Ling Liu1, Lu Bai1, Dengbin Yu1, Junfeng Zhai1, Shaojun Dong2.
Abstract
Using mediator as electron acceptor for biochemical oxygen demand (BOD) measurement was developed in the last decade (BODMed). However, until now, no BOD(Med) in a flow system has been reported. This work for the first time describes a flow system of BOD(Med) method (BOD(Med)-FS) by using potassium ferricyanide as mediator and carbon fiber felt as substrate material for microbial immobilization. The system can determine the BOD value within 30 min and possesses a wider analytical linear range for measuring glucose-glutamic acid (GGA) standard solution from 2 up to 200 mg L(-1) without the need of dilution. The analytical performance of the BOD(Med)-FS is comparable or better than that of the previously reported BOD(Med) method, especially its superior long-term stability up to 2 months under continuous operation. Moreover, the BOD(Med)-FS has same determination accuracy with the conventional BOD5 method by measuring real samples from a local wastewater treatment plant (WWTP).Entities:
Keywords: BOD(Med); Flow system; Native biofilm; Rapid BOD
Mesh:
Substances:
Year: 2015 PMID: 25863368 DOI: 10.1016/j.talanta.2015.02.001