Stability of edentulous, atrophic mandibles after insertion of different dental implants. A biomechanical study.

T Torsiglieri1, S Raith2, A Rau3, H Deppe3, F Hölzle4, T Steiner4.   

Abstract

OBJECTIVES: Fractures of the atrophic edentulous mandible are a rare complication that can become severe after the insertion of dental implants. This in vitro study investigated the effects of different implant settings varying in number, diameter, and length. and the influence of a fixed bar.
MATERIALS AND METHODS: In biomechanical experiments on artificial mandibles, an unmodified reference group, four implant settings with two different implants, and the effect of adding a fixed bar to these settings were tested. All specimens were loaded with incisal biting forces until failure due to fracture.
RESULTS: Implants weakened all specimens significantly compared with those in the reference group. Without a fixed bar, four short and thick implants showed the best results, with high significance. With a fixed bar, four long and thin implants withstood the highest loads. The addition of fixed bars reduced the differences between the implant settings. Fixed bars did not show increased stability for all groups; however, these groups showed a higher mean strength.
CONCLUSIONS: Four implants with a short and thick design should be the first choice when implants are placed without a fixed bar in an atrophic mandible. With a fixed bar, four long and thin implants should be used.
Copyright © 2015 European Association for Cranio-Maxillo-Facial Surgery. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Dental implant; Mandibular atrophy; Mandibular fracture

Mesh:

Substances:

Year:  2015        PMID: 25865491     DOI: 10.1016/j.jcms.2015.03.001

Source DB:  PubMed          Journal:  J Craniomaxillofac Surg        ISSN: 1010-5182            Impact factor:   2.078


  2 in total

1.  Twist removal of healed vs. nonhealed implants-a mechanical and histological study in mini pigs.

Authors:  Ricardo de Oliveira Silva; Fabrício Passador; Paulo Henrique Ferreira Caria
Journal:  Int J Implant Dent       Date:  2016-11-25

2.  Evaluating the biomechanical effects of implant diameter in case of facial trauma to an edentulous atrophic mandible: a 3D finite element analysis.

Authors:  Aysa Ayali; Kani Bilginaylar
Journal:  Head Face Med       Date:  2017-05-02       Impact factor: 2.151

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.