| Literature DB >> 25884838 |
Ross A Clark1, Stephanie Vernon2, Benjamin F Mentiplay3, Kimberly J Miller4, Jennifer L McGinley5, Yong Hao Pua6, Kade Paterson7, Kelly J Bower8,9.
Abstract
BACKGROUND: The Microsoft Kinect has been used previously to assess spatiotemporal aspects of gait; however the reliability of this system for the assessment of people following stroke has not been established. This study examined the reliability and additional information that the Kinect provides when instrumenting a gait assessment in people living with stroke.Entities:
Mesh:
Year: 2015 PMID: 25884838 PMCID: PMC4333881 DOI: 10.1186/s12984-015-0006-8
Source DB: PubMed Journal: J Neuroeng Rehabil ISSN: 1743-0003 Impact factor: 4.262
Participants’ characteristics recorded during the first assessment session
|
|
|
|---|---|
| Age (yrs) | 68 ± 15 |
| Height (cm) | 166.7 ± 9.4 |
| Body Mass (kg) | 72.5 ± 11.9 |
| Gender (n = male) | 21 |
| Mini-mental State exam score (/30) | 27.1 ± 2.7 |
| Lesion type (% Infarct) | 73 |
| Lesion side (% Right) | 63 |
| Time since stroke (months) | 21 ± 19 |
| Hypertension (%) | 70 |
| Dyslipidaemia (%) | 47 |
| Type 2 Diabetes mellitus (%) | 27 |
| 10 m Walk test speed (m/s) | 0.97 ± 0.35 |
| 10 m Walk test steps (n) | 20 ± 9 |
| Timed up and go (s) | 17.7 ± 10.5 |
| Functional reach (cm) | 28.5 ± 7.4 |
| Step test (n) | 9.3 ± 4.9 |
| Static double leg eyes closed balance – Path velocity (cm/s) | 2.06 ± 1.05 |
| Static weight-bearing asymmetry (% Body mass) | 46.3 ± 8.5 |
| Medial-lateral weight shifting ability (no. of shifts in 30 seconds) | 10.0 ± 3.7 |
Values are reported as mean ± standard deviation unless otherwise indicated.
Test-retest reliability measures for the Kinect-derived gait variables
|
|
|
|
|
|
|
|---|---|---|---|---|---|
|
|
|
| |||
| Affected step length (mm) | 507 ± 147 | 513 ± 169 | 0.97 (0.95 to 0.99) | 29 | 80 |
| Unaffected step length (mm) | 520 ± 166 | 520 ± 144 | 0.98 (0.96 to 0.99) | 21 | 58 |
| Step length asymmetry (%) | 15.0 ± 15.6 | 17.4 ± 19.4 | 0.89 (0.76 to 0.95) | 5.2 | 14.3 |
| Affected foot swing velocity (m/s) | 3.18 ± 0.91 | 3.16 ± 0.90 | 0.97 (0.94 to 0.99) | 0.16 | 0.44 |
| Unaffected foot swing velocity (m/s) | 3.08 ± 1.3 | 3.09 ± 1.29 | 0.99 (0.97 to 0.99) | 0.13 | 0.36 |
| Foot swing velocity asymmetry (%) | 12.6 ± 13.7 | 14.0 ± 13.0 | 0.80 (0.58 to 0.90) | 6.1 | 17.0 |
| Mean velocity (m/s) | 0.89 ± 0.33 | 0.90 ± 0.32 | 0.98 (0.96 to 0.99) | 0.05 | 0.13 |
| Peak velocity (m/s) | 1.22 ± 0.39 | 1.22 ± 0.39 | 0.98 (0.95 to 0.99) | 0.06 | 0.15 |
| Peak – Mean velocity difference (%) | 41.8 ± 18.4 | 39.8 ± 18.1 | 0.96 (0.92 to 0.98) | 3.7 | 10.2 |
Abbreviations: SD = standard deviation; ICC = intraclass correlation coefficient; CI = confidence interval; SEM = standard error of measurement; MDC = minimum detectable change.
Redundancy analysis of the Kinect-derived gait variables compared to the clinical measures of manually assessed time (MT) and manually assessed number of steps (MS) using Spearman’s correlation
|
|
|
|
|---|---|---|
| Affected step length (mm) | −0.86* | −0.93* |
| Unaffected step length (mm) | −0.93* | −0.96* |
| Step length asymmetry (%) | 0.45* | 0.39 |
| Affected foot swing velocity (mm/s) | −0.71* | −0.76* |
| Unaffected foot swing velocity (mm/s) | −0.78* | −0.80* |
| Foot swing velocity asymmetry (%) | 0.25 | 0.20 |
| Mean velocity (mm/s) | −0.92* | −0.92* |
| Peak velocity (mm/s) | −0.92* | −0.93* |
| Peak – Mean velocity difference (%) | 0.66* | 0.65* |
*significant at P < 0.05.
The correlation between MT and MS is 0.93*.
Redundancy analysis of the Kinect-derived gait variables using Spearman’s correlation matrix
|
|
|
|
|
|
|---|---|---|---|---|
|
|
| |||
| Peak – Mean velocity difference | 0.53* | −0.59* | −0.60* | 0.21 |
| Foot swing velocity asymmetry | 0.02 | 0.06 | −0.16 | |
| Unaffected foot swing velocity | −0.26 | 0.91* | ||
| Affected foot | −0.31 |
*significant at P < 0.05.
Results for the clinically-derived and Kinect-derived multiple regression models
|
|
|
| ||||
|---|---|---|---|---|---|---|
|
|
|
|
|
|
| |
|
| ||||||
| Timed up and go | 0.89 | 0.88** | 81 | 0.95 | 0.94** | 63 |
|
| ||||||
| Static double leg eyes closed balance | 0.41 | 0.37* | −8 | 0.47 | 0.36* | −5 |
| Static weight-bearing asymmetry | 0.49 | 0.44** | 102 | 0.50 | 0.38* | 108 |
|
| ||||||
| Functional reach | 0.43 | 0.39* | 109 | 0.65 | 0.58** | 100 |
| Step test | 0.61 | 0.58** | 70 | 0.83 | 0.79** | 52 |
| Medial-lateral weight shifting ability | 0.48 | 0.44** | 60 | 0.75 | 0.69** | 46 |
Clinically-derived model included 2 input variables: 10MWT time; total number of steps. Kinect-derived model included 5 input variables: Mean velocity; Foot swing velocity asymmetry; Peak – Mean velocity difference; Step length asymmetry; Affected foot swing velocity.
*p-values <0.01; **p-values <0.001.
R2, Ezekiel equation adjusted R2 and the Akaike Information Criterion (AIC) score are provided for comparison of model strengths.