Correlation between Initial BIC and the Insertion Torque/Depth Integral Recorded with an Instantaneous Torque-Measuring Implant Motor: An in vivo Study.

Paolo Capparé1, Raffaele Vinci1, Danilo Alessio Di Stefano1, Tonino Traini1, Giuseppe Pantaleo2, Enrico Felice Gherlone1, Giorgio Gastaldi3.   

Abstract

BACKGROUND: Quantitative intraoperative evaluation of bone quality at implant placement site and postinsertion implant primary stability assessment are two key parameters to perform implant-supported rehabilitation properly. A novel micromotor has been recently introduced allowing to measure bone density at implant placement site and to record implant insertion-related parameters, such as the instantaneous, average and peak insertion torque values, and the insertion torque/depth integral.
PURPOSE: The aim of this study was to investigate in vivo if any correlation existed between initial bone-to-implant contact (BIC) and bone density and integral values recorded with the instrument.
MATERIALS AND METHODS: Twenty-five patients seeking for implant-supported rehabilitation of edentulous areas were consecutively treated. Before implant placement, bone density at the insertion site was measured. For each patient, an undersized 3.3 × 8-mm implant was placed, recording the insertion torque/depth integral values. After 15 minutes, the undersized implant was retrieved with a 0.5 mm-thick layer of bone surrounding it. Standard implants were consequently placed. Retrieved implants were analyzed for initial BIC quantification after fixation, dehydration, acrylic resin embedment, sections cutting and grinding, and toluidine-blue and acid fuchsine staining. Correlation between initial BIC values, bone density at the insertion site, and the torque/depth integral values was investigated by linear regression analysis.
RESULTS: A significant linear correlation was found to exist between initial BIC and (a) bone density at the insertion site (R = 0.96, explained variance R(2)  = 0.92) and (b) torque/depth integral at placement (R = 0.81, explained variance R(2)  = 0.66).
CONCLUSIONS: The system provided quantitative, reliable data correlating significantly with immediate postinsertion initial BIC, and could therefore represent a valuable tool both for clinical research and for the oral implantologist in his/her daily clinical practice.
© 2015 Wiley Periodicals, Inc.

Entities:  

Keywords:  bone density; bone to implant contact; dental implant; histomorphometry; implant stability; insertion torque; primary stability

Mesh:

Substances:

Year:  2015        PMID: 25876078     DOI: 10.1111/cid.12294

Source DB:  PubMed          Journal:  Clin Implant Dent Relat Res        ISSN: 1523-0899            Impact factor:   3.932


  10 in total

Review 1.  Efficacy of parathyroid hormone supplementation on the osseointegration of implants: a systematic review.

Authors:  Fawad Javed; Mohammad D Al Amri; Sergio Varela Kellesarian; Abdulaziz A Al-Kheraif; Fahim Vohra; José Luis Calvo-Guirado; Hans Malmstrom; Georgios E Romanos
Journal:  Clin Oral Investig       Date:  2015-12-22       Impact factor: 3.573

2.  Inhibition of cathepsin K promotes osseointegration of titanium implants in ovariectomised rats.

Authors:  Chun Yi; Ke-Yi Hao; Ting Ma; Ye Lin; Xi-Yuan Ge; Yu Zhang
Journal:  Sci Rep       Date:  2017-03-17       Impact factor: 4.379

Review 3.  Osseodensification - A novel approach in implant dentistry.

Authors:  Umesh Y Pai; Shobha J Rodrigues; Karishma S Talreja; Mahesh Mundathaje
Journal:  J Indian Prosthodont Soc       Date:  2018 Jul-Sep

4.  Correlation between Implant Geometry, Bone Density, and the Insertion Torque/Depth Integral: A Study on Bovine Ribs.

Authors:  Danilo Alessio Di Stefano; Paolo Arosio; Vittoria Perrotti; Giovanna Iezzi; Antonio Scarano; Adriano Piattelli
Journal:  Dent J (Basel)       Date:  2019-03-05

5.  Bone Density and Implant Primary Stability. A Study on Equine Bone Blocks.

Authors:  Francesco Orlando; Federico Arosio; Paolo Arosio; Danilo Alessio Di Stefano
Journal:  Dent J (Basel)       Date:  2019-07-01

Review 6.  Stability of Dental Implants and Thickness of Cortical Bone: Clinical Research and Future Perspectives. A Systematic Review.

Authors:  Danilo Alessio Di Stefano; Paolo Arosio; Paolo Capparè; Silvia Barbon; Enrico Felice Gherlone
Journal:  Materials (Basel)       Date:  2021-11-25       Impact factor: 3.623

7.  New Osseodensification Implant Site Preparation Method to Increase Bone Density in Low-Density Bone: In Vivo Evaluation in Sheep.

Authors:  Paolo Trisi; Marco Berardini; Antonello Falco; Michele Podaliri Vulpiani
Journal:  Implant Dent       Date:  2016-02       Impact factor: 2.454

8.  The effect of undersizing and tapping on bone to implant contact and implant primary stability: A histomorphometric study on bovine ribs.

Authors:  Danilo Alessio Di Stefano; Vittoria Perrotti; Gian Battista Greco; Claudia Cappucci; Paolo Arosio; Adriano Piattelli; Giovanna Iezzi
Journal:  J Adv Prosthodont       Date:  2018-06-12       Impact factor: 1.904

9.  Effects of implant length and 3D bone-to-implant contact on initial stabilities of dental implant: a microcomputed tomography study.

Authors:  Jui-Ting Hsu; Aaron Yu-Jen Wu; Lih-Jyh Fuh; Heng-Li Huang
Journal:  BMC Oral Health       Date:  2017-11-21       Impact factor: 2.757

10.  New Implant Design with Midcrestal and Apical Wing Thread for Increased Implant Stability in Single Postextraction Maxillary Implant.

Authors:  Antonio Scarano; Bartolomeo Assenza; Francesco Inchingolo; Filiberto Mastrangelo; Felice Lorusso
Journal:  Case Rep Dent       Date:  2019-09-05
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.