or the Iowa Oral Performance Instrument (aka commercial tools).Entities:
Mesh:
Year: 2015 PMID: 25889828 PMCID: PMC4403897 DOI: 10.1186/s12984-015-0024-6
Source DB: PubMed Journal: J Neuroeng Rehabil ISSN: 1743-0003 Impact factor: 4.262
A comparison of the IOPI, KSW& MOST tools for oro-lingual pressure measurement
|
|
|
|
|
|---|---|---|---|
| Visualisation of signal | Excellent | Good | Good |
| Quantitative data | Yes (optional) | Yes | Yes |
| Swallow Environment | Lab only | Clinic/Lab | Clinic/Lab |
| Sensor Array | 3 air filled sensors (hand-held & fixed position versions) | 1 liquid filled sensor (hand-held) | Custom-fit mouthpiece with four sensors |
| Ease of use | Needs skilled user | Easy to use | Reported to be easy to use |
| Data Integrity | Good-With fixed sensors | Poor–movement artefacts | Good |
| Intrusiveness | Intrusive-gag risk | Intrusive and fragile | Intrusive and effect normal swallow |
| Patient usage | Fixed position | Hand held device | Fixed position |
| Pressure Data | Swallowing, isotonic | Isotonic only | Swallowing, isotonic |
| Costs: | |||
| (a) Hardware | (a) d **€80-90k | (a) ** €4k | (a) Not available |
| (b) Probes | (b) **€40/single use | (b) ** €10/single use | (b) Not available |
a * KSW: Kay Swallowing Workstation.
b IOPI: Iowa Oral Performance Instrument.
c MOST: Madison Oral Strengthening Therapeutic device.
d ** €: Euro.
Reported gender differences in maximum isometric tongue strength pressures (PmaxS) of normal adults
|
|
|
|
|
|
|
|---|---|---|---|---|---|
| Vitorino, 2010 [ | IOPI | N = 75 | 56.81 ± 7.21 kPa | 56.37 ± 6.9 kPa | Yes |
| (m = 35, f = 40) | |||||
| Stierwalt & Youmans, 2007 [ | IOPI | N = 200 | 63.24 ± 13.86 kPa | 57.15 ± 13.50 kPa | Yes |
| (m = 80, f = 120) | |||||
| Youmans & Stierwalt, 2006 [ | IOPI | N = 90 | 64.0 ± 13.7 kPa | 55.9 ± 12.5 kPa | Yes |
| (m = 45, f = 45) | |||||
| Crow & Ship, 1996 [ | IOPI | N = 99 | 74.8 ± 18.9 kPa | 64.7 ± 19.6 kPa | Yes |
| (m = 52, f = 47) | |||||
| Nicosia et al., 2000 [ | KSW | N = 20 | Not provided | Not provided | No |
*Maximum pressure during an isometric tongue strength task.
[Mean and standard deviations are reported].
Figure 1The components of OroPress.
Figure 2Cross-sectional representation of OroPress sensor.
Figure 3A graphical representation of the calibration of the OroPress sensor.
Figure 4Positioning of OroPress sensor for measuring oro-lingual pressure.
Figure 5A graphical representation of the parameters used to measure isometric tongue strength and isometric tongue endurance.
Gender differences for normally distributed parameters-PmaxS and PmaxE
|
|
|
| |
|---|---|---|---|
| PMaxS (all) | 528.48 | 131.54 | 428.37, 574.37 |
| Males | 573.74 | 143.11 | 497.49, 650.00 |
| Females | 488.24 | 108.96 | 434.06, 542.43 |
| PMaxE (all) | 547.52 | 142.59 | 497.76, 597.27 |
| Males | 556.20 | 138.33 | 482.49, 629.91 |
| Females | 539.80 | 149.83 | 465.29, 614.31 |
Abbreviations: PMaxS, maximum oro-lingual pressure generated during isometric tongue strength task; PmaxE: maximum oro-lingual pressure generated during the isometric tongue endurance task; SD, Standard deviation; CI, Confidence interval.
Gender differences for not-normally distributed parameters-t100 and Pt100
|
|
| |
|---|---|---|
| t100 (all) | a *11.43 | 8.70, 17.88 |
| Males | *10.25 | 7.96, 16.80 |
| Females | *13.15 | 9.89, 20.50 |
| Pt100 (all) | b †4407.25 | 3448.35, 5775.60 |
| Males | †4246.10 | 3347.28, 6015.93 |
| Females | †4407.25 | 3629.98, 5775.60 |
Abbreviations: t100, the time for which the oro-lingual pressure is maintained above 100 mmHg during the isometric tongue endurance task; Pt100, the pressure–time product for the region where the oro-lingual pressure is maintained above 100 mmHg during the isometric tongue endurance task; IQR, Inter-quartile range.
a. *Seconds.
b. †mmHg/Sec.