| Literature DB >> 25861577 |
Sang-Hee Park1, Chang-Ki Kim2, Hye-Ran Jeong3, Hyunjin Son3, Seong-Han Kim1, Mi-Sun Park1.
Abstract
OBJECTIVES: A fast and accurate diagnosis is necessary to control and eliminate tuberculosis (TB). In Korea, TB continues to be a serious public health problem. In this study, diagnostic tests on clinical samples from patients suspected to have TB were performed and the sensitivity and specificity of the various techniques were compared. The main objective of the study was to compare various diagnostic tests and evaluate their sensitivity and specificity for detecting tuberculosis.Entities:
Keywords: acid-fast bacilli culture; real-time polymerase chain reaction; smear microscopy; tuberculosis
Year: 2014 PMID: 25861577 PMCID: PMC4301634 DOI: 10.1016/j.phrp.2014.10.006
Source DB: PubMed Journal: Osong Public Health Res Perspect ISSN: 2210-9099
Locationwise data of total samples tested and positivity rates for Mycobacterium tuberculosis.
| Location | No. of samples tested | No. of positive cultures | Rate of positive culture (%) |
|---|---|---|---|
| Seoul | 24,000 | 1,855 | 7.7 |
| Busan | 16,595 | 790 | 4.8 |
| Incheon | 9,626 | 483 | 5.0 |
| Gyeonggi-do | 26,788 | 1,554 | 5.8 |
| Gangwon-do | 13,815 | 295 | 2.1 |
| Chungcheongbuk-do | 6,230 | 211 | 3.4 |
| Chungcheongnam-do | 14,090 | 567 | 4.0 |
| Jeollabuk-do | 6,982 | 291 | 4.2 |
| Jeollanam-do | 18,319 | 460 | 2.5 |
| Gyeongsangbuk-do | 12,899 | 679 | 5.3 |
| Gyeongsangnam-do | 16,000 | 723 | 4.5 |
| Jeju-do | 4,104 | 117 | 2.9 |
| Osong (central center) | 792 | 191 | 24.1 |
| Total | 170,240 | 8,216 | 4.8 |
Figure 1Monthly distribution of sample numbers tested and detections of Mycobacterium tuberculosis (culture results).
Comparisons of smear microscopy, real-time PCR, and culture results.
| Solid culture | Liquid culture | |||
|---|---|---|---|---|
| Smear microscopy | Real-time PCR | Smear microscopy | Real-time PCR | |
| No. of samples | 167,538 | 3,692 | 10,144 | 1,583 |
| True positive | 4,633 | 2,185 | 1,524 | 826 |
| False positive | 684 | 298 | 105 | 51 |
| True negative | 158,697 | 1,146 | 7,604 | 649 |
| False negative | 3,524 | 63 | 911 | 57 |
| Sensitivity (95% CI) | 56.8% (55.7–57.9) | 97.2% (96.4–97.84) | 62.6% (60.6–64.5) | 93.5% (91.7–95.1) |
| Specificity (95% CI) | 99.6% (99.5–99.6) | 79.4% (77.2–81.4) | 98.6% (98.4–98.9) | 92.7% (90.5–94.5) |
| <0.001 | <0.001 | <0.001 | <0.001 | |
CI = confidence interval; PCR = polymerase chain reaction.