Infusion pump-mediated mechanical hemolysis in pediatric patients.

Jonathan Hughes1, Janet McNaughton1, Jennifer Andrews1, Tracy George2, Cassandra Bergero3, Kimberly Pyke-Grimm3, Susan A Galel1, Christopher Gonzalez1, Lawrence Tim Goodnough4, Magali J Fontaine5.   

Abstract

CONTEXT: Hemoglobinuria was observed after packed red blood cell transfusion in a series of patients at our pediatric treatment center. Laboratory testing was suggestive of intravascular hemolysis with no support for an immunohematologic process.
OBJECTIVE: We investigated these adverse events to define a quality improvement plan and to prevent future hemolytic adverse events. Multiple factors were investigated, and the only change identified was the implementation of a new infusion pump (Pump A) that replaced a previous model (Pump B).
DESIGN: In vitro pump analyses, a retrospective review of urinalyses, and prospective urinalysis and nursing surveillances were also performed.
RESULTS: In in vitro analysis of the pumps, irradiated units with higher hematocrit at a low flow rate through Pump A had a greater than thirty-fold increase in free hemoglobin from baseline compared to minimal free hemoglobin changes seen with Pump B. Irradiated units with a lower hematocrit had a minimal change in free hemoglobin from baseline with both Pumps A and B at either low or high flow rate. Subsequently, only units with lower hematocrits were issued for transfusion of pediatric patients, and Pump A was replaced by Pump B in the outpatient unit. Retrospective and prospective surveillances found no additional unexplained cases of gross hemoglobinuria associated with transfusion.
CONCLUSION: The investigation determined that infusion of higher hematocrit units using a specific commercial pump was associated with mechanical hemolysis. The change to units with lower hematocrit through an alternative pump has been an effective corrective action to date.
© 2015 by the Association of Clinical Scientists, Inc.

Entities:  

Keywords:  corrective action; infusion pump; mechanical hemolysis; quality improvement

Mesh:

Substances:

Year:  2015        PMID: 25887866

Source DB:  PubMed          Journal:  Ann Clin Lab Sci        ISSN: 0091-7370            Impact factor:   1.256


  1 in total

1.  Hemolysis risk after packed red blood cells transfusion with infusion pumps.

Authors:  Ana Maria Miranda Martins Wilson; Maria Angélica Sorgini Peterlini; Mavilde da Luz Gonçalves Pedreira
Journal:  Rev Lat Am Enfermagem       Date:  2018-10-11
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.