An In Vitro Comparison of Fracture Load of Zirconia Custom Abutments with Internal Connection and Different Angulations and Thicknesses: Part II.

Roya Zandparsa1, Abdalah Albosefi1.   

Abstract

PURPOSE: The purpose of part II of this in vitro study was to compare the fracture load of two-piece zirconia custom abutments with different thicknesses and angulations.
MATERIALS AND METHODS: Forty zirconia custom abutments were divided into four groups as follows: group A1: 0.7 mm thickness and 0° angulations; group A2: 0.7 mm thickness and 15° angulations; group B1: 1 mm thickness and 0° angulations; group B2: 1 mm thickness and 15° angulations. As in part I, in all groups, implant replicas were mounted in self-cure acrylic jigs to support the abutments. The zirconia custom abutments were engaged in the implant replicas using a manual torque wrench. All jigs were secured and mounted in a metallic vice and subjected to shear stress till failure using a universal testing machine with a 0.5 mm/min crosshead speed with the force transferred to the lingual surface of the zirconia custom abutments 2 mm below the incisal edge. The test specimens used in this study did not include a crown. The universal testing machine was controlled via a computer software system, which also completed the stress-strain diagram and recorded the breaking fracture load. The fracture loads were recorded for comparison among the groups and subjected to statistical analysis (two-way ANOVA and Kolmogorov-Smirnov).
RESULTS: The mean fracture load of zirconia custom abutments across the groups (A1 to B2) ranged from 432 ± 97 N to 746 ± 275 N. The angulated zirconia custom abutment exhibited the highest fracture load, which was statistically significant (p = 0.045). The thickness of the zirconia custom abutment also had a positive influence on the strength of the specimens (p = 0.005).
CONCLUSIONS: In this study, the 15° angulated zirconia custom abutments showed the highest fracture load of those investigated. The 1 mm thick zirconia custom abutments also exhibited significantly higher fracture load compared to 0.7 mm abutments. CLINICAL IMPLICATIONS: The results of this in vitro study will help dental practitioners with their decision-making process in selecting the type of custom abutment to be used clinically.
© 2015 by the American College of Prosthodontists.

Entities:  

Keywords:  Ceramic abutments stability; fracture load; implant/abutment connection, ceramics; shear stress; zirconia

Mesh:

Substances:

Year:  2015        PMID: 25865430     DOI: 10.1111/jopr.12292

Source DB:  PubMed          Journal:  J Prosthodont        ISSN: 1059-941X            Impact factor:   2.752


  3 in total

1.  Mechanical stability of angulated zirconia abutments supporting maxillary anterior single crowns on narrow-diameter implants.

Authors:  Ruiyang Ma; Ping Yu; Yuqiang Zhang; Chenyang Xie; Xin Tan; Jikui Sun; Haiyang Yu
Journal:  Clin Oral Investig       Date:  2022-09-26       Impact factor: 3.606

2.  Clinical and Laboratory Outcomes of Angled Screw Channel Implant Prostheses: A Systematic Review.

Authors:  Vanya Rasaie; Jaafar Abduo; Mehran Falahchai
Journal:  Eur J Dent       Date:  2022-02-21

3.  Effect of Implant Platform Connection and Abutment Material on Removal Torque and Implant Hexagon Plastic Deformation.

Authors:  Paulo Henrique de Souza Pereira; Marina Amaral; Kusai Baroudi; Rafael Pino Vitti; Mohammad Zakaria Nassani; Laís Regiane da Silva-Concílio
Journal:  Eur J Dent       Date:  2019-12-03
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.